  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64569",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the states of AZ, CO, DE, FL, ID, LMA, MI, NH, NJ, NM, PA, RI, TX, and VA.",
      "recall_number": "Z-1045-2013",
      "product_description": "Zimmer NexPosure, retractor portals of the NexPosure MIS Access System, Part Numbers or REF numbers: N1830R80-100, N1830R80-090¿, N1830R80-080¿, N1830R80-070¿, N1830R80-060, N1830R50-100, N1830R50-090, N1830R50-080, ¿ N1830R50-070, N1830R50-060, N1830L80-100, N1830L80-090, N1830L80-080, N1830L80-070, N1830L80-060, N1830L50-100, N1830L50-090, N1830L50-080, N1830L50-070, N1830L50-060, ¿ N182450-085, N182450-075, N182450-065, N182450-055, N182450-045.¿ Retractor access system that allows placement of posterior spinal fixation implants.",
      "product_quantity": "58",
      "reason_for_recall": "Possibility that the inner pouch may be compromised. The product is packaged in two pouches - a sterile inner pouch within an external pouch. NexPosure Retractors contained in a breached inner pouch have an increased risk of contamination (i.e.; loss of product sterility) if proper aseptic technique is not followed when transferring the inner package into the sterile field. Patient infection may o",
      "recall_initiation_date": "20130208",
      "center_classification_date": "20130401",
      "termination_date": "20141209",
      "report_date": "20130410",
      "code_info": "Lot Number  61225572          61664689          61664689          61657007          61657015          61657015          61267745          61267745          61267750          61267750          61267753          61267753          61657016          61657016          61801814          61895685          61657017          61657018          61657018          61267756          61267756          61267759          61267759          61267760          61267760          61267761          61267761          61267761          61873348          61657019          61657019          61657031          61657031"
    }
  ]
}