      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69863",
      "recalling_firm": "Maquet Datascope Corp - Cardiac Assist Division",
      "address_1": "1300 MacArthur Blvd.",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-1062-2015",
      "product_description": "CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units.  Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx    An electromechanical system used to inflate and deflate intra-aortic balloons.",
      "product_quantity": "838 units-US; 364 units - OUS",
      "reason_for_recall": "Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.",
      "recall_initiation_date": "20141216",
      "center_classification_date": "20150206",
      "termination_date": "20170609",
      "report_date": "20150218",
      "code_info": "Sensation 7 fr.34 cc IAB Demonstration Unit    0684-DC-0433  0684-DO-0433  0684-DC-0469  0684-DO-0469    Sensation 7 Fr. 40 cc IAB Demonstration Unit  0684-DC-0434  0684-DO-0434  0684-DC-0470  0684-DO-0470    Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units   0684-DC-0567  0684-DO-0567  0684-DC-0568  0684-DO-0568    Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units  0684-DC-0575  0684-DO-0575  0684-DC-0576  0684-DO-0576    Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit  0684-DC-0272  0684-DO-0272"
    }
  ]
}