      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5750-013-01  00-5750-013-02  00-5750-013-06  00-5750-014-01  00-5750-014-02  00-5750-014-05  00-5750-014-06  00-5750-015-01  00-5750-015-02  00-5750-015-05  00-5750-015-06  00-5750-016-01  00-5750-016-02  00-5750-016-05  00-5750-016-06  00-5750-017-01  00-5750-017-02  00-5750-013-01 61951107  00-5750-013-01 61990633  00-5750-013-01 62003770  00-5750-013-01 62088332  00-5750-013-01 62168463  00-5750-013-02 61970680  00-5750-013-02 62004806  00-5750-013-02 62093265  00-5750-013-02 62166656  00-5750-013-06 62120693  00-5750-014-01 61827938  00-5750-014-01 11400140  00-5750-014-01 61844939  00-5750-014-01 61879304  00-5750-014-01 11400196  00-5750-014-01 61905173  00-5750-014-01 61929402  00-5750-014-01 61951674  00-5750-014-01 61964632  00-5750-014-01 61972167  00-5750-014-01 62004808  00-5750-014-01 62029480  00-5750-014-01 62072208  00-5750-014-01 62087759  00-5750-014-01 62093268  00-5750-014-01 62121194  00-5750-014-01 62134274  00-5750-014-01 62159864  00-5750-014-01 62170341  00-5750-014-01 62170342  00-5750-014-01 62170343  00-5750-014-05 62121555  00-5750-014-05 62166655  00-5750-014-02 61791935  00-5750-014-02 61791936  00-5750-014-02 61813666  00-5750-014-02 11400111  00-5750-014-02 61843131  00-5750-014-02 61862167  00-5750-014-02 61870724  00-5750-014-02 61890017  00-5750-014-02 61905264  00-5750-014-02 61914484  00-5750-014-02 61951689  00-5750-014-02 61970681  00-5750-014-02 61978269  00-5750-014-02 61990915  00-5750-014-02 62011560  00-5750-014-02 62022989  00-5750-014-02 62059374  00-5750-014-02 62072514  00-5750-014-02 62088280  00-5750-014-02 62093271  00-5750-014-02 62132082  00-5750-014-02 62134707  00-5750-014-02 62159269  00-5750-014-02 62170344  00-5750-014-02 62170384  00-5750-014-02 62171084  00-5750-014-06 62121229  00-5750-014-06 62135959  00-5750-014-06 62170385  00-5750-015-01 61797750  00-5750-015-01 61804265  00-5750-015-01 61809715  00-5750-015-01 61810409  00-5750-015-01 61819936  00-5750-015-01 61820642  00-5750-015-01 61829061  00-5750-015-01 61834876  00-5750-015-01 61842352  00-5750-015-01 61847666  00-5750-015-01 61863943  00-5750-015-01 61870821  00-5750-015-01 61878293  00-5750-015-01 61889443  00-5750-015-01 61895335  00-5750-015-01 11400226  00-5750-015-01 61902970  00-5750-015-01 61905209  00-5750-015-01 61909358  00-5750-015-01 61908315  00-5750-015-01 61916173  00-5750-015-01 61916174  00-5750-015-01 61921446  00-5750-015-01 61931666  00-5750-015-01 61941604  00-5750-015-01 61952956  00-5750-015-01 61958754  00-5750-015-01 61963888  00-5750-015-01 61966239  00-5750-015-01 61966240  00-5750-015-01 61979149  00-5750-015-01 61986108  00-5750-015-01 61996141  00-5750-015-01 62002081  00-5750-015-01 62002082  00-5750-015-01 62011329  00-5750-015-01 62018053  00-5750-015-01 62018142  00-5750-015-01 62028178  00-5750-015-01 62029873  00-5750-015-01 62029875  00-5750-015-01 62035840  00-5750-015-01 62042331  00-5750-015-01 62049388  00-5750-015-01 62055039  00-5750-015-01 62060477  00-5750-015-01 62064338  00-5750-015-01 62086643  00-5750-015-01 62091506  00-5750-015-01 62091507  00-5750-015-01 62086644  00-5750-015-01 62106697  00-5750-015-01 62116013  00-5750-015-01 62121558  00-5750-015-01 62121556  00-5750-015-01 62132523  00-5750-015-01 62132525  00-5750-015-01 62140119  00-5750-015-01 62148243  00-5750-015-01 62152391  00-5750-015-01 62155453  00-5750-015-01 62170954  00-5750-015-01 62170955  00-5750-015-01 62170956  00-5750-015-01 62170957  00-5750-015-01 62170958  00-5750-015-01 62172733  00-5750-015-01 62172742  00-5750-015-01 62172749  00-5750-015-01 62172737  00-5750-015-01 62172750  00-5750-015-05 61829059  00-5750-015-05 61979151  00-5750-015-05 62003579  00-5750-015-05 62075016  00-5750-015-05 62122660  00-5750-015-05 62133157  00-5750-015-05 62170401  00-5750-015-02 61794716  00-5750-015-02 61810412  00-5750-015-02 61811776  00-5750-015-02 61820645  00-5750-015-02 61824008  00-5750-015-02 61834877  00-5750-015-02 61838758  00-5750-015-02 61847670  00-5750-015-02 61862132  00-5750-015-02 61872331  00-5750-015-02 61876301  00-5750-015-02 61879242  00-5750-015-02 61885147  00-5750-015-02 61886280  00-5750-015-02 61890706  00-5750-015-02 61890043  00-5750-015-02 61904206  00-5750-015-02 61910227  00-5750-015-02 61910229  00-5750-015-02 61910889  00-5750-015-02 61918992  00-5750-015-02 61923590  00-5750-015-02 61936794  00-5750-015-02 61950709  00-5750-015-02 61950710  00-5750-015-02 61960710  00-5750-015-02 61964117  00-5750-015-02 61970682  00-5750-015-02 61973267  00-5750-015-02 61986430  00-5750-015-02 61986739  00-5750-015-02 61996356  00-5750-015-02 61996357  00-5750-015-02 62011576  00-5750-015-02 62011577  00-5750-015-02 62011578  00-5750-015-02 62011579  00-5750-015-02 62022109  00-5750-015-02 62020823  00-5750-015-02 62028437  00-5750-015-02 62033069  00-5750-015-02 62036622  00-5750-015-02 62043322  00-5750-015-02 62050766  00-5750-015-02 62060528  00-5750-015-02 62071255  00-5750-015-02 62075660  00-5750-015-02 62079865  00-5750-015-02 62087941  00-5750-015-02 62092277  00-5750-015-02 62094385  00-5750-015-02 62106698  00-5750-015-02 62107126  00-5750-015-02 62117325  00-5750-015-02 62121774  00-5750-015-02 62132865  00-5750-015-02 62132878  00-5750-015-02 62141732  00-5750-015-02 62141927  00-5750-015-02 62147818  00-5750-015-02 62152394  00-5750-015-02 62156825  00-5750-015-02 62166658  00-5750-015-02 62166748  00-5750-015-02 62171101  00-5750-015-02 62171102  00-5750-015-02 62171757  00-5750-015-02 62171762  00-5750-015-02 62171764  00-5750-015-02 62173170  00-5750-015-06 61797739  00-5750-015-06 61830366  00-5750-015-06 62003582  00-5750-015-06 62091562  00-5750-015-06 62122538  00-5750-015-06 62141936  00-5750-016-01 61794718  00-5750-016-01 61796369  00-5750-016-01 61804735  00-5750-016-01 61823887  00-5750-016-01 61834861  00-5750-016-01 61843134  00-5750-016-01 61879330  00-5750-016-01 61886333  00-5750-016-01 61890709  00-5750-016-01 61905190  00-5750-016-01 61905191  00-5750-016-01 61914487  00-5750-016-01 61929422  00-5750-016-01 61936004  00-5750-016-01 61946957  00-5750-016-01 61955579  00-5750-016-01 61962742  00-5750-016-01 61962744  00-5750-016-01 61976418  00-5750-016-01 61982466  00-5750-016-01 61996959  00-5750-016-01 62005925  00-5750-016-01 62007016  00-5750-016-01 62013511  00-5750-016-01 62022325  00-5750-016-01 62022324  00-5750-016-01 62034567  00-5750-016-01 62044626  00-5750-016-01 62063739  00-5750-016-01 62072219  00-5750-016-01 62080345  00-5750-016-01 62092539  00-5750-016-01 62093019  00-5750-016-01 62112670  00-5750-016-01 62119497  00-5750-016-01 62125812  00-5750-016-01 62137994  00-5750-016-01 62144301  00-5750-016-01 62152969  00-5750-016-01 62155068  00-5750-016-01 62171793  00-5750-016-05 61836353  00-5750-016-05 61970691  00-5750-016-05 11006472  00-5750-016-05 62093021  00-5750-016-05 62100379  00-5750-016-05 62126817  00-5750-016-02 61796371  00-5750-016-02 61798627  00-5750-016-02 61809722  00-5750-016-02 61819934  00-5750-016-02 61823891  00-5750-016-02 61823892  00-5750-016-02 61829840  00-5750-016-02 61844041  00-5750-016-02 61846258  00-5750-016-02 61867611  00-5750-016-02 61876284  00-5750-016-02 11400202  00-5750-016-02 61888046  00-5750-016-02 61893643  00-5750-016-02 61900478  00-5750-016-02 61905269  00-5750-016-02 61905271  00-5750-016-02 61914490  00-5750-016-02 61922259  00-5750-016-02 61936008  00-5750-016-02 61936677  00-5750-016-02 61946960  00-5750-016-02 61959246  00-5750-016-02 61959248  00-5750-016-02 11400318  00-5750-016-02 61968824  00-5750-016-02 61968857  00-5750-016-02 61976430  00-5750-016-02 61988466  00-5750-016-02 61989668  00-5750-016-02 61997351  00-5750-016-02 62005963  00-5750-016-02 62014079  00-5750-016-02 62014080  00-5750-016-02 62022326  00-5750-016-02 62022327  00-5750-016-02 62038034  00-5750-016-02 62046244  00-5750-016-02 62064361  00-5750-016-02 62073119  00-5750-016-02 62080827  00-5750-016-02 62088862  00-5750-016-02 62094391  00-5750-016-02 62106555  00-5750-016-02 62112673  00-5750-016-02 62130983  00-5750-016-02 62138093  00-5750-016-02 62146939  00-5750-016-02 62154050  00-5750-016-02 62163077  00-5750-016-02 62171794  00-5750-016-06 61843136  00-5750-016-06 61982989  00-5750-016-06 62007752  00-5750-016-06 62094390  00-5750-016-06 62138249  00-5750-017-01 61830532  00-5750-017-01 61870826  00-5750-017-01 61889195  00-5750-017-01 11400212  00-5750-017-01 11400228  00-5750-017-01 61943192  00-5750-017-01 61970695  00-5750-017-01 61977555  00-5750-017-01 62003098  00-5750-017-01 62036516  00-5750-017-01 62075677  00-5750-017-01 62134276  00-5750-017-01 62164340  00-5750-017-01 62172421  00-5750-017-05 62134278  00-5750-017-02 61830533  00-5750-017-02 61870831  00-5750-017-02 61890024  00-5750-017-02 61904207  00-5750-017-02 61906387  00-5750-017-02 61926131  00-5750-017-02 61931674  00-5750-017-02 61958023  00-5750-017-02 61963893  00-5750-017-02 61984229  00-5750-017-02 62003103  00-5750-017-02 62036517  00-5750-017-02 62081907  00-5750-017-02 62091452  00-5750-017-02 62146658  00-5750-017-02 62172422  00-5750-017-06 62134277  00-5750-017-06 62173183",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "CR-FLEX GSF PRECOAT SZ C-L¿  CR-FLEX GSF PRECOAT SZ C-R¿  CR-FLEX GSF PRECOAT SZ D-L¿  CR-FLEX GSF PRECOAT SZ D-R¿  CR-FLEX GSF PRECOAT SZ E-L¿  CR-FLEX GSF PRECOAT SZ E-R¿  CR-FLEX GSF PRECOAT SZ F-L¿  CR-FLEX GSF PRECOAT SZ F-R¿  CR-FLEX GSF PRECOAT SZ G-L¿  CR-FLEX GSK PRECOAT SZ G-R¿    \" These devices are indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision of previous arthroplasty procedures. \" These devices are indicated for cemented use only. \" The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1059-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}