  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82313",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide domestic distribution, and worldwide international distribution.",
      "recall_number": "Z-1053-2019",
      "product_description": "Urodiagnost, Product No. 708033    Product Usage:  The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including:   Urograms with injection or infusion   Retrograde pyelography   Cystography and urethrography   Micturition cystography   Cystoscopy, endopscopy   Transurethral resection   Percutaneous nephroscopy   Urethrorenoscopy   Percutaneous nephrolithotomy   Serial angiography with image intensifier and DSI   Tomography",
      "product_quantity": "1969",
      "reason_for_recall": "The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.",
      "recall_initiation_date": "20190228",
      "center_classification_date": "20190323",
      "termination_date": "20200803",
      "report_date": "20190403",
      "code_info": "Serial codes 1  2  3  4  5  7  8  11  13  14  15  16  17  18  21  48  59  60  64"
    }
  ]
}