{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "El Segundo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89880",
      "recalling_firm": "Karl Storz Endoscopy",
      "address_1": "2151 E Grand Ave",
      "address_2": "N/A",
      "postal_code": "90245-5017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY.  O.U.S.: Mexico, Lebanon, and South Korea",
      "recall_number": "Z-1061-2022",
      "product_description": "CYSTO-URETHRO-FIBERSCOPE REF 11272C1    CYSTOSCOPE, 15FR X 370MM, ADULT  REF 11272C2 and Kit Number 11272CK2    CYSTO-URETHRO-FIBERSCOPE  REF 11272CU1 and Kit Number 11272CUK1",
      "product_quantity": "All Serial Numbers manufactured/distributed since January 2018",
      "reason_for_recall": "Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.",
      "recall_initiation_date": "20220401",
      "center_classification_date": "20220512",
      "report_date": "20220518",
      "code_info": "11272C1 n/a Flexible Cysto-Urethroscope Fiberscope Z18449US-BD (08-2018) UDI Code: 4048551226100; 11272C2 11272CK2 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551226100; 11272CU1 11272CUK1 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551226162;  All Serial Numbers manufactured/distributed since January 2018."
    }
  ]
}