  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73248",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including all states in including PR except ND, VT & WY.  Foreign Distribution:  ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, BURKINA FASO, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE,HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRA N, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA  .",
      "recall_number": "Z-1053-2016",
      "product_description": "LightSpeed QX/I (4-slice MDAS)    LightSpeed 4.0 CT Scanner System - The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. PET Discovery ST system - The GE Discovery ST system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging system. LightSpeed Pro16 - The Lightspeed 5.0 CT Scanner System is indicated for head and whole body X-ray computed tomography applications. 16 Slice BGO PET/CT System - The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems. 6 Slice BGO PET/CT System - The GE Discovery LS System is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery LS is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery LS can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.",
      "product_quantity": "104",
      "reason_for_recall": "GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.",
      "recall_initiation_date": "20160210",
      "center_classification_date": "20160304",
      "termination_date": "20160921",
      "report_date": "20160316",
      "code_info": "Mfg Lot or Serial # System ID  00000006841PT1 PET368725CN8  00000004411PT5 4512-1  00000004411PT5 4512-1  00000006841PT1 PET368725CN8  00000005376PT9 PET357769CN9  00000005670PT5 082426140014  00000006610PT0 082445020003  00000006610PT0 082445020003  00000006610PT0 082445020003  00000006119PT2 082445030003  00000005815PT6 082445050003  00000006146PT5 082445090002  00000005808PT1 082445100008  00000006177PT0 082445100009  00000006175PT4 082445130004  00000005670PT5 082426140014  00000005808PT1 082445100008  00000006500PT3 082445160007  00000006242PT2 082445190002  00000006200PT0 082445250002  00000005251PT4 H4858964  00000005121PT9 M4143813  00000005352PT0 M4144860  00000005323PT1 X105967801  00000006194PT5 HU1167PT01  00000006762PT9 0004168403  00000006762PT9 0004168403  00000005809PT9 IE1005PT01  00000004588PT0 IL1062NM18  00000006120PT0 A5107746  UNK51012720002 A5107746  UNK51012720001 A5215931  00000006003PT8 A5333025  00000004619PT3 A99263301  00000006292PT7 YD0024  00000006293PT5 YD0025  00000006039PT2 YE0001  00000006273PT7 YE0002  00000006272PT9 YE0003  00000006345PT3 YE0004  00000006346PT1 YE0005  00000006808PT0 YE0006  00000006488PT1 YE0007  00000006403PT0 YE0008  00000006368PT5 YE0011  00000006274PT5 YE0077  00000005920PT4 YE0101  00000006141PT6 YE0102  00000006076PT4 YE0103  00000006404PT8 YE0104  00000006496PT4 YE0105  00000006405PT5 YE0106  00000006453PT5 YE0107  00000006636PT5 YE0108  00000006331PT3 YE0110  00000006347PT9 YE0111  00000006454PT3 YE0112  00000006115PT0 YE0201  00000006407PT1 YE0202  00000006611PT8 YE0203  00000006245PT5 YE0205  00000006810PT6 YE0210  00000006843PT7 YE0211  00000005813PT1 YE0214  00000006367PT7 YE0109  00000006304PT0 YE0204  00000005813PT1 YE0214  00000004967PT6 0850450042  00000004967PT6 0850450042  00000004967PT6 0850450042  00000004967PT6 0850450042  00000004463PT6 0850450055  00000005495PT7 0850450056X1  00000005342PT1 0850450057  00000005342PT1 0850450057  00000005878PT4 0850450058  00000006176PT2 0850450059  00000006176PT2 0850450059  00000005785PT1 0834450001  00000005785PT1 0834450001  00000005719PT0 2535997DST  00000005719PT0 2535997DST  00000006486PT5 2546945DST  00000006148PT1 FE816135  UNK51012720004 FE816135  00000004732PT4 FEG81300  00000006302PT4 NP739714  UNK51012720000 Not Available  00000006201PT8 C001PE03  00000006201PT8 C001PE03  00000006294PT3 259130PE01  00000006294PT3 259130PE01  00000005184PT7 0853450020  00000004676PT3 0853450021  00000005784PT4 0853450023  00000005880PT0 0853450026  00000006244PT8 0853450038  00000006244PT8 0853450038  00000006501PT1 0853450062  00000004624PT3 06008PET01  00000005169PT8 01293DST01  UNK51012720003 10495DST01  00000005892PT5 561883DST  00000006457PT6 388559737PETMET"
    }
  ]
}