  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68672",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Distribution to New York.",
      "recall_number": "Z-1050-2015",
      "product_description": "Plum A+ Infusion System with Hospira MedNet List 20792-04-25, Pole Clamp Assembly Part Number 820-95443-008.  A cassette based multi-function infusion system. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual. The pole clamp assembly is used to provide a mechanism to tighten and loosen the grip of the pole clamp to remove the infuser from the IV pole.",
      "product_quantity": "102 infusors",
      "reason_for_recall": "Plum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the pole clamp to the pole.",
      "recall_initiation_date": "20140611",
      "center_classification_date": "20150204",
      "termination_date": "20161128",
      "report_date": "20150211",
      "code_info": "List Number: 20792-04-25;     Serial Numbers: 17737372 ;  17737375 ;  17737377 ;  17737378 ;  17737383 ;  17737385 ;  17737390 ;  17737394 ;  17737395 ;  17737396 ;  17737397 ;  17737398 ;  17737399 ;  17737400 ;  17737402 ;  17737405 ;  17737408 ;  17737409 ;  17737414 ;  17737416 ;  17737419 ;  17737420 ;  17737421 ;  17737422 ;  17737424 ;  17737425 ;  17737426 ;  17737429 ;  17737430 ;  17737431 ;  17737432 ;  17737433 ;  17737435 ;  17737438 ;  17737440 ;  17737442 ;  17737443 ;  17737444 ;  17737445 ;  17737446 ;  17737447 ;  17737448 ;  17737449 ;  17737450 ;  17737451 ;  17737452 ;  17737454 ;  17737456 ;  17737458 ;  17737459 ;  17737460 ;  17737462 ;  17737463 ;  17737464 ;  17737467 ;  17737470 ;  17737475 ;  17737476 ;  17737478 ;  17737479 ;  17737480 ;  17737483 ;  17737485 ;  17737486 ;  17737489 ;  17737490 ;  17737492 ;  17737498 ;  17737500 ;  17737501 ;  17737502 ;  17737506 ;  17737507 ;  17737508 ;  17737511 ;  17737514 ;  17737516 ;  17737518 ;  17737520 ;  17737521 ;  17737522 ;  17737524 ;  17737525 ;  17737526 ;  17737531 ;  17737533 ;  17737535 ;  17737539 ;  17737542 ;  17737545 ;  17737546 ;  17737550 ;  17737551 ;  17737556 ;  17737557 ;  17737558 ;  17737559 ;  17737560 ;  17737562 ;  17737564 ;  17737569 ;  17737575"
    }
  ]
}