  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82272",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX.    Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.",
      "recall_number": "Z-1060-2019",
      "product_description": "Integrated ECG cable with 3-lead leadwires labeled as the following:      1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED  GRABBER LEAD WIRE, AHA, 3.6 M/12 FT;      2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED  GRABBER LEAD WIRE, IEC, 3.6 M/12 FT    Product Usage:  The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.",
      "product_quantity": "288 devices (23 US, 265 OUS)",
      "reason_for_recall": "The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.",
      "recall_initiation_date": "20190222",
      "center_classification_date": "20190326",
      "termination_date": "20200819",
      "report_date": "20190403",
      "code_info": "1.  Catalog Number 2106309-001, GTIN 00840682137928;      2.  Catalog Number 2106309-002, GTIN 00840682137829"
    }
  ]
}