  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70240",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1052-2015",
      "product_description": "Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04.      Product Usage:  The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.",
      "product_quantity": "67963",
      "reason_for_recall": "Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.",
      "recall_initiation_date": "20141211",
      "center_classification_date": "20150204",
      "termination_date": "20151023",
      "report_date": "20150211",
      "code_info": "Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04:  42-5279-002-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 10mm;  42-5279-002-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 12mm;  42-5279-002-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 14mm;  42-5279-004-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 10mm;  42-5279-004-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 12mm;  42-5279-004-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 14mm;  42-5279-006-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 10mm;  42-5279-006-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 12mm;  42-5279-006-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 14mm;  42-5279-008-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 10mm;  42-5279-008-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 12mm;  42-5279-008-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 14mm;  42-5279-003-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 10mm;  42-5279-003-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 11mm;  42-5279-003-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 12mm;  42-5279-003-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 13mm;  42-5279-003-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 14mm;  42-5279-005-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 10mm;  42-5279-005-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 11mm;  42-5279-005-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 12mm;  42-5279-005-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 13mm;  42-5279-005-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 14mm;  42-5279-007-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 10mm;  42-5279-007-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 11mm;  42-5279-007-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 12mm;  42-5279-007-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 13mm;  42-5279-007-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 14mm."
    }
  ]
}