  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73248",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including all states in including PR except ND, VT & WY.  Foreign Distribution:  ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, BURKINA FASO, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE,HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRA N, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA  .",
      "recall_number": "Z-1051-2016",
      "product_description": "LightSpeed 16 (MDAS only  , , 2339985  LightSpeed 4.0 CT Scanner System - The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. PET Discovery ST system - The GE Discovery ST system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging system. LightSpeed Pro16 - The Lightspeed 5.0 CT Scanner System is indicated for head and whole body X-ray computed tomography applications. 16 Slice BGO PET/CT System - The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems. 6 Slice BGO PET/CT System - The GE Discovery LS System is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery LS is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery LS can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.",
      "product_quantity": "324",
      "reason_for_recall": "GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.",
      "recall_initiation_date": "20160210",
      "center_classification_date": "20160304",
      "termination_date": "20160921",
      "report_date": "20160316",
      "code_info": "Mfg Lot or Serial # System ID  00000350495CN8 A5333025  00001001853YM9 AL4004CT01  00000953838YM0 AL4023CT01  00001001879YM4 0002876927  00000357924CN0 1577LS16  00001001879YM4 0002876927  00001001879YM4 0002876927  00001001879YM4 0002876927  00001010424YM8 0910212127  00001010461YM0 0910212129  00001010461YM0 0910212129  00001010471YM9 0910213084  00000980093YM9 0910212119  00000980093YM9 0910212119  00000994084YM2 0910212123X  00000984698YM1 0910213080  00001010387YM7 0910215023  00001010475YM0 0910215025  00001001897YM6 AH5703CT01  00000980120YM0 AH4001PE01  00000988410YM7 110002CT02  00000953828YM1 110017CT01  00000980106YM9 110040CT03  00000953853YM9 110059CT01  00000988410YM7 110002CT02  00000980111YM9 030355CT03  00000980111YM9 030355CT03  00001002103YM8 2858149  00000383739CN0 CT383739CN0  00000980108YM5 BF1012CT01  00000980108YM5 BF1012CT01  00000980108YM5 BF1012CT01  00000988391YM9 CT988391YM9  00000988391YM9 CT988391YM9  00000988391YM9 CT988391YM9  00001002107YM9 082421120191  00000364134CN7 082445160009  00000365398CN7 082445040002  00000363607CN3 082421100316  00001002106YM1 082421110048  00000953856YM2 082421120171  00001001852YM1 082421120187  00001010423YM0 082421100294  00001010444YM6 082421100301  00000941008YM5 082421020154  00000988393YM5 082421030172  00000996079YM0 082421030175  00000988398YM4 082421050047  00000996082YM4 082421100157  00000984688YM2 082421100270  00000988395YM0 082421100276  00000988397YM6 082421100278  00000984685YM8 082421100283  00001010426YM3 082421160169  00001010436YM2 082421160176  00000980095YM4 082421180052  00001010457YM8 082421180061  00001023062YM1 082421180064  00000984699YM9 082421010160  00000941008YM5 082421020154  00000994089YM1 082421100284  00001002104YM6 082421120188  00000984700YM5 082421130071  00000988386YM9 082421130073  00001010451YM1 082421100307  00001010455YM2 082421290042  00001010478YM4 EE1020CT01  00001010478YM4 EE1020CT01  00000953833YM1 FI1052CT02  00000994076YM8 A9885025  00000994074YM3 C2195260  00000994074YM3 C2195260  00000994076YM8 A9885025  00000994086YM7 GOLD010094  00000363551CN3 X265413001  00000988394YM3 M1612804  00000984701YM3 M2311502  00000994086YM7 GOLD010094  00001023052YM2 M4162509  00000953834YM9 M4189251  00000953830YM7 M4217029  00000953837YM2 M4400629  00001001899YM2 M4480007  00000980112YM7 M4490502  00000980094YM7 M4505261  00000997212YM6 M4854012  00001001883YM6 M6065310  00000994076YM8 A9885025  00000941016YM8 X100719101  00001023052YM2 M4162509  00000363551CN3 X265413001  00000365551CN3 X265413001  00001023056YM3 X706835004  00000363551CN3 X265413001  00000363551CN3 X265413001  00000984701YM3 M2311502  00000988394YM3 M1612804  00000997212YM6 M4854012  00001001883YM6 M6065310  00000994074YM3 C2195260  00000984701YM3 M2311502  00000997212YM6 M4854012  00001023052YM2 M4162509  00001001883YM6 M6065310  00000980117YM6 GOLD010585  00000980117YM6 GOLD010585  00001001860YM4 420127CT01  00001010403YM2 439532CT01  00001023050YM6 442365CT01  00000980113YM5 PROS01CT04  00001023051YM4 UHAN01CT07  00000980113YM5 PROS01CT04  00000980113YM5 PROS01CT04  00000988408YM1 393115CT01  00000953843YM0 408749CT01  UNK2378661-305 439532CT01  UNK23786610001 CC4821CT03  00000941024YM2 HC1649CT01  00000953843YM0 408749CT01  00000988384YM4 HC4912CT01  00000988382YM8 HC4912CT02  00000953855YM4 HC7946CT03  UNK23786610000 HC7946CT03  UNK2378661-305 439532CT01  00000988382YM8 HC4912CT02  00001001866YM1 PC0075CT05  00000988385YM1 PC0815CT05  UNK2378661-300 PC4176CT02  00000941022YM6 PSTU02CT06  00000953843YM0 408749CT01  00000980105YM1 UC1923CT05  00000941024YM2 HC1649CT01  00000980105YM1 UC1923CT05  00000941022YM6 PSTU02CT06  00000953855YM4 HC7946CT03  00000988384YM4 HC4912CT01  00001001878YM6 00257CTS04  00000953843YM0 408749CT01  00000953843YM0 408749CT01  00001001888YM5 H10632CT08  00001001888YM5 H10632CT08  00001001888YM5 H10632CT08  00001001888YM5 H10632CT08  00001001888YM5 H10632CT08  00001001886YM9 H2094CT11  00001023049YM8 H2112CT02  00000984691YM6 H5826CT01  00001010448YM7 0828210052  00001023047YM2 HU1114CT01  00000988402YM4 083021061102211  00001010441YM2 083021602200264  00000408732CN6 0002777955  00000408732CN6 0002777955  00000341516CN3 IL1020NM31  00000980121YM8 IL1082CT02  00000341516CN3 IL1020NM31  00000341516CT3 IL1082CT03  00001001875YM2 A5348207  00000994075YM0 A5258505  00000941013YM5 A257527001  UNK2378661-302 A5111130  00001010481YM8 A5118059  00000996077YM4 A5124913  00001001869YM5 A5128405  00000364957CN1 A5129956  00001001869YM5 A5128405  00001010476YM8 A5311007  00001023057YM1 A5379305  00001010440YM4 A5387105  00001001874YM5 A5618325  00001010473YM5 D5249901  00000953846YM3 A224857323  00000980115YM0 A5110016  UNK2378661-303 A5127399  00000984684YM1 A5131848  00000941018YM4 A5144908  00000953852YM1 A5164340  00000980115YM0 A5110016  00000364182CN6 A5215931  00000984692YM4 A5358103  00000988388YM5 A5361302  00000980116YM8 A5413819  00000980115YM0 A5110016  UNK2378661-304 B5343102  00000980114YM3 H5382003  00000941036YM6 QT3029  00000941014YM3 QT3030  00000941031YM7 QT3031  00000941033YM3 QT3032  00000953829YM9 QT3033  00000953850YM5 QT3034  00000953847YM1 QT3036  00000980100YM2 QT3037  00000980098YM8 QT3039  00000980096YM2 QT4049  00000980101YM0 QT4050  00000994079YM2 QT4051  00001001854YM7 QT4056  00001001855YM4 QT4057  00001010392YM7 QT4061  00000996072YM5 QT4062  00001010401YM6 QT4063  00001023043YM1 QT4065  00001023042YM3 QT4066  00001010416YM4 QT4077  00000953831YM5 QW3056  00000953849YM7 QW3057  00000980097YM0 QW3059  00000988399YM2 QW3064  00001001857YM0 QW3068  00001023044YM9 QW4087  00000953849YM7 QW3057  0000000G16ME01 QX0044  00000953827YM3 QY0012  00000980099YM6 QZ0001  00000941032YM5 QZ0021  00001023046YM4 SG0090  00000941035YM8 XF0010  00000941010YM1 XF0018  00000941025YM9 QT3027  00000941030YM9 QT3028  00000953832YM3 QT3035  00001023058YM9 QT3035  00000980103YM6 QT3040  00000980102YM8 QT3041  00000980107YM7 QT3042  00000941034YM1 QT3044  00000984686YM6 QT3045  00000984697YM3 QT3047  00000988400YM8 QT3048  00000996076YM6 QT4052  00000980110YM1 QT4054  00001001856YM2 QT4055  00001001895YM0 QT4058  00000988389YM3 QT4060  00001010452YM9 QT4064  00001023045YM6 QT4074  00001010449YM5 QV0030  00000941026YM7 QW3051  00000941028YM3 QW3060  00000941020YM0 QW3077  00000941023YM4 QW3078  00000988392YM7 QW3080  00001010453YM7 QW4088  00000988409YM9 QY0014  00000996073YM3 QY0041  00000980119YM2 QZ0020  00000941012YM7 XF0022  00000988404YM0 XF0149  00001010437YM0 0850210570  00001010439YM6 0850210572  00000984687YM4 0850210560  00000994072YM7 0850210562  00000941017YM6 KW1042CT01  00001023064YM7 KW1008CT03  00000941027YM5 KW1013CT01  00000941017YM6 KW1042CT01  00000941017YM6 KW1042CT01  00001010472YM7 LB1160CT01  00001010472YM7 LB1160CT01  00001010472YM7 LB1160CT01  00000953854YM7 LY1018CT01  00001010454YM5 0834210024  00001010454YM5 0834210024  00000354679KN3 314LS2543020  00001001867YM9 449CT2893654  00000361838CN6 44LS2551149  00000365448CN0 57CTLS16  00000953851YM3 MA1013CT02  UNK2378661-301 DVT0122  00001010463YM6 0920219033  00000361714CN9 0843210103  00001010432YM1 PL1019CT02  00001001885YM1 PL1235CT03  00001001885YM1 PL1235CT03  00001010460YM2 RU1078CT02  00001023060YM5 RU2526CT01  00001001898YM4 RU2705CT01  00000994077YM6 RU1158CT01  00001023065YM4 SA1387CT01  00000984683YM3 0847210020  00001001893YM5 ZA1071CT01  00000994088YM3 ZA1127CT02  00000994088YM3 ZA1127CT02  00001010442YM0 ZA2051CT01  00001023048YM0 ZA2244CT01  00001001891YM9 EPI28404  00000996081YM6 FPA99409  00001010462YM8 YP308712  00001010419YM8 082421140154  00001001858YM8 MPX30218  00001010446YM1 C003CT01  00001023063YM9 L006CT01  00000984694YM0 475230CT04  00000988396YM8 529170CT05  00000988396YM8 529170CT05  00000988396YM8 529170CT05  00000994078YM4 0853210154  00001010465YM1 01017CTS01  00001001892YM7 06013CTS01  00000953842YM2 38006CTS01  00001010406YM5 UK1085CT01  00000988387YM7 AE1021CT02  00001001878YM6 00257CTS04  00000980109YM3 00416CTS03  00000941019YM2 00432CTS03  00000953844YM8 00468CTS01  00000984693YM2 00664CTS02  00001001873YM7 01211CTS02  00001001872YM9 70365CTS03  00001001887YM7 00094CTS01  00000953835YM6 10470CTS01  00000953845YM5 10476CTS01  00000953848YM9 10479CTS01  00000988403YM2 10486CTS01  00001001896YM8 77266CTS02  00001001896YM8 77266CTS02  00000953841YM4 00314CTS02  00001010390YM1 941747LS16  00001010390YM1 941747LS16  00001010474YM3 928669CT  00001001861YM2 770638LS16  00001001861YM2 770638LS16  00000344959CN2 ZWL0470"
    }
  ]
}