  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73248",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including all states in including PR except ND, VT & WY.  Foreign Distribution:  ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, BURKINA FASO, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE,HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRA N, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA  .",
      "recall_number": "Z-1050-2016",
      "product_description": "Discovery ST, , 2339985  LightSpeed 4.0 CT Scanner System - The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. PET Discovery ST system - The GE Discovery ST system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging system. LightSpeed Pro16 - The Lightspeed 5.0 CT Scanner System is indicated for head and whole body X-ray computed tomography applications. 16 Slice BGO PET/CT System - The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems. 6 Slice BGO PET/CT System - The GE Discovery LS System is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery LS is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery LS can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.",
      "product_quantity": "522",
      "reason_for_recall": "GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.",
      "recall_initiation_date": "20160210",
      "center_classification_date": "20160304",
      "termination_date": "20160921",
      "report_date": "20160316",
      "code_info": "Mfg Lot or Serial # System ID  00000312256CN1 4860LS16GS  00000312256CN1 4860LS16GS  00000312256CN1 4860LS16GS  00000294153CN2 0910212097  00000309306CN9 0910212101  00000309960CN3 0910212104  00000309960CN3 0910212104  00000311052CN5 0910212105  00000321402CN0 0910212106  00000321402CN0 0910212106  00000337340CN4 0910212115  00000337340CN4 0910212115  00000337340CN4 0910212115  00000297003CN6 0910213064  00000313645CN4 0910213070  00000301209CN3 0910215018  00000313097CN8 0910216036  00000337344CN6 0910216041  00000297460CN8 AH1410CT06  00000325191CN5 AH5362CT02  00000310728CN1 110037CT01  00000310728CN1 110037CT01  00000299935CN7 190020CT03  00000311058CN2 190052CT06  00000297386CN5 190059CT04  00000311058CN2 190052CT06  00000311426CN1 030018CT01  00000297596CN9 BA4142CT01  00000297596CN9 BA4142CT01  00000297596CN9 BA4142CT01  00000327271CN3 10026  00000346702CN4 1350651  00000310682CN0 2850998  00000310682CN0 2850998  00000327815CN7 CT327815CN7  00000332075CN1 BG4037CT01  UNK23399850000 BG4037CT01  00000316663CN4 BG4042CT01  00000316663CN4 BG4042CT01  00000327081CN6 BG4608CT01  00000314635CN4 BG4600CT01  00000357475CN3 CM1007CT02  00000357475CN3 CM1007CT02  00000357475CN3 CM1007CT02  00000361978CN0 082445020003  00000361978CN0 082445020003  00000361978CN0 082445020003  00000352887CN4 082445030003  00000347756CN9 082445100008  00000347756CN9 082445100008  00000309961CN1 082445150001  00000336259CN7 082445150008  00000359854CN7 082445160007  00000354676CN9 082445250002  00000294148CN2 HLGSCTD10002  00000294148CN2 HLGSCTD10002  00000299483CN8 082421111101  00000334898CN4 082421140150  00000322331CN0 082421160145  00000298653CN7 082421030115  00000333002CN4 082421030162  00000295411CN3 082421030340  00000299934CN0 082421040093  00000295644CN9 082421040096  00000299483CN8 082421111101  00000308733CN5 082421040103  00000299951CN4 082421030354  00000299951CN4 082421030354  00000299951CN4 82421030354  00000322333CN6 082421040120  00000295409CN7 082421050034  00000323331CN9 082421050038  00000343124CN4 082421060040  00000317099CN0 082421070116  00000330529CN9 082421070130  00000297900CN3 082421070368  00000323328CN5 082421080059  00000320550CN7 082421100258  00000323310CN3 082421100259  00000322857CN4 082421100260  00000322857CN4 082421100260  00000335318CN2 082421100265  00000336002CN1 082421130067  00000329307CN3 082421160152  00000342614CN5 082421160155  00000295117CN6 082421210084  00000319305CN9 082421220038  00000297900CN3 082421070368  00000331165CN1 082421010103  00000335317CN4 082421010157  00000333000CN8 082421020090  00000287474CN1 082421130055  00000295411CN3 082421030340  00000308614CN7 082421130065  00000336002CN1 082421130067  00000314492CN0 082421140117  00000356887CN0 082421220038  00000336001CN3 082421220044  00000323308CN7 082421020070  00000332999CN2 082421140145  00000295644CN9 082421040096  00000346880CN8 2532905LIG16  00000337342CN0 374521381CT  00000337342CN0 374521381CT  00000295118CN4 DK1018CT01  00000297992CN0 DK1025CT02  00000313915CN1 11390744CT  00000313915CN1 11390744CT  00000323858CN1 FI1003CT03  00000295643CN1 FI1017CT02  00000323858CN1 FI1003CT03  00000295643CN1 FI1017CT02  00000304685CN1 A4167313  00000319308CN3 B4195716  00000313650CN4 C6010809  UNK23399850001 GOLD001266  00000298318CN7 GOLD010093  00000310120CN1 X966077004  00000313650CN4 C6010809  00000324460CN5 M1020808  00000326112CN0 M1279722  00000311808CN0 M2312304  00000293009CN7 M2322212  00000307022CN4 M23737001  00000311059CN0 M2486504  00000305523CN3 M2800910  00000317193CN1 M2848414  00000298344CN3 M40345153  00000298327CN8 M4047054  00000298318CN7 GOLD010093  00000320672CN9 M4160064  00000303583CN9 M4163021  00000324985CN1 M4174504  00000298659CN4 M4180123  00000310268CN8 M4191025  00000326084CN1 M4192518  00000325672CN4 M4198036  00000326326CN6 M4198185  00000304683CN6 M4215536  00000336000CN5 M4409416  00000336003CN9 M4491526  00000309305CN1 M4491923  00000319062CN6 M4497247  00000302751CN3 M4516011  00000293499CN0 M5481223  00000315232CN9 M5647104  00000326111CN2 M8050904  00000296278CN5 M9202219  00000297389CN9 M9481828  00000297389CN9 M9481828  00000315232CN9 M5647104  00000305523CN3 M2800910  00000297001CN0 270494CT02  00000310266CN2 X751962001  00000311422CN0 X90281601  00000318593CN1 X965468001  00000319063CN4 X976365001  00000322856CN6 X98683601  00000329304CN0 X993729001  00000336000CN5 M4409416  00000325672CN4 M4198036  00000304683CN6 M4215536  00000298327CN8 M4047054  00000311059CN0 M2486504  00000336000CN5 M4409416  00000329304CN0 X993729001  00000319063CN4 X976365001  00000309305CN1 M4491923  00000317193CN1 M2848414  00000307022CN4 M23737001  00000324985CN1 M4174504  00000319062CN6 M4497247  00000298659CN4 M4180123  00000311422CN0 X90281601  00000326111CN2 M8050904  00000319062CN6 M4497247  00000319062CN6 M4497247  00000311059CN0 M2486504  00000298327CN8 M4047054  00000307022CN4 M23737001  00000304683CN6 M4215536  00000324985CN1 M4174504  00000304683CN6 M4215536  00000329303CN2 GOLD010082  00000329303CN2 GOLD010082  00000329303CN2 GOLD010082  00000309522CN1 GOLD001588  00000297427CN7 GOLD001604  00000293010CN5 GOLD002151  00000319948CN6 GOLD002182  00000293010CN5 GOLD002151  00000297427CN7 GOLD001604  00000319948CN6 GOLD002182  00000309522CN1 GOLD001588  00000296977CN2 410014CT01  00000288652CN1 CC0618CT02  00000313361CN8 HC0399CT01  00000301205CN1 HC1627CT04  00000310261CN3 HC2252CT01  00000294145CN8 HC3076CT01  00000302750CN5 HC3314CT02  00000327818CN1 HC4438CT01  00000311056CN6 PC1268CT03  00000317102CN2 PC1401CT01  00000302754CN7 PC7479CT02  00000299235CN2 PC8118CT03  00000294146CN6 UC1590CT01  00000296977CN2 410014CT01  00000310261CN3 HC2252CT01  00000294146CN6 UC1590CT01  00000302754CN7 PC7479CT02  00000309302CN8 HC3428CT02  00000310261CN3 HC2252CT01  00000302754CN7 PC7479CT02  00000305953CN2 05252CTS01  00000305953CN2 05252CTS01  00000294146CN6 UC1590CT01  00000327079CN0 GP1036CT01  00000297599CN3 0828210042  00000309521CN3 0828210045  00000322338CN5 0828210046  00000330530CN7 0828210047  00000323258CN4 0828210094  00000323258CN4 0828210094  00000325192CN3 HU1137CT02  00000319061CN8 Unlocated  00000317789CN6 GOLD010275  00000313648CN8 HU1008CT02  00000297465CN7 HU1039CT03  00000297465CN7 HU1039CT03  00000300567CN5 083021840009012  00000331758CN3 08302100210134X  00000342119CN5 083021042101703  00000345300CN8 083021242100044  00000307945CN6 083021592012120  00000332732CN7 083021602101433  00000311812CN2 083021802100123  00000331758CN3 08302100210134X  00000336854CN5 083021824542213  00000322337CN7 083021842100841  00000336854CN5 083021824542213  00000333821CN7 IR1050CT02  00000330534CN9 IL1000CT01  00000314489CN6 A5816714  00000298365CN8 A206873404  00000294150CN8 A224457303  00000314489CN6 A5816714  00000298364CN1 A5109820  00000329300CN8 A5114983  UNK23399850002 A5127399  00000297387CN3 A5158020  00000298788CN1 A5248810  00000297006CN9 A5279809  00000297467CN3 A5315535  00000305241CN2 A5333021  00000325670CN8 A5364424  00000318594CN9 A5366613  00000297991CN2 A5375119  00000297991CN2 A5375119  00000298788CN1 A5248810  00000308141CN1 A5630204  00000314489CN6 A5816714  00000296974CN9 A5824009  00000316659CN2 A95819201  00000297991CN2 A5375119  00000308982CN8 QW2001  00000294505CN3 QW2006  00000295885CN8 QW2007  00000294505CN3 QW2006  00000304155CN5 QW2021  00000304690CN1 QW2023  00000304692CN7 QW2024  00000305242CN0 QW2025  00000307953CN0 QW2026  00000305954CN0 QW2027  00000305522CN5 QW2028  00000310727CN3 QW3032  00000303514CN4 QW3033  00000311054CN1 QW3036  00000304691CN9 QW3037  00000310876CN8 QW3039  00000303582CN1 QW3040  00000313918CN5 QW3042  00000317976CN9 QW3044  00000309301CN0 QW3045  00000314632CN1 QW3052  00000324987CN7 QW3054  00000325671CN6 QW3055  00000320665CN3 QW3058  00000331765CN8 QW3070  00000304158CN9 QY0002  00000297463CN2 QY0006  00000310122CN7 QY0058  00000313651CN2 QZ0064  00000352579CN7 YE0201  00000357355CN7 YE0202  00000361979CN8 YE0203  00000354674CN4 YE0205  00000365101CN5 YE0210  00000366340CN8 YE0211  00000348807CN9 YE0214  00000293890CN0 QW2002  00000292947CN9 QW2004  00000294768CN7 QW2005  00000298654CN5 QW2009  00000297005CN1 QW2010  00000299761CN7 QW2012  00000298655CN2 QW2013  00000301085CN7 QW2014  00000301207CN7 QW2015  00000295646CN4 QW2018  00000299762CN5 QW2019  00000301206CN9 QW2022  00000310264CN7 QW2029  00000307949CN8 QW3031  00000313364CN2 QW3041  00000297007CN7 QW3043  00000322855CN8 QW3047  00000326660CN8 QW3047  00000321451CN7 QW3048  00000322120CN7 QW3048  00000326662CN4 QW3053  00000297005CN1 QW2010  00000298361CN7 QX0014  00000297007CN7 QW3043  00000297007CN7 QW3043  00000303436CN0 QY0017  00000296978CN0 QY0044  00000301211CN9 QZ0006  00000298785CN7 QZ0008  00000294147CN4 QZ0009  00000293891CN8 QZ0036  00000295415CN4 QZ0039  00000297597CN7 SG0033  00000292948CN7 XG0009  00000354889CN8 YE0204  00000305243CN8 QW3034  00000305243CN8 QW3034  00000348807CN9 YE0214  00000311418CN8 JO1000CT02  00000298652CN9 JO1009CT03  00000298652CN9 JO1009CT03  00000307946CN4 ZA6002CT02  00000295412CN1 0850210523X  00000295413CN9 0850210532  00000309310CN1 0850210548X  00000314641CN2 0850210549  00000323317CN8 0850210550X  00000329993CN0 0850210825  00000329305CN7 0850210552  00000335459CN4 0850210554  00000332082CN7 0850210555  00000334466CN0 0850210873  00000334880CN2 0850210557  00000336853CN7 0850210558X  00000336257CN1 0850210837  00000329993CN0 0850210825  00000336257CN1 0850210837  00000334466CN0 0850210873  00000322332CN8 KW1012CT02  00000321964CN9 LB1050CT01  00000309523CN9 110807CT01  00000294851CN1 0834210019  00000294851CN1 0834210019  00000294502CN0 CT294502CN0  00000346391CN6 2535997DST  00000346391CN6 2535997DST  00000359395CN1 2546945DST  00000295414CN7 73801LS16S  00000294502CN0 CT294502CN0  00000294502CN0 CT294502CN0  00000307947CN2 NO1052CT02  00000293035CN2 OM1043CT01  00000293035CN2 OM1043CT01  00000297995CN3 0843210037  00000301208CN5 0843210038  00000343087CN3 0843210116  00000343087CN3 0843210116  00000298359CN1 600042CT01  00000305951CN6 600062CT01  00000298359CN1 600042CT01  00000305951CN6 600062CT01  00000314495CN3 PPR56604  00000314495CN3 PPR56604  00000314495CN3 PPR56604  00000314493CN8 809780LS16  00000310681CN2 809740LS16  00000297897CN1 RE1001CT02  00000297897CN1 RE1001CT02  00000328357CN9 RU2116CT01  00000296276CN9 SA1011CT02  00000328360CN3 SA1012CT08  00000328359CN5 SA1007CT03  00000317094CN1 SA1014CT02  00000310875CN0 SA1279CT01  00000001380806 0847210020  00000001380806 0847210020  00000303589CN6 SK1003CT01  00000327082CN4 SK1056CT01  00000323315CN2 ZA1000CT02  00000323315CN2 ZA1000CT02  00000310270CN4 ZA1247CT01  00000318926CN3 ZA1442CT02  00000314636CN2 ZA2103CT01  00000314636CN2 ZA2103CT01  00000295879CN1 UPW00600  00000308732CN7 082421160123  00000329306CN5 DE188430  00000327272CN1 DEL26410  00000329991CN4 DEL89957  00000329991CN4 DEL89957  00000309308CN5 DEN27243  00000297385CN7 DPB15832  00000293038CN6 DPC32410  00000030411CN8 DPM32132  00000313369CN1 DPM99916  00000297902CN9 FPG03601  00000305945CN8 FPG03720  00000304109CN2 FPG08400  00000305945CN8 FPG03720  00000330699CN0 FPG79903  00000310874CN3 MPX02003  00000329992CN2 NP739714  00000294144CN1 NPV13924  00000310874CN3 MPX02003  00000294144CN1 NPV13924  00000294144CN1 NPV13924  00000297902CN9 FPG03601  00000295879CN1 UPW00600  00000313369CN1 DPM99916  00000304109CN2 FPG08400  00000293038CN6 DPC32410  00000304109CN2 FPG08400  00000295879CN1 UPW00600  00000293038CN6 DPC32410  00000297902CN9 FPG03601  00000313369CN1 DPM99916  00000299766CN6 A010CT07  00000311423CN8 A010CT08  00000325188CN1 A010CT09  00000354675CN1 C011PE01  00000323859CN9 259130CT02  00000297002CN8 260010CT13  00000297002CN8 260010CT13  00000304112CN6 426020CT02  00000323859CN9 259130CT02  00000323859CN9 259130CT02  00000296274CN4 501030CT02  00000297466CN5 529040CT01  00000296274CN4 501030CT02  00000313362CN6 899010CT02  00000346701CN6 900010PE08  00000297002CN8 260010CT13  00000346701CN6 900010PE08  00000296274CN4 501030CT02  00000304112CN6 426020CT02  00000304112CN6 426020CT02  00000299949CN8 0853210138  00000312925CN1 0853210142  00000312925CN1 0853210142  00000318883CN6 0853210145  00000318883CN6 0853210145  00000328912CN1 0853210150  00000338965CN7 0853450020  00000346743CN8 0853450023  00000354795CN7 0853450038  00000354795CN7 0853450038  00000359858CN8 0853450062  00000298792CN3 TN1013CT03  00000298319CN5 TN1006CT03  00000294852CN9 TN1168CT01  00000310879CN2 AE1005CT02  00000310879CN2 AE1005CT02  00000305239CN6 05006CTS01  00000307948CN0 05027CTS03  00000305953CN2 05252CTS01  00000293032CN9 00243CTS05  00000293032CN9 00243CTS05  00000294856CN0 00361CTS02  00000318595CN6 00537CTS01  00000294042CN7 00557CTS01  00000305952CN4 01051CTS02  00000293032CN9 00243CTS05  00000300489CN2 00020CTS01  00000299314CN5 00095CTS01  00000312924CN4 00160CTS02  00000311813CN0 00257CTS02  00000314497CN9 10079CTM62  00000294343CN9 10441CTS01  00000312815CN4 10458CTS01  00000311420CN4 70050CTS02  00000313653CN8 70328CTS03  00000310732CN3 80040CTS02  00000322119CN9 01027CTS01  00000323259CN2 00432CTS03  00000295880CN9 00298CTS07  00000305524CN1 00216CTS04  00000298358CN3 00133CTS02  00000310262CN1 PM1633CT04  00000311810CN6 763588SPCT  00000336801CN6 936538PERC  00000316259CN1 210351MSDLS  00000328245CN6 706378LS16  00000304684CN4 714633CTM  00000294765CN3 251368ACT  00000316660CN0 402887CT16  00000293034CN5 419584CT  00000318591CN5 309680CT  00000317568CN4 214647LS16  00000336801CN6 936538PERC  00000318591CN5 309680CT  00000316660CN0 402887CT16  00000293034CN5 419584CT  00000317568CN4 214647LS16  00000294765CN3 251368ACT  00000316259CN1 210351MSDLS  00000311810CN6 763588SPCT  00000328245CN6 706378LS16  00000294345CN4 SMGCT15  00000296976CN4 38826196CTOCC  00000296976CN4 38826196CTOCC   00000323258CN #N/A"
    }
  ]
}