  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Rochester",
      "address_1": "150 Verona St",
      "reason_for_recall": "Software error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters.  As such, the user expects that the measurement is taken calculated in centimeters, consistent with other values in the report.  In actuality, the measurement is taken in millimeters.  When this updated measurement is used in an average calculation, the result appears incorrect as two measurements are interpreted as centimeters while the user corrected value is interpreted as millimeters.  If the user selects a Calc Result display as Min or Max, the values are also interpreted as millimeters when centimeters were expected.",
      "address_2": "",
      "product_quantity": "US: 14 units; Foreign: 16 units",
      "code_info": "Serial Numbers: US units: 106, 112, 114, 116, 1021, 1037, 1038, 1044, 1057, 1067, 1072, 1074, 1079, 1090; OUS units: 1005, 1006, 1009, 1011, 1015, 1017, 1018, 1020, 1031, 1041, 1042, 1043, 1045, 1081, 1086, 1089.",
      "center_classification_date": "20170119",
      "distribution_pattern": "Worldwide Distribution - US including SC, GA, TX, OH, IA, NY and foreign distribution to Switzerland and Malaysia",
      "state": "NY",
      "product_description": "Carestream Touch Prime, Catalog # 1738830, and Touch Prime XE, Catalog # 1738822, Ultrasound System  Diagnostic ultrasound imaging or fluid flow analysis of the human body",
      "report_date": "20170125",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Carestream Health Inc",
      "recall_number": "Z-1052-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76213",
      "termination_date": "20170307",
      "more_code_info": "",
      "recall_initiation_date": "20161121",
      "postal_code": "14608-1733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}