  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67261",
      "recalling_firm": "Synthes, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA (nationwide) Distribution.",
      "recall_number": "Z-1057-2014",
      "product_description": "Synthes Driving Cap/Threaded    Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.",
      "product_quantity": "143",
      "reason_for_recall": "The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.",
      "recall_initiation_date": "20140107",
      "center_classification_date": "20140220",
      "termination_date": "20160818",
      "report_date": "20140226",
      "code_info": "Part No. 03.010.523, Lot Nos. 8279162 & 8429004",
      "more_code_info": ""
    }
  ]
}