      "status": "Terminated",
      "city": "Irwindale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67505",
      "recalling_firm": "Biosense Webster, Inc.",
      "address_1": "15715 Arrow Hwy",
      "address_2": "",
      "postal_code": "91706-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including CA, KY, NJ, MA, and UT and Internationally to Belgium, Austria, and Germany.",
      "recall_number": "Z-1045-2014",
      "product_description": "CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124    WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE.",
      "product_quantity": "8 units total (5 units in US)",
      "reason_for_recall": "Biosense Webster has initiated a recall of the  CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only.  If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.",
      "recall_initiation_date": "20140212",
      "center_classification_date": "20140220",
      "termination_date": "20140523",
      "report_date": "20140226",
      "code_info": "Serial No. 13201, 11125, 11519, 11320, 11115.",
      "more_code_info": ""
    }
  ]
}