  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Oxford",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73459",
      "recalling_firm": "Villa Radiology Systems LLC",
      "address_1": "91 Willenbrock Rd Ste B1",
      "address_2": "N/A",
      "postal_code": "06478-1036",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AL  AZ  CA  CO  CT  DC  FL  IA  IL  IN  LA  MA  MD  MI  MN  MO  MS  NJ  NY  OH  PA  Puerto Rico  TN  TX  UT  WI",
      "recall_number": "Z-1063-2017",
      "product_description": "Juno DRF; Model: 709020    Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.",
      "product_quantity": "66 distributed in US",
      "reason_for_recall": "The analysis showed that the incident was caused by a chain of four faults:  1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.  2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area.  3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation).  4.) Delayed activation of the emergency stop.",
      "recall_initiation_date": "20160219",
      "center_classification_date": "20170120",
      "report_date": "20170201",
      "code_info": "10110770  11030805  11030806  11070845  11070848  11070849  11090864  11110900  12020978  12020982  12030999  12041003  12061014  12061018  12061020  12081040  12091074  12101104  12121186  12121187  13031215  13101263  14021301  14031313  14041321  14051332  14071353  14081359  14091368  14091371  14121400  14121403  15021426  15031440  15041444  15091473  15091478  15101484  15101487"
    }
  ]
}