  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70138",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355-1406",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including Canada and Mexico.",
      "recall_number": "Z-1059-2015",
      "product_description": "LINAC Accelerators: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI,      \t  Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2,   MEVATRON KDS-2, MEVATRON KDS-2, MEVATRON KD-2, MEVATRON K2,   MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, \t MEVATRON M2 6300, MEVATRON MDX-2, MEVATRON MD-2; The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.\t      Product Usage:  The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.",
      "product_quantity": "310",
      "reason_for_recall": "A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.",
      "recall_initiation_date": "20141219",
      "center_classification_date": "20150205",
      "termination_date": "20170126",
      "report_date": "20150211",
      "code_info": "model numbers: 08139789, 05863472, 05857912, 07360717, 05857920, 04504200, 01940035, 05500371,   09822685, 09411588, 09822693, 08515520, 01940753, 01924500, 09401506,  09401407, 09401316, 09401746, 09401654, with serial numbers: 3397  3986  5087  70-4119  70-4144  70-4262  70-4343  4078  5120  5170  5282  5506  70-4146  5527  2958  3855  5517  3753  3702  3751  2863  3695  5364  3904  70-4287  70-4175  3102  3781  3203  5060  3798  70-4316  5205  3383  5688  5743  3832  2972  3942  70-4122  4067  70-4283  5190  5157  2646  5343  5844  3953  5148  5391  3931  5300  3802  70-4172  70-4143  70-4296  3079  2516  70-4235  2962  70-4257  5474  2834  70-4363  70-4368  3791  3793  3738  3817  3055  5833  5899  2586  5207  3458  3673  70-4378  5355  3515  70-4159  3524  3574  3519  5281  5894  5374  5558  3225  3631  3835  3367  3296  2702  3591  3267  2912  3905  3922  3924  2896  2814  3773  3679  3533  3709  5367  3618  70-4354  70-4356  5540  3103  5488  3859  3009  4019  5523  5590  5657  3624  2887  5350  5784  2974  5821  3503  70-4344  3285  3979  3541  3413  3774  2996  2125  5843  3504  3318  2628  70-4129  5851  3354  70-4183  3888  3066  3361  5098  3756  5385  2689  3305  2228  3604  3629  3136  5114  3748  3788  5700  2183  5410  2644  70-4323  3666  3825  2507  3241  70-4207  3987  4036  3989  5865  70-4166  3617  3736  5118  5548  3075  3769  2845  3842  5861  70-4188  5748  5222  3013  3484  3302  3068  70-4317  5737  5830  2596  5656  5245  5092  3303  3778  5088  4021  3245  5665  3815  3133  5823  3846  70-4178  2532  3907  70-4174  3534  5353  3826  3025  5630  70-4185  3639  3274  3664  2527  3761  5096  3801  3949  3878  3049  2894  5572  3699  3873  5417  4059  3805  5500  2965  5388  3547  2045  3116  3678  3186  70-4362  70-4379  5226  3623  3804  3453  3602  70-4142  2827  3657  3948  70-4135  3755  3754  5601  5640  3153  3622  5033  3514  5177  3084  3731  2855  4065  3147  5707  5398  4091  3644  4018  2906  5532  70-4306  70-4161  3114  3667  5424  3565  5755  3711  5326  5525  5145  3016  5699  4079  4082  70-4113  4015  3981  5419  3998  3975  5093  3479  5481  5396  70-4347  5008  2067  2818  3939  3089  5095  3901  5153  3244  5179  2244  5154"
    }
  ]
}