      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5412-015-02 61777756  00-5412-015-02 11003149  00-5412-015-02 61782977  00-5412-015-02 61796047  00-5412-015-02 61806563  00-5412-015-02 61806562  00-5412-015-02 61867212  00-5412-015-02 61867213  00-5412-015-02 61872035  00-5412-015-02 61871424  00-5412-015-02 11004422  00-5412-015-02 11004477  00-5412-015-02 61876226  00-5412-015-02 61885240  00-5412-015-02 61889370  00-5412-015-02 61894436  00-5412-015-02 61899511  00-5412-015-02 11001440  00-5412-015-02 61904154  00-5412-015-02 61916029  00-5412-015-02 61921361  00-5412-015-02 61928677  00-5412-015-02 61940207  00-5412-015-02 61952635  00-5412-015-02 61963807  00-5412-015-02 61968948  00-5412-015-02 61988106  00-5412-015-02 11005840  00-5412-015-02 61983536  00-5412-015-02 61975655  00-5412-015-02 61995014  00-5412-015-02 61988107  00-5412-015-02 62001897  00-5412-015-02 62013575  00-5412-015-02 62005775  00-5412-015-02 62013586  00-5412-015-02 62013587  00-5412-015-02 11006300  00-5412-015-02 62020383  00-5412-015-02 62025800  00-5412-015-02 62025801  00-5412-015-02 62041160  00-5412-015-02 62032711  00-5412-015-02 62054603  00-5412-015-02 62056958  00-5412-015-02 62062122  00-5412-015-02 62068182  00-5412-015-02 62080694  00-5412-015-02 62074905  00-5412-015-02 62086883  00-5412-015-02 62104646  00-5412-015-02 62104205  00-5412-015-02 62110690  00-5412-015-02 62117168  00-5412-015-02 62122805  00-5412-015-02 62128264  00-5412-016-01 61777757  00-5412-016-01 61777758  00-5412-016-01 61777751  00-5412-016-01 61789818  00-5412-016-01 61782969  00-5412-016-01 61782978  00-5412-016-01 61789819  00-5412-016-01 61789821  00-5412-016-01 61789820  00-5412-016-01 11003194  00-5412-016-01 61796051  00-5412-016-01 61796048  00-5412-016-01 61796049  00-5412-016-01 61796050  00-5412-016-01 11003207  00-5412-016-01 61806564  00-5412-016-01 61806565  00-5412-016-01 61810244  00-5412-016-01 61810277  00-5412-016-01 61810278  00-5412-016-01 61822640  00-5412-016-01 61824650  00-5412-016-01 61831766  00-5412-016-01 61831767  00-5412-016-01 61831768  00-5412-016-01 61831769  00-5412-016-01 61837030  00-5412-016-01 61855584  00-5412-016-01 61858029  00-5412-016-01 61862203  00-5412-016-01 61862254  00-5412-016-01 11004099  00-5412-016-01 61867302  00-5412-016-01 61871426  00-5412-016-01 61871427  00-5412-016-01 61872036  00-5412-016-01 61876227  00-5412-016-01 61876228  00-5412-016-01 61885241  00-5412-016-01 61872037  00-5412-016-01 11004486  00-5412-016-01 11004488  00-5412-016-01 11004571  00-5412-016-01 11004569  00-5412-016-01 61885242  00-5412-016-01 61894438  00-5412-016-01 61889371  00-5412-016-01 61889372  00-5412-016-01 61889373  00-5412-016-01 61894437  00-5412-016-01 61904155  00-5412-016-01 61910532  00-5412-016-01 61899549  00-5412-016-01 61910533  00-5412-016-01 11004855  00-5412-016-01 61916030  00-5412-016-01 61910534  00-5412-016-01 61904156  00-5412-016-01 11004880  00-5412-016-01 11004921  00-5412-016-01 61921362  00-5412-016-01 61921363  00-5412-016-01 61928678  00-5412-016-01 61928679  00-5412-016-01 61933415  00-5412-016-01 61933416  00-5412-016-01 61936720  00-5412-016-01 11005190  00-5412-016-01 11005070  00-5412-016-01 61940208  00-5412-016-01 11005071  00-5412-016-01 61963809  00-5412-016-01 61950171  00-5412-016-01 61952636  00-5412-016-01 61952637  00-5412-016-01 61975656  00-5412-016-01 61958443  00-5412-016-01 61963808  00-5412-016-01 61975657  00-5412-016-01 61958444  00-5412-016-01 61983537  00-5412-016-01 61986587  00-5412-016-01 61988108  00-5412-016-01 61983538  00-5412-016-01 61968949  00-5412-016-01 11005766  00-5412-016-01 61968951  00-5412-016-01 61975660  00-5412-016-01 61975658  00-5412-016-01 61968950  00-5412-016-01 61975659  00-5412-016-01 61991994  00-5412-016-01 61995015  00-5412-016-01 61983550  00-5412-016-01 11005970  00-5412-016-01 61988109  00-5412-016-01 11005971  00-5412-016-01 61995017  00-5412-016-01 61991990  00-5412-016-01 61995016  00-5412-016-01 62001900  00-5412-016-01 62001901  00-5412-016-01 11006008  00-5412-016-01 62001898  00-5412-016-01 11006100  00-5412-016-01 62005777  00-5412-016-01 62005776  00-5412-016-01 62013577  00-5412-016-01 62013576  00-5412-016-01 62013589  00-5412-016-01 62001899  00-5412-016-01 11006234  00-5412-016-01 62013588  00-5412-016-01 62020384  00-5412-016-01 62020386  00-5412-016-01 62020385  00-5412-016-01 62020387  00-5412-016-01 62032712  00-5412-016-01 62032713  00-5412-016-01 11006309  00-5412-016-01 62041161  00-5412-016-01 62041162  00-5412-016-01 62041164  00-5412-016-01 62041163  00-5412-016-01 62046174  00-5412-016-01 62046173  00-5412-016-01 62054604  00-5412-016-01 62054605  00-5412-016-01 62056959  00-5412-016-01 62056961  00-5412-016-01 11006457  00-5412-016-01 62056960  00-5412-016-01 62062123  00-5412-016-01 62062124  00-5412-016-01 62068183  00-5412-016-01 62062125  00-5412-016-01 62074907  00-5412-016-01 62080688  00-5412-016-01 62074906  00-5412-016-01 62074908  00-5412-016-01 62080686  00-5412-016-01 62086884  00-5412-016-01 62080687  00-5412-016-01 62080689  00-5412-016-01 62086885  00-5412-016-01 62092463  00-5412-016-01 62092462  00-5412-016-01 62094870  00-5412-016-01 62103504  00-5412-016-01 11006982  00-5412-016-01 62104181  00-5412-016-01 11007026  00-5412-016-01 62104207  00-5412-016-01 62104207R  00-5412-016-01 62104180  00-5412-016-01 62110691  00-5412-016-01 62110692  00-5412-016-01 62110699  00-5412-016-01 62122806  00-5412-016-01 62128265  00-5412-016-01 62128266  00-5412-016-01 62128267  00-5412-016-01 62132476  00-5412-017-01 61777762  00-5412-017-01 61777761  00-5412-017-01 61782970  00-5412-017-01 61782980  00-5412-017-01 61789827  00-5412-017-01 61789828  00-5412-017-01 61782971  00-5412-017-01 61789826  00-5412-017-01 61789840  00-5412-017-01 61796054  00-5412-017-01 11003172  00-5412-017-01 11003197  00-5412-017-01 61801505  00-5412-017-01 61796055  00-5412-017-01 61801507  00-5412-017-01 61801526  00-5412-017-01 61806570  00-5412-017-01 61810246  00-5412-017-01 11003218  00-5412-017-01 61801506  00-5412-017-01 61801508  00-5412-017-01 61815751  00-5412-017-01 61810281  00-5412-017-01 61810282  00-5412-017-01 61815752  00-5412-017-01 61815753  00-5412-017-01 61824653  00-5412-017-01 61822644  00-5412-017-01 61831774  00-5412-017-01 61824652  00-5412-017-01 61837034  00-5412-017-01 61837035  00-5412-017-01 61871431  00-5412-017-01 61871433  00-5412-017-01 61872040  00-5412-017-01 61867303  00-5412-017-01 61867303R  00-5412-017-01 61885245  00-5412-017-01 61885246  00-5412-017-01 61885247  00-5412-017-01 61876231  00-5412-017-01 11004611  00-5412-017-01 61889376  00-5412-017-01 11004423  00-5412-017-01 61894442  00-5412-017-01 61899513  00-5412-017-01 61894443  00-5412-017-01 61899551  00-5412-017-01 61904159  00-5412-017-01 11004762  00-5412-017-01 61916033  00-5412-017-01 61910535  00-5412-017-01 61910567  00-5412-017-01 61916034  00-5412-017-01 11004927  00-5412-017-01 61916035  00-5412-017-01 11004963  00-5412-017-01 11004964  00-5412-017-01 61921366  00-5412-017-01 61921367  00-5412-017-01 11005044  00-5412-017-01 61940212  00-5412-017-01 61928682  00-5412-017-01 11005301  00-5412-017-01 61936723  00-5412-017-01 61940213  00-5412-017-01 61936722  00-5412-017-01 61950175  00-5412-017-01 61952640  00-5412-017-01 61958448  00-5412-017-01 61975664  00-5412-017-01 61975665  00-5412-017-01 61952641  00-5412-017-01 61963813  00-5412-017-01 61963812  00-5412-017-01 61968955  00-5412-017-01 61986590  00-5412-017-01 61968954  00-5412-017-01 61983542  00-5412-017-01 61991992  00-5412-017-01 61988112  00-5412-017-01 61958447  00-5412-017-01 11005768  00-5412-017-01 61975666  00-5412-017-01 61975667  00-5412-017-01 61983544  00-5412-017-01 61983543  00-5412-017-01 61991996  00-5412-017-01 61983552  00-5412-017-01 61995023  00-5412-017-01 61995022  00-5412-017-01 62001906  00-5412-017-01 62005780  00-5412-017-01 62005805  00-5412-017-01 62005781  00-5412-017-01 62013580  00-5412-017-01 62013581  00-5412-017-01 62013592  00-5412-017-01 62001905  00-5412-017-01 62020393  00-5412-017-01 62020392  00-5412-017-01 62025804  00-5412-017-01 62032716  00-5412-017-01 62032717  00-5412-017-01 62041169  00-5412-017-01 62025803  00-5412-017-01 62041168  00-5412-017-01 62041170  00-5412-017-01 11006518  00-5412-017-01 62054608  00-5412-017-01 62054609  00-5412-017-01 11006585  00-5412-017-01 62056964  00-5412-017-01 62056965  00-5412-017-01 62056965R  00-5412-017-01 62074911  00-5412-017-01 62062131  00-5412-017-01 62062132  00-5412-017-01 62062130  00-5412-017-01 62068188  00-5412-017-01 62068189  00-5412-017-01 62074912  00-5412-017-01 11006765  00-5412-017-01 11006815  00-5412-017-01 62086890  00-5412-017-01 62092467  00-5412-017-01 62080692  00-5412-017-01 62092469  00-5412-017-01 62104647  00-5412-017-01 11007002  00-5412-017-01 62104206  00-5412-017-01 62046178  00-5412-017-01 62092468  00-5412-017-01 62110695  00-5412-017-01 62110696  00-5412-017-01 62080695  00-5412-017-01 62117175  00-5412-017-01 62122808  00-5412-017-01 62128270  00-5412-017-02 61777752  00-5412-017-02 61777763  00-5412-017-02 11003153  00-5412-017-02 61782981  00-5412-017-02 61782982  00-5412-017-02 61789829  00-5412-017-02 61796057  00-5412-017-02 61796057R  00-5412-017-02 61789841  00-5412-017-02 61796056  00-5412-017-02 61801509  00-5412-017-02 61796058  00-5412-017-02 61801510  00-5412-017-02 61801511  00-5412-017-02 61810247  00-5412-017-02 11003217  00-5412-017-02 61810283  00-5412-017-02 61806572  00-5412-017-02 61822645  00-5412-017-02 11003238  00-5412-017-02 61810284  00-5412-017-02 61824655  00-5412-017-02 61822646  00-5412-017-02 61824654  00-5412-017-02 61837036  00-5412-017-02 61837037  00-5412-017-02 61862204  00-5412-017-02 61871437  00-5412-017-02 61872042  00-5412-017-02 61871436  00-5412-017-02 61872041  00-5412-017-02 61885249  00-5412-017-02 11004424  00-5412-017-02 61885250  00-5412-017-02 61889377  00-5412-017-02 61876232  00-5412-017-02 61885248  00-5412-017-02 61894444  00-5412-017-02 61894445  00-5412-017-02 61899514  00-5412-017-02 11004761  00-5412-017-02 11004734  00-5412-017-02 61904160  00-5412-017-02 61899552  00-5412-017-02 61867215  00-5412-017-02 61904161  00-5412-017-02 61910568  00-5412-017-02 61916036  00-5412-017-02 61916037  00-5412-017-02 61921368  00-5412-017-02 61921369  00-5412-017-02 11005011  00-5412-017-02 61928683  00-5412-017-02 61933420  00-5412-017-02 61936724  00-5412-017-02 11005310  00-5412-017-02 11005302  00-5412-017-02 61958450  00-5412-017-02 61963814  00-5412-017-02 61963815  00-5412-017-02 61952642  00-5412-017-02 61950173  00-5412-017-02 61940214  00-5412-017-02 61950174  00-5412-017-02 61958449  00-5412-017-02 61968956  00-5412-017-02 61952643  00-5412-017-02 61975670  00-5412-017-02 61975669  00-5412-017-02 61983546  00-5412-017-02 61991993  00-5412-017-02 61991997  00-5412-017-02 61983545  00-5412-017-02 61975668  00-5412-017-02 61968957  00-5412-017-02 61968958  00-5412-017-02 11005889  00-5412-017-02 61983553  00-5412-017-02 61983547  00-5412-017-02 61986591  00-5412-017-02 61988113  00-5412-017-02 61975671  00-5412-017-02 61995024  00-5412-017-02 62001907  00-5412-017-02 61995025  00-5412-017-02 11006045  00-5412-017-02 62001909  00-5412-017-02 62005782  00-5412-017-02 62005783  00-5412-017-02 62005784  00-5412-017-02 62013583  00-5412-017-02 62013584  00-5412-017-02 62013593  00-5412-017-02 62001908  00-5412-017-02 62013582  00-5412-017-02 62020394  00-5412-017-02 62020395  00-5412-017-02 62032718  00-5412-017-02 62025806  00-5412-017-02 62032719  00-5412-017-02 62032720  00-5412-017-02 62025805  00-5412-017-02 62041171  00-5412-017-02 62041172  00-5412-017-02 11006492  00-5412-017-02 11006519  00-5412-017-02 62046179  00-5412-017-02 62046180  00-5412-017-02 62054610  00-5412-017-02 62054611  00-5412-017-02 62056969  00-5412-017-02 62056966  00-5412-017-02 62062134  00-5412-017-02 62062133  00-5412-017-02 62068190  00-5412-017-02 62074913  00-5412-017-02 62074914  00-5412-017-02 62092472  00-5412-017-02 11006758  00-5412-017-02 62086892  00-5412-017-02 62092471  00-5412-017-02 62086891  00-5412-017-02 62104657  00-5412-017-02 62104632  00-5412-017-02 11007001  00-5412-017-02 62092470  00-5412-017-02 62104183  00-5412-017-02 62104183R  00-5412-017-02 62110697  00-5412-017-02 11007081  00-5412-017-02 62117176  00-5412-017-02 62080693  00-5412-017-02 62110698  00-5412-017-02 62122810  00-5412-017-02 62122809  00-5412-017-02 62128271  00-5412-017-02 62132481  00-5412-017-02 62128272  00-5412-018-01 61777753  00-5412-018-01 61782983  00-5412-018-01 61867216  00-5412-018-01 61872043  00-5412-018-01 61889378  00-5412-018-01 61910536  00-5412-018-01 61958451  00-5412-018-01 61958488  00-5412-018-01 61971299  00-5412-018-01 61975673  00-5412-018-01 61975672  00-5412-018-01 61986592  00-5412-018-01 11005767  00-5412-018-01 61988114  00-5412-018-01 61995026  00-5412-018-01 62005785  00-5412-018-01 77001458  00-5412-018-01 77001460  00-5412-018-01 77001459  00-5412-018-01 77001459R    00-5412-018-01 62001910  00-5412-018-01 62001910R  00-5412-018-01 77001462  00-5412-018-01 77001461  00-5412-018-01 77001483  00-5412-018-02 61777754  00-5412-018-02 61782984  00-5412-018-02 11003184  00-5412-018-02 11003184R  00-5412-018-02 11003166  00-5412-018-02 61801512  00-5412-018-02 61806574  00-5412-018-02 61810285  00-5412-018-02 61850060  00-5412-018-02 11004266  00-5412-018-02 61867304  00-5412-018-02 61872044  00-5412-018-02 11004519  00-5412-018-02 61889379  00-5412-018-02 61910537  00-5412-018-02 61928684  00-5412-018-02 61971308  00-5412-018-02 61991998  00-5412-018-02 61983548  00-5412-018-02 11005930  00-5412-018-02 61995027  00-5412-018-02 61988115  00-5412-018-02 62001911  00-5412-018-02 62001911R  00-5412-018-02 62005786  00-5412-018-02 77001454  00-5412-018-02 77001452  00-5412-018-02 77001457  00-5412-018-02 77001456  00-5412-018-02 77001455  00-5412-018-02 77001453  00-5412-018-02 62117177",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "GENDER SOLUTIONS NATURAL-KNEE FLEX SYSTEM FEMORAL COMPONENT¿  N-K FLX GSM POR FEM SZ 2 RT     N-K FLX GSM POR FEM SZ 3 LT     N-K FLX GSM POR FEM SZ 4 LT     N-K FLX GSM POR FEM SZ 4 RT     N-K FLX GSM POR FEM SZ 5 LT     N-K FLX GSM POR FEM SZ 5 RT         Components with CSTi porous coating are indicated for uncemented or cemented use in skeletally mature individuals with intact medial and lateral collateral ligaments undergoing primary surgery for rehabilitating knees damaged as a result of Noninflammatory Degenerative Joint Disease (NIDJD) or Inflammatory Joint Disease (IJD).",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1061-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}