  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73212",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Drive",
      "address_2": "N/A",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -  US including AZ and MO.  Foreign distribution to Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, Ecuador, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-1045-2016",
      "product_description": "Maquet Getinge Group 1180.19XO Extension Device.  Manufacturer MAQUET GmbH,   Used to support the upper and lower extremities immediately before, during and after fracture surgery. The Extension Device (1180.19F0) is a table top module for use with the Maquet MAGNUS operating table system and it is attached to the base mounting point of the MAGNUS operating table.",
      "product_quantity": "2 within US; 247 OUS",
      "reason_for_recall": "MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0).  The Extension Device is manufactured by Maquet GmbH, in Germany, and is a table top module for use with the Maquet MAGNUS operating table system.  Maquet GmbH was made aware of a potential issue that can result in the Extension Device becoming unlocked from the table top, causing it to fall down.",
      "recall_initiation_date": "20160107",
      "center_classification_date": "20160303",
      "termination_date": "20170525",
      "report_date": "20160309",
      "code_info": "Catalog Number 1180.19F0; Serial numbers 00003 and 00004."
    }
  ]
}