{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82313",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide domestic distribution, and worldwide international distribution.",
      "recall_number": "Z-1054-2019",
      "product_description": "Omnidiagnost Classic, Product Codes 70859, 708023, 708024, 708025    Product Usage:  The Philips OmniDiagnost is a multi-functional, universal X-ray system with remote-controlled facilities. It performs a wide range of routine examinations and interventional procedures include:   Gastro-intestinal examination   Urography   Bronchography   Arthrography   Hysterography   Myelography   Lymphography   Skeletal radiography   Tomography   Venography/phlebography Special procedures that require subtracted and non-subtracted digital imaginginclude:   Angiography   Peripheral angiography (bolus chasing) Vascular and non-vascular interventional procedures.",
      "product_quantity": "1696",
      "reason_for_recall": "The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.",
      "recall_initiation_date": "20190228",
      "center_classification_date": "20190323",
      "termination_date": "20200803",
      "report_date": "20190403",
      "code_info": "Serial numbers 1  2  3  4  7  9  10  11  12  13  14  15  16  17  18  19  20  22  23  24  25  26  27  28  29  30  31  32  33  34  36  38  39  41  42  44  46  47  48  49  50  51  52  53  54  56  58  60  61  62  64  65  66  70  71  72  73  74  75  76  77  78  79  80  81  82  83  84  85  86  87  88  89  90  91  92  94  95  96  97  98  101  102  103  104  104  106  107  108  109  110  111  111  112  113  115  116  117  118  118  119  120  121  123  123  124  126  127  128  129  130  131  132  133  134  135  135  136  137  138  139  140  142  143  145  146  146  147  148  149  150  151  152  153  154  156  157  158  159  161  164  165  167  173  183  195  198  199  216  217  227  251  253  259  274  281  286  297  317  323  327  333  336  340  348  353  353  358  369  378  380  390  391  393  402  421  423  426  428  438  441  466  471  473  474  481  486  498  504  514  520  530  570  574  592  598  608  664  1489  1529  3019  13124  38351  50428  76596  76948  82655  83871  83958  85163  86113  86303  86326  9341573  41355410  4163250111  708023019087  970478028909  _000364  _073  _082  _146  _302  _332  _345  _364  _371  ~~~~~~~~~~1  ~~~1  096B  1  1*015  112  11591  12  122  1232  126%  131  13310  136  14*009-Omni  141%  163  175  175  178  182  185  190  201  202  210  221  224  225  237  248  26  266  268  275  280  282  284  294  306  319  324  3243  326  3278012607  339  355  365  370  3797350-9  381  389  392  3A57204  401  4042910/000134  407  411  413554106  416  4188010/0171  419  4349690/000217  436  447  463  4665620/000374  4665620/355  4700180000326  4831320/480  60  609  66  68  708023015453  708023015995  708023018193  7080231896  70802322  7080235106  710376025944  78  8  83  84850  86549  86965  86977  87%  87164  89  90  96  97  98  99  C@20236  C@20238  C@20239  C@20241  C@20286  C@21319  C@21325  C@21338  C@21339  C@21353  C@21388  C@21389  C@21390  C@21398  C@21402  C@21406  C@21414  C@22528  C@22599  C@226  C@98168  C@99218  C@99224  GEN023210  GEN023233  M16331  M17264  M19163  M19164  M19169  M19269  M20825  M21186  M22124  S100000980/000001  S100002676/000001  S100002705/000001  S103815/1  S1048831  S3F000861A/117337  S3F000862A/117338  SN98170  URF.092  URF.130  URF.186  URF.218"
    }
  ]
}