  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76043",
      "recalling_firm": "The Binding Site Group, Ltd.",
      "address_1": "8 Calthorpe Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. distribution to the following; MA, GA, NY, IA, CA, IL, UT, NJ, NC, MD, MN.    Foreign distribution to the following; UK, Germany, Spain, France.",
      "recall_number": "Z-1055-2017",
      "product_description": "Digital RID Plate Reader and Software    Product Code: AD400",
      "product_quantity": "",
      "reason_for_recall": "If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.",
      "recall_initiation_date": "20121009",
      "center_classification_date": "20170119",
      "termination_date": "20170404",
      "report_date": "20170125",
      "code_info": "UDI 05051700000376",
      "more_code_info": ""
    }
  ]
}