{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5764-013-51 61874481  00-5764-013-51 61892716  00-5764-013-51 61946510  00-5764-013-51 61954552  00-5764-013-51 61977566  00-5764-013-51 62003780  00-5764-013-51 62022335  00-5764-013-51 62029870  00-5764-013-51 62107937  00-5764-013-51 62121782  00-5764-013-51 62155687  00-5764-013-51 62159335  00-5764-013-52 61800630  00-5764-013-52 61808793  00-5764-013-52 61819071  00-5764-013-52 61885786  00-5764-013-52 61892717  00-5764-013-52 61914463  00-5764-013-52 61949724  00-5764-013-52 61954981  00-5764-013-52 61977567  00-5764-013-52 62003781  00-5764-013-52 62023325  00-5764-013-52 62030453  00-5764-013-52 62096017  00-5764-013-52 62102077  00-5764-013-52 62127093  00-5764-013-52 62137523  00-5764-013-52 62168360  00-5764-013-52 62169039  00-5764-013-52 62172720  00-5764-013-52 62172721  00-5764-013-52 62173154  00-5764-014-51 61793522  00-5764-014-51 61796450  00-5764-014-51 61802094  00-5764-014-51 61816554  00-5764-014-51 61817772  00-5764-014-51 61817773  00-5764-014-51 11400112  00-5764-014-51 61787880  00-5764-014-51 61829076  00-5764-014-51 61829078  00-5764-014-51 61834863  00-5764-014-51 61835849  00-5764-014-51 61842366  00-5764-014-51 61858497  00-5764-014-51 61868886  00-5764-014-51 61868887  00-5764-014-51 61881032  00-5764-014-51 61881034  00-5764-014-51 61885761  00-5764-014-51 61888051  00-5764-014-51 61892383  00-5764-014-51 61896059  00-5764-014-51 61896058  00-5764-014-51 61897301  00-5764-014-51 61902968  00-5764-014-51 61910188  00-5764-014-51 61914468  00-5764-014-51 61918704  00-5764-014-51 61918705  00-5764-014-51 61924472  00-5764-014-51 61926153  00-5764-014-51 61936817  00-5764-014-51 61943203  00-5764-014-51 61946974  00-5764-014-51 61958271  00-5764-014-51 61958748  00-5764-014-51 61965573  00-5764-014-51 61965574  00-5764-014-51 61967192  00-5764-014-51 61967194  00-5764-014-51 61974066  00-5764-014-51 61974068  00-5764-014-51 61979793  00-5764-014-51 61984251  00-5764-014-51 61986228  00-5764-014-51 61985016  00-5764-014-51 61986231  00-5764-014-51 62017272  00-5764-014-51 62023806  00-5764-014-51 62023333  00-5764-014-51 62030266  00-5764-014-51 62029858  00-5764-014-51 62030268  00-5764-014-51 62030269  00-5764-014-51 62030578  00-5764-014-51 62038009  00-5764-014-51 62038007  00-5764-014-51 62038008  00-5764-014-51 62039076  00-5764-014-51 62049410  00-5764-014-51 62049412  00-5764-014-51 62071907  00-5764-014-51 62076334  00-5764-014-51 62084449  00-5764-014-51 62085549  00-5764-014-51 62096020  00-5764-014-51 62096024  00-5764-014-51 62100483  00-5764-014-51 62100482  00-5764-014-51 62101706  00-5764-014-51 62106707  00-5764-014-51 62107956  00-5764-014-51 62110811  00-5764-014-51 62119558  00-5764-014-51 62119557  00-5764-014-51 62120493  00-5764-014-51 62131261  00-5764-014-51 62132092  00-5764-014-51 62143628  00-5764-014-51 62151117  00-5764-014-51 62151121  00-5764-014-51 62151252  00-5764-014-51 62152410  00-5764-014-51 62152412  00-5764-014-51 62162780  00-5764-014-51 62169040  00-5764-014-51 62169058  00-5764-014-51 62169061  00-5764-014-51 62169059  00-5764-014-51 62169060  00-5764-014-51 62170082  00-5764-014-51 62170076  00-5764-014-51 62170079  00-5764-014-51 62170080  00-5764-014-51 62173155  00-5764-014-51 62173156  00-5764-014-51 62173167  00-5764-014-51 62173168  00-5764-014-51 62173194  00-5764-014-51 62173193  00-5764-014-52 61796361  00-5764-014-52 61804253  00-5764-014-52 61803334  00-5764-014-52 61809580  00-5764-014-52 61816405  00-5764-014-52 61819068  00-5764-014-52 61821435  00-5764-014-52 61822979  00-5764-014-52 61832458  00-5764-014-52 61833619  00-5764-014-52 61837343  00-5764-014-52 61837344  00-5764-014-52 61842369  00-5764-014-52 61854802  00-5764-014-52 61859250  00-5764-014-52 61868889  00-5764-014-52 61868890  00-5764-014-52 61869869  00-5764-014-52 61874487  00-5764-014-52 61881657  00-5764-014-52 61881659  00-5764-014-52 61881660  00-5764-014-52 61892387  00-5764-014-52 61893247  00-5764-014-52 61894545  00-5764-014-52 61899648  00-5764-014-52 61901228  00-5764-014-52 61902966  00-5764-014-52 61907467  00-5764-014-52 61915329  00-5764-014-52 61921477  00-5764-014-52 61926156  00-5764-014-52 61932311  00-5764-014-52 61946459  00-5764-014-52 61956807  00-5764-014-52 61956808  00-5764-014-52 61958269  00-5764-014-52 61965505  00-5764-014-52 61964453  00-5764-014-52 61972316  00-5764-014-52 61972317  00-5764-014-52 61972754  00-5764-014-52 61976340  00-5764-014-52 61976342  00-5764-014-52 61980592  00-5764-014-52 61984199  00-5764-014-52 61992067  00-5764-014-52 61992069  00-5764-014-52 62018156  00-5764-014-52 62023807  00-5764-014-52 62019511  00-5764-014-52 62026104  00-5764-014-52 62024586  00-5764-014-52 62030580  00-5764-014-52 62032079  00-5764-014-52 62039173  00-5764-014-52 62039171  00-5764-014-52 62040748  00-5764-014-52 62046929  00-5764-014-52 62047537  00-5764-014-52 62048449  00-5764-014-52 62063766  00-5764-014-52 62064493  00-5764-014-52 62069351  00-5764-014-52 62077281  00-5764-014-52 62077309  00-5764-014-52 62085550  00-5764-014-52 62085552  00-5764-014-52 62098784  00-5764-014-52 62098834  00-5764-014-52 62099264  00-5764-014-52 62099786  00-5764-014-52 62102049  00-5764-014-52 62107957  00-5764-014-52 62107995  00-5764-014-52 62107999  00-5764-014-52 62114385  00-5764-014-52 62115960  00-5764-014-52 62116005  00-5764-014-52 62121784  00-5764-014-52 62132886  00-5764-014-52 62132887  00-5764-014-52 62133207  00-5764-014-52 62133209  00-5764-014-52 62146949  00-5764-014-52 62154048  00-5764-014-52 62160342  00-5764-014-52 62160892  00-5764-014-52 62165941  00-5764-014-52 62170952  00-5764-014-52 62170953  00-5764-014-52 62170397  00-5764-014-52 62170951  00-5764-014-52 62171097  00-5764-014-52 62171098  00-5764-014-52 62171099  00-5764-014-52 62171584  00-5764-014-52 62170948  00-5764-014-52 62171096  00-5764-014-52 62174422  00-5764-015-51 61779535  00-5764-015-51 61793548  00-5764-015-51 61778270  00-5764-015-51 61793550  00-5764-015-51 61780342  00-5764-015-51 61817771  00-5764-015-51 61819065  00-5764-015-51 61779538  00-5764-015-51 61780343  00-5764-015-51 61780344  00-5764-015-51 61819067  00-5764-015-51 61821147  00-5764-015-51 61827947  00-5764-015-51 61827948  00-5764-015-51 61827949  00-5764-015-51 61829861  00-5764-015-51 61829859  00-5764-015-51 61837319  00-5764-015-51 61839696  00-5764-015-51 61839697  00-5764-015-51 61819232  00-5764-015-51 61839701  00-5764-015-51 61846271  00-5764-015-51 61846295  00-5764-015-51 11400127  00-5764-015-51 61855939  00-5764-015-51 61856755  00-5764-015-51 11400159  00-5764-015-51 61866732  00-5764-015-51 61866733  00-5764-015-51 61872345  00-5764-015-51 61876931  00-5764-015-51 61880113  00-5764-015-51 61880112  00-5764-015-51 61881692  00-5764-015-51 61883280  00-5764-015-51 61883281  00-5764-015-51 61883283  00-5764-015-51 61884198  00-5764-015-51 61885186  00-5764-015-51 61885185  00-5764-015-51 61902963  00-5764-015-51 61902964  00-5764-015-51 61904203  00-5764-015-51 61906073  00-5764-015-51 61908309  00-5764-015-51 61916101  00-5764-015-51 61915325  00-5764-015-51 61915327  00-5764-015-51 61915328  00-5764-015-51 61918735  00-5764-015-51 61924478  00-5764-015-51 61926148  00-5764-015-51 61933846  00-5764-015-51 61933818  00-5764-015-51 61949719  00-5764-015-51 61950809  00-5764-015-51 61952891  00-5764-015-51 61953109  00-5764-015-51 61958030  00-5764-015-51 61958031  00-5764-015-51 61958327  00-5764-015-51 61958329  00-5764-015-51 61963907  00-5764-015-51 61963908  00-5764-015-51 61964178  00-5764-015-51 61971181  00-5764-015-51 61971133  00-5764-015-51 61972194  00-5764-015-51 61972190  00-5764-015-51 11400326  00-5764-015-51 61978305  00-5764-015-51 61978545  00-5764-015-51 61984216  00-5764-015-51 61985033  00-5764-015-51 61986426  00-5764-015-51 61986732  00-5764-015-51 61986425  00-5764-015-51 61986731  00-5764-015-51 61990932  00-5764-015-51 62014088  00-5764-015-51 62013572  00-5764-015-51 62015860  00-5764-015-51 62015863  00-5764-015-51 62023803  00-5764-015-51 62023804  00-5764-015-51 62024965  00-5764-015-51 62026103  00-5764-015-51 62026945  00-5764-015-51 62030613  00-5764-015-51 62030641  00-5764-015-51 62039186  00-5764-015-51 62039946  00-5764-015-51 62040768  00-5764-015-51 62044625  00-5764-015-51 62050091  00-5764-015-51 62051206  00-5764-015-51 62051061  00-5764-015-51 62051207  00-5764-015-51 62052132  00-5764-015-51 62064405  00-5764-015-51 62068557  00-5764-015-51 62072217  00-5764-015-51 62073121  00-5764-015-51 62080805  00-5764-015-51 62084279  00-5764-015-51 62084447  00-5764-015-51 62092535  00-5764-015-51 62096018  00-5764-015-51 62096629  00-5764-015-51 62099061  00-5764-015-51 62099278  00-5764-015-51 62099947  00-5764-015-51 62099950  00-5764-015-51 11006800  00-5764-015-51 62103221  00-5764-015-51 62109175  00-5764-015-51 62107993  00-5764-015-51 62107994  00-5764-015-51 62114466  00-5764-015-51 62117288  00-5764-015-51 62117289  00-5764-015-51 62121804  00-5764-015-51 62122578  00-5764-015-51 62127094  00-5764-015-51 62128369  00-5764-015-51 62127493  00-5764-015-51 62128382  00-5764-015-51 62132116  00-5764-015-51 62132118  00-5764-015-51 62132869  00-5764-015-51 62146907  00-5764-015-51 62146914  00-5764-015-51 62152632  00-5764-015-51 62152649  00-5764-015-51 62162211  00-5764-015-51 62160897  00-5764-015-51 62169025  00-5764-015-51 62169026  00-5764-015-51 62169027  00-5764-015-51 62169028  00-5764-015-51 62169064  00-5764-015-51 62169065  00-5764-015-51 62170094  00-5764-015-51 62170097  00-5764-015-51 62170102  00-5764-015-51 62170104  00-5764-015-51 62170402  00-5764-015-51 62170403  00-5764-015-51 62170404  00-5764-015-51 62171080  00-5764-015-51 62173192  00-5764-015-51 62174388  00-5764-015-51 62174389  00-5764-015-51 62174418  00-5764-015-51 62174419  00-5764-015-51 62174420  00-5764-015-52 61794729  00-5764-015-52 61794731  00-5764-015-52 61785877  00-5764-015-52 61785878  00-5764-015-52 61806293  00-5764-015-52 61806294  00-5764-015-52 61780360  00-5764-015-52 61782061  00-5764-015-52 61780358  00-5764-015-52 61780359  00-5764-015-52 61781294  00-5764-015-52 61814773  00-5764-015-52 61782059  00-5764-015-52 61817767  00-5764-015-52 61817768  00-5764-015-52 61780357  00-5764-015-52 61782058  00-5764-015-52 61782060  00-5764-015-52 61820658  00-5764-015-52 61820669  00-5764-015-52 61820732  00-5764-015-52 61821146  00-5764-015-52 61822977  00-5764-015-52 61824793  00-5764-015-52 61832298  00-5764-015-52 61832299  00-5764-015-52 61833582  00-5764-015-52 61837336  00-5764-015-52 61819248  00-5764-015-52 61819250  00-5764-015-52 61819270  00-5764-015-52 61845036  00-5764-015-52 11400124  00-5764-015-52 61853393  00-5764-015-52 61853394  00-5764-015-52 61854830  00-5764-015-52 11400123  00-5764-015-52 61866727  00-5764-015-52 61871541  00-5764-015-52 61871542  00-5764-015-52 61876293  00-5764-015-52 61876294  00-5764-015-52 61876929  00-5764-015-52 61878281  00-5764-015-52 61878285  00-5764-015-52 61878286  00-5764-015-52 61876295  00-5764-015-52 61880111  00-5764-015-52 61879238  00-5764-015-52 61881691  00-5764-015-52 61885151  00-5764-015-52 61889072  00-5764-015-52 61889073  00-5764-015-52 61896056  00-5764-015-52 61897298  00-5764-015-52 61900472  00-5764-015-52 61900474  00-5764-015-52 61905204  00-5764-015-52 61900475  00-5764-015-52 61904939  00-5764-015-52 61910884  00-5764-015-52 61916102  00-5764-015-52 61916167  00-5764-015-52 61917767  00-5764-015-52 61916165  00-5764-015-52 61916166  00-5764-015-52 61920457  00-5764-015-52 61920458  00-5764-015-52 61927979  00-5764-015-52 61927981  00-5764-015-52 61929419  00-5764-015-52 61932307  00-5764-015-52 61935999  00-5764-015-52 61936001  00-5764-015-52 61943100  00-5764-015-52 61947025  00-5764-015-52 61949720  00-5764-015-52 61952892  00-5764-015-52 61954534  00-5764-015-52 61954551  00-5764-015-52 61958765  00-5764-015-52 61960727  00-5764-015-52 61960730  00-5764-015-52 11400323  00-5764-015-52 61965494  00-5764-015-52 61965498  00-5764-015-52 61971185  00-5764-015-52 61971734  00-5764-015-52 61971735  00-5764-015-52 61971736  00-5764-015-52 61976427  00-5764-015-52 61977573  00-5764-015-52 11400331  00-5764-015-52 61980327  00-5764-015-52 61980331  00-5764-015-52 61980588  00-5764-015-52 61980589  00-5764-015-52 61987069  00-5764-015-52 61988203  00-5764-015-52 61987068  00-5764-015-52 62003783  00-5764-015-52 62005010  00-5764-015-52 62013660  00-5764-015-52 62014093  00-5764-015-52 62017638  00-5764-015-52 62017639  00-5764-015-52 62018580  00-5764-015-52 62027210  00-5764-015-52 62028434  00-5764-015-52 62028459  00-5764-015-52 62028460  00-5764-015-52 62033741  00-5764-015-52 62033742  00-5764-015-52 62036666  00-5764-015-52 62038027  00-5764-015-52 62039089  00-5764-015-52 62040769  00-5764-015-52 62040770  00-5764-015-52 62042055  00-5764-015-52 62042058  00-5764-015-52 62051073  00-5764-015-52 62052008  00-5764-015-52 62052684  00-5764-015-52 11400407  00-5764-015-52 62055603  00-5764-015-52 62064397  00-5764-015-52 62063754  00-5764-015-52 62074510  00-5764-015-52 62077305  00-5764-015-52 62077869  00-5764-015-52 62085543  00-5764-015-52 62085555  00-5764-015-52 62085958  00-5764-015-52 62094393  00-5764-015-52 62096027  00-5764-015-52 62097833  00-5764-015-52 62097834  00-5764-015-52 62098073  00-5764-015-52 62098802  00-5764-015-52 62098847  00-5764-015-52 62098848  00-5764-015-52 62110241  00-5764-015-52 62110820  00-5764-015-52 62110822  00-5764-015-52 62111573  00-5764-015-52 62112040  00-5764-015-52 62119555  00-5764-015-52 62119556  00-5764-015-52 62123093  00-5764-015-52 62123102  00-5764-015-52 62123109  00-5764-015-52 62129369  00-5764-015-52 62129497  00-5764-015-52 62129499  00-5764-015-52 62129501  00-5764-015-52 62133835  00-5764-015-52 62143321  00-5764-015-52 62142625  00-5764-015-52 62147830  00-5764-015-52 62148253  00-5764-015-52 62148806  00-5764-015-52 62149021  00-5764-015-52 62152332  00-5764-015-52 62152339  00-5764-015-52 62160344  00-5764-015-52 62160908  00-5764-015-52 62163115  00-5764-015-52 62163090  00-5764-015-52 62163092  00-5764-015-52 62164681  00-5764-015-52 62171082  00-5764-015-52 62171103  00-5764-015-52 62171104  00-5764-015-52 62171105  00-5764-015-52 62171106  00-5764-015-52 62171795  00-5764-015-52 62171797  00-5764-015-52 62171798  00-5764-015-52 62171852  00-5764-015-52 62171796  00-5764-015-52 62171851  00-5764-015-52 62172424  00-5764-015-52 62172425  00-5764-015-52 62171849  00-5764-015-52 62173197  00-5764-015-52 62173198  00-5764-015-52 62174394  00-5764-015-52 62174395  00-5764-016-51 61792793  00-5764-016-51 61792794  00-5764-016-51 61804255  00-5764-016-51 61806271  00-5764-016-51 61814754  00-5764-016-51 61821411  00-5764-016-51 61829141  00-5764-016-51 61832279  00-5764-016-51 61835848  00-5764-016-51 61855948  00-5764-016-51 61861376  00-5764-016-51 61866734  00-5764-016-51 61880092  00-5764-016-51 61884397  00-5764-016-51 61889176  00-5764-016-51 61889178  00-5764-016-51 61906078  00-5764-016-51 61906079  00-5764-016-51 61912350  00-5764-016-51 61915332  00-5764-016-51 61926158  00-5764-016-51 61933850  00-5764-016-51 61948147  00-5764-016-51 61956821  00-5764-016-51 61958019  00-5764-016-51 61971189  00-5764-016-51 61971190  00-5764-016-51 61977411  00-5764-016-51 61982477  00-5764-016-51 61982478  00-5764-016-51 61990937  00-5764-016-51 61995694  00-5764-016-51 61995696  00-5764-016-51 61997354  00-5764-016-51 62005967  00-5764-016-51 62021945  00-5764-016-51 62030615  00-5764-016-51 62030617  00-5764-016-51 62039992  00-5764-016-51 62050760  00-5764-016-51 62063780  00-5764-016-51 62077021  00-5764-016-51 62097837  00-5764-016-51 62097838  00-5764-016-51 62111956  00-5764-016-51 62112984  00-5764-016-51 62118647  00-5764-016-51 62129502  00-5764-016-51 62129823  00-5764-016-51 62133843  00-5764-016-51 62141734  00-5764-016-51 62154079  00-5764-016-51 62154053  00-5764-016-51 62157969  00-5764-016-51 62170085  00-5764-016-51 62169062  00-5764-016-51 62169063  00-5764-016-51 62170087  00-5764-016-52 61793531  00-5764-016-52 61794710  00-5764-016-52 61798624  00-5764-016-52 61807823  00-5764-016-52 61816385  00-5764-016-52 61816386  00-5764-016-52 61817749  00-5764-016-52 61819049  00-5764-016-52 61832270  00-5764-016-52 61834999  00-5764-016-52 61835000  00-5764-016-52 61857910  00-5764-016-52 61857909  00-5764-016-52 61863969  00-5764-016-52 61878280  00-5764-016-52 61879235  00-5764-016-52 61880390  00-5764-016-52 61890013  00-5764-016-52 61900476  00-5764-016-52 61900477  00-5764-016-52 61906080  00-5764-016-52 61907464  00-5764-016-52 61915334  00-5764-016-52 61916171  00-5764-016-52 61927986  00-5764-016-52 61933852  00-5764-016-52 61939198  00-5764-016-52 61949725  00-5764-016-52 61954907  00-5764-016-52 61955570  00-5764-016-52 61960732  00-5764-016-52 61964191  00-5764-016-52 61964450  00-5764-016-52 61964451  00-5764-016-52 61977415  00-5764-016-52 61977420  00-5764-016-52 61977569  00-5764-016-52 61984192  00-5764-016-52 61990934  00-5764-016-52 61990935  00-5764-016-52 61997092  00-5764-016-52 61997655  00-5764-016-52 62005011  00-5764-016-52 62022987  00-5764-016-52 62030644  00-5764-016-52 62036758  00-5764-016-52 62040775  00-5764-016-52 62052777  00-5764-016-52 11400420  00-5764-016-52 62072221  00-5764-016-52 62077311  00-5764-016-52 62085964  00-5764-016-52 62098850  00-5764-016-52 62098851  00-5764-016-52 62110243  00-5764-016-52 62111640  00-5764-016-52 62122673  00-5764-016-52 62127562  00-5764-016-52 62134718  00-5764-016-52 62140669  00-5764-016-52 62149048  00-5764-016-52 62154294  00-5764-016-52 62156828  00-5764-016-52 62163088  00-5764-016-52 62168669  00-5764-016-52 62168677  00-5764-016-52 62168688  00-5764-016-52 62168693  00-5764-017-51 61807828  00-5764-017-51 61835350  00-5764-017-51 61907463  00-5764-017-51 61944906  00-5764-017-51 61967907  00-5764-017-51 61996353  00-5764-017-51 62034482  00-5764-017-51 62087938  00-5764-017-51 62126104  00-5764-017-51 62136387  00-5764-017-51 62117139  00-5764-017-51 62171585  00-5764-017-52 61806276  00-5764-017-52 61835845  00-5764-017-52 61907462  00-5764-017-52 61917771  00-5764-017-52 61951114  00-5764-017-52 61967909  00-5764-017-52 61982476  00-5764-017-52 61996355  00-5764-017-52 62067956  00-5764-017-52 62090171  00-5764-017-52 62136388  00-5764-017-52 62156850  00-5764-017-52 62171591",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "LPS-FLEX GSF OPT SZ C-L¿  LPS-FLEX GSF OPT SZ C-R¿  LPS-FLEX GSF OPT SZ D-L¿  LPS-FLEX GSF OPT SZ D-R¿  LPS-FLEX GSF OPT SZ E-L¿  LPS-FLEX GSF OPT SZ E-R¿  LPS-FLEX GSK OPT SZ F-L¿  LPS-FLEX GSF OPT SZ F-R¿  LPS-FLEX GSF OPT SZ G-L¿  LPS-FLEX GSF OPT SZ G-R¿",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1056-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}