  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5960-012-51 62163094  00-5960-012-52 114687  00-5960-012-52 62162221  00-5960-013-51 61418001  00-5960-013-51 11006251  00-5960-013-51 62119540  00-5960-013-51 62130947  00-5960-013-51 62174403  00-5960-013-52 62120588  00-5960-013-52 62130963  00-5960-014-51 11400330  00-5960-014-51 11006176  00-5960-014-51 11006333  00-5960-014-51 62120591  00-5960-014-51 62131197  00-5960-014-51 62134257  00-5960-014-51 62170032  00-5960-014-52 62012423  00-5960-014-52 11006544  00-5960-014-52 62061891  00-5960-014-52 62121191  00-5960-014-52 62131248  00-5960-014-52 62134703  00-5960-014-52 62171835  00-5960-014-52 62174384  00-5960-015-51 62121224  00-5960-015-51 62131267  00-5960-015-51 62134706  00-5960-015-51 62168365  00-5960-015-52 61641772  00-5960-015-52 62121768  00-5960-015-52 62121770  00-5960-015-52 62131277  00-5960-015-52 62135957  00-5960-016-51 61793525  00-5960-016-51 61784879  00-5960-016-51 62121790  00-5960-016-51 62121792  00-5960-016-51 62131249  00-5960-016-51 62131284  00-5960-016-51 11400431  00-5960-016-51 62171784  00-5960-016-52 61794705  00-5960-016-52 11005960  00-5960-016-52 62122521  00-5960-016-52 62133156  00-5960-016-52 62137533  00-5960-017-51 114691  00-5960-017-51 62122653  00-5960-017-51 62131221  00-5960-017-51 62137511  00-5960-017-51 62168367  00-5960-017-52 62122857  00-5960-017-52 62128361  00-5960-017-52 62132861  00-5960-018-51 11004155  00-5960-018-51 11005166  00-5960-018-51 11005910  00-5960-018-52 11004157  00-5960-018-52 11005164  00-5960-018-52 11005165    00-5950-015-05 61823885  00-5950-015-05 61858477  00-5950-015-05 61876281  00-5950-015-05 61953097  00-5950-015-05 61999987  00-5950-015-05 62030633  00-5950-015-05 62054070  00-5950-015-05 62091541  00-5950-015-05 62138242  00-5950-015-05 62172417  00-5950-012-02 61777545  00-5950-013-01 61854815  00-5950-013-01 62092032  00-5950-013-01 62171087  00-5950-013-02 62005882  00-5950-013-02 62030313  00-5950-013-06 11003339  00-5950-013-06 62171088  00-5950-014-01 61817750  00-5950-014-01 61827935  00-5950-014-01 61832285  00-5950-014-01 61835326  00-5950-014-01 61857891  00-5950-014-01 61857892  00-5950-014-01 61900460  00-5950-014-01 61900462  00-5950-014-01 61906384  00-5950-014-01 61921448  00-5950-014-01 61933834  00-5950-014-01 61940956  00-5950-014-01 61964633  00-5950-014-01 61974381  00-5950-014-01 61980308  00-5950-014-01 11400334  00-5950-014-01 61992035  00-5950-014-01 61992036  00-5950-014-01 61999556  00-5950-014-01 62005884  00-5950-014-01 62011561  00-5950-014-01 62022990  00-5950-014-01 62030319  00-5950-014-01 62039177  00-5950-014-01 62051191  00-5950-014-01 62062520  00-5950-014-01 62092262  00-5950-014-01 62097716  00-5950-014-01 62100467  00-5950-014-01 62107134  00-5950-014-01 62124802  00-5950-014-01 62131186  00-5950-014-01 62142614  00-5950-014-01 62160335  00-5950-014-01 62166742  00-5950-014-01 62171089  00-5950-014-05 61835327  00-5950-014-05 62058212  00-5950-014-05 62122662  00-5950-014-02 61827936  00-5950-014-02 61830369  00-5950-014-02 61844039  00-5950-014-02 61847663  00-5950-014-02 61857895  00-5950-014-02 61857896  00-5950-014-02 61900463  00-5950-014-02 61901179  00-5950-014-02 61922252  00-5950-014-02 61939801  00-5950-014-02 61939802  00-5950-014-02 61964638  00-5950-014-02 61980573  00-5950-014-02 61980574  00-5950-014-02 11400337  00-5950-014-02 61992037  00-5950-014-02 61999557  00-5950-014-02 62007000  00-5950-014-02 62011562  00-5950-014-02 62019491  00-5950-014-02 62030320  00-5950-014-02 62040699  00-5950-014-02 62058769  00-5950-014-02 62068569  00-5950-014-02 62092285  00-5950-014-02 62097717  00-5950-014-02 62097718  00-5950-014-02 62119167  00-5950-014-02 62126098  00-5950-014-02 62126099  00-5950-014-02 62137073  00-5950-014-02 62149010  00-5950-014-02 62148759  00-5950-014-02 62167498  00-5950-014-02 62171787  00-5950-014-06 61835328  00-5950-014-06 62039178  00-5950-014-06 62101538  00-5950-014-06 62126114  00-5950-014-06 62171788  00-5950-015-01 61803564  00-5950-015-01 61808803  00-5950-015-01 61817775  00-5950-015-01 61819932  00-5950-015-01 61824790  00-5950-015-01 61824789  00-5950-015-01 61826931  00-5950-015-01 61832286  00-5950-015-01 61832287  00-5950-015-01 61835923  00-5950-015-01 61844895  00-5950-015-01 61858472  00-5950-015-01 61858475  00-5950-015-01 61867605  00-5950-015-01 61872333  00-5950-015-01 61901180  00-5950-015-01 61901181  00-5950-015-01 61901216  00-5950-015-01 61905267  00-5950-015-01 61910195  00-5950-015-01 61915307  00-5950-015-01 61926127  00-5950-015-01 61934959  00-5950-015-01 11400291  00-5950-015-01 61940960  00-5950-015-01 61940961  00-5950-015-01 61940959  00-5950-015-01 61952879  00-5950-015-01 61952882  00-5950-015-01 61957848  00-5950-015-01 61962866  00-5950-015-01 61962867  00-5950-015-01 61967197  00-5950-015-01 61972738  00-5950-015-01 61980575  00-5950-015-01 61982303  00-5950-015-01 61986418  00-5950-015-01 61986701  00-5950-015-01 61992806  00-5950-015-01 61999985  00-5950-015-01 61999986  00-5950-015-01 62007696  00-5950-015-01 62007697  00-5950-015-01 62015839  00-5950-015-01 62019492  00-5950-015-01 62024905  00-5950-015-01 62024902  00-5950-015-01 62030587  00-5950-015-01 62030588  00-5950-015-01 62030590  00-5950-015-01 62041498  00-5950-015-01 62041499  00-5950-015-01 62054885  00-5950-015-01 62054083  00-5950-015-01 62062229  00-5950-015-01 62070830  00-5950-015-01 62076992  00-5950-015-01 62076995  00-5950-015-01 62084436  00-5950-015-01 62091542  00-5950-015-01 62092023  00-5950-015-01 62097752  00-5950-015-01 62097754  00-5950-015-01 62102059  00-5950-015-01 62107982  00-5950-015-01 62107983  00-5950-015-01 62122861  00-5950-015-01 62129351  00-5950-015-01 62129352  00-5950-015-01 62131877  00-5950-015-01 62131884  00-5950-015-01 62137976  00-5950-015-01 62137536  00-5950-015-01 62148241  00-5950-015-01 62152619  00-5950-015-01 62152620  00-5950-015-01 62161594  00-5950-015-01 62171842  00-5950-015-01 62171843  00-5950-015-01 62171844  00-5950-015-01 62171841  00-5950-015-02 61803566  00-5950-015-02 61808804  00-5950-015-02 61819042  00-5950-015-02 61819933  00-5950-015-02 61822944  00-5950-015-02 61822945  00-5950-015-02 61822946  00-5950-015-02 61824788  00-5950-015-02 61827933  00-5950-015-02 61829833  00-5950-015-02 61829834  00-5950-015-02 61835395  00-5950-015-02 61844896  00-5950-015-02 61844897  00-5950-015-02 61844898  00-5950-015-02 61850215  00-5950-015-02 61855958  00-5950-015-02 61859223  00-5950-015-02 61868863  00-5950-015-02 61876912  00-5950-015-02 61898491  00-5950-015-02 11400235  00-5950-015-02 61901217  00-5950-015-02 61906068  00-5950-015-02 61910865  00-5950-015-02 61910199  00-5950-015-02 61910864  00-5950-015-02 61916099  00-5950-015-02 61916146  00-5950-015-02 61923568  00-5950-015-02 61929399  00-5950-015-02 61934961  00-5950-015-02 61937758  00-5950-015-02 61937794  00-5950-015-02 61941366  00-5950-015-02 61941367  00-5950-015-02 61954545  00-5950-015-02 61958304  00-5950-015-02 61958306  00-5950-015-02 61962869  00-5950-015-02 61963682  00-5950-015-02 61967898  00-5950-015-02 61967899  00-5950-015-02 61967900  00-5950-015-02 61982460  00-5950-015-02 61982461  00-5950-015-02 61988451  00-5950-015-02 61994856  00-5950-015-02 61994859  00-5950-015-02 62002027  00-5950-015-02 62002028  00-5950-015-02 62009304  00-5950-015-02 62009305  00-5950-015-02 62013490  00-5950-015-02 62014076  00-5950-015-02 62022034  00-5950-015-02 62022035  00-5950-015-02 62026094  00-5950-015-02 62027197  00-5950-015-02 62027195  00-5950-015-02 62032083  00-5950-015-02 62032735  00-5950-015-02 62032082  00-5950-015-02 62046881  00-5950-015-02 62046885  00-5950-015-02 62052667  00-5950-015-02 62052666  00-5950-015-02 62057261  00-5950-015-02 62072516  00-5950-015-02 62072518  00-5950-015-02 62078819  00-5950-015-02 62078818  00-5950-015-02 62085947  00-5950-015-02 62090395  00-5950-015-02 62091410  00-5950-015-02 62091411  00-5950-015-02 62090394  00-5950-015-02 62102061  00-5950-015-02 62109655  00-5950-015-02 62119543  00-5950-015-02 62122863  00-5950-015-02 62129494  00-5950-015-02 62131897  00-5950-015-02 62131905  00-5950-015-02 62134709  00-5950-015-02 62134279  00-5950-015-02 62142616  00-5950-015-02 62148260  00-5950-015-02 62154072  00-5950-015-02 62162760  00-5950-015-02 62169023  00-5950-015-02 62171090  00-5950-015-02 62171091  00-5950-015-02 62171093  00-5950-015-02 62171092  00-5950-015-06 61842346  00-5950-015-06 61847686  00-5950-015-06 61876911  00-5950-015-06 61953101  00-5950-015-06 61992810  00-5950-015-06 62034464  00-5950-015-06 62056780  00-5950-015-06 11400434  00-5950-015-06 62098057  00-5950-015-06 62129496  00-5950-015-06 62169022  00-5950-016-01 61806297  00-5950-016-01 61807851  00-5950-016-01 61821132  00-5950-016-01 61821418  00-5950-016-01 61824772  00-5950-016-01 61824769  00-5950-016-01 61830374  00-5950-016-01 61829844  00-5950-016-01 61835333  00-5950-016-01 61839713  00-5950-016-01 61844588  00-5950-016-01 61859226  00-5950-016-01 61867612  00-5950-016-01 61873441  00-5950-016-01 61895319  00-5950-016-01 61904920  00-5950-016-01 61909345  00-5950-016-01 61909488  00-5950-016-01 61909490  00-5950-016-01 61910202  00-5950-016-01 61915311  00-5950-016-01 61932288  00-5950-016-01 61932290  00-5950-016-01 61940965  00-5950-016-01 61940963  00-5950-016-01 61947195  00-5950-016-01 61952885  00-5950-016-01 61953103  00-5950-016-01 61963894  00-5950-016-01 61963897  00-5950-016-01 61968873  00-5950-016-01 61974018  00-5950-016-01 61974021  00-5950-016-01 61978298  00-5950-016-01 61986182  00-5950-016-01 11400340  00-5950-016-01 61989678  00-5950-016-01 61989680  00-5950-016-01 61989681  00-5950-016-01 11400351  00-5950-016-01 61997353  00-5950-016-01 62002041  00-5950-016-01 11400358  00-5950-016-01 62008287  00-5950-016-01 62009108  00-5950-016-01 62009126  00-5950-016-01 62009127  00-5950-016-01 62018074  00-5950-016-01 62023795  00-5950-016-01 62023796  00-5950-016-01 11400383  00-5950-016-01 62030594  00-5950-016-01 62030595  00-5950-016-01 62040772  00-5950-016-01 62054095  00-5950-016-01 62054910  00-5950-016-01 62061642  00-5950-016-01 62064898  00-5950-016-01 62077017  00-5950-016-01 62077020  00-5950-016-01 62090397  00-5950-016-01 62090398  00-5950-016-01 62096622  00-5950-016-01 62102064  00-5950-016-01 62102066  00-5950-016-01 62102065  00-5950-016-01 62114422  00-5950-016-01 62114432  00-5950-016-01 62119546  00-5950-016-01 62124279  00-5950-016-01 62131228  00-5950-016-01 62138328  00-5950-016-01 62138329  00-5950-016-01 62149041  00-5950-016-01 62154070  00-5950-016-01 62163074  00-5950-016-01 62170901  00-5950-016-01 62170899  00-5950-016-05 61819051  00-5950-016-05 61822951  00-5950-016-05 11400161  00-5950-016-05 61876285  00-5950-016-05 61904193  00-5950-016-05 61953104  00-5950-016-05 61958321  00-5950-016-05 62009314  00-5950-016-05 62014089  00-5950-016-05 62024313  00-5950-016-05 62090142  00-5950-016-05 62098060  00-5950-016-05 62124282  00-5950-016-05 62145401  00-5950-016-05 62170902  00-5950-016-02 11002977  00-5950-016-02 61791939  00-5950-016-02 61806298  00-5950-016-02 61808808  00-5950-016-02 11400105  00-5950-016-02 61821420  00-5950-016-02 61822949  00-5950-016-02 61822950  00-5950-016-02 11400114  00-5950-016-02 61824770  00-5950-016-02 61824771  00-5950-016-02 61829843  00-5950-016-02 61829842  00-5950-016-02 61830371  00-5950-016-02 61830372  00-5950-016-02 61844234  00-5950-016-02 61854819  00-5950-016-02 61860220  00-5950-016-02 61868868  00-5950-016-02 11400205  00-5950-016-02 61894250  00-5950-016-02 61894257  00-5950-016-02 61894397  00-5950-016-02 61894396  00-5950-016-02 61908287  00-5950-016-02 61908289  00-5950-016-02 61910207  00-5950-016-02 61910870  00-5950-016-02 61916151  00-5950-016-02 61916152  00-5950-016-02 61921451  00-5950-016-02 61933696  00-5950-016-02 61933836  00-5950-016-02 61939769  00-5950-016-02 61950746  00-5950-016-02 61955291  00-5950-016-02 61958755  00-5950-016-02 61958756  00-5950-016-02 61964944  00-5950-016-02 61964134  00-5950-016-02 61964439  00-5950-016-02 61974026  00-5950-016-02 61979778  00-5950-016-02 61986506  00-5950-016-02 61986508  00-5950-016-02 61990285  00-5950-016-02 61990287  00-5950-016-02 11400347  00-5950-016-02 61998832  00-5950-016-02 62010344  00-5950-016-02 62011268  00-5950-016-02 62011563  00-5950-016-02 62012286  00-5950-016-02 62012287  00-5950-016-02 62011564  00-5950-016-02 62014091  00-5950-016-02 62015851  00-5950-016-02 62015852  00-5950-016-02 62025570  00-5950-016-02 62025571  00-5950-016-02 62024959  00-5950-016-02 62029864  00-5950-016-02 62030343  00-5950-016-02 62039976  00-5950-016-02 62043146  00-5950-016-02 62055077  00-5950-016-02 62059367  00-5950-016-02 62065378  00-5950-016-02 62068499  00-5950-016-02 62078472  00-5950-016-02 62077295  00-5950-016-02 62088635  00-5950-016-02 62098516  00-5950-016-02 62098793  00-5950-016-02 62101547  00-5950-016-02 62107949  00-5950-016-02 62116018  00-5950-016-02 62116017  00-5950-016-02 62119548  00-5950-016-02 62124811  00-5950-016-02 62131841  00-5950-016-02 62138683  00-5950-016-02 62139369  00-5950-016-02 62149543  00-5950-016-02 62154998  00-5950-016-02 62162233  00-5950-016-02 62170048  00-5950-016-02 62170398  00-5950-016-02 62170049  00-5950-016-02 62170050  00-5950-016-02 62170399  00-5950-016-02 62169049  00-5950-016-06 61822953  00-5950-016-06 61859228  00-5950-016-06 61876286  00-5950-016-06 61903676  00-5950-016-06 61953107  00-5950-016-06 61958324  00-5950-016-06 62009315  00-5950-016-06 62024957  00-5950-016-06 62088879  00-5950-016-06 62098062  00-5950-016-06 62124283  00-5950-016-06 62162224  00-5950-016-06 62169048  00-5950-017-01 61800617  00-5950-017-01 61800616  00-5950-017-01 61814766  00-5950-017-01 61814767  00-5950-017-01 61815711  00-5950-017-01 61829065  00-5950-017-01 61829066  00-5950-017-01 61834859  00-5950-017-01 11400135  00-5950-017-01 61840473  00-5950-017-01 11400139  00-5950-017-01 61844902  00-5950-017-01 61854822  00-5950-017-01 11400156  00-5950-017-01 61860221  00-5950-017-01 61899627  00-5950-017-01 61903675  00-5950-017-01 61912355  00-5950-017-01 61918974  00-5950-017-01 61916155  00-5950-017-01 61934963  00-5950-017-01 11400282  00-5950-017-01 61936776  00-5950-017-01 61936778  00-5950-017-01 61946988  00-5950-017-01 61951696  00-5950-017-01 61950760  00-5950-017-01 11400313  00-5950-017-01 61960714  00-5950-017-01 11400317  00-5950-017-01 61960713  00-5950-017-01 61960712  00-5950-017-01 61973243  00-5950-017-01 61973244  00-5950-017-01 61980313  00-5950-017-01 61984231  00-5950-017-01 61990637  00-5950-017-01 61996360  00-5950-017-01 61996361  00-5950-017-01 11400354  00-5950-017-01 61996945  00-5950-017-01 11400357  00-5950-017-01 61999571  00-5950-017-01 62005000  00-5950-017-01 62005001  00-5950-017-01 62012670  00-5950-017-01 62018146  00-5950-017-01 62018672  00-5950-017-01 62034610  00-5950-017-01 62034607  00-5950-017-01 62038017  00-5950-017-01 62038018  00-5950-017-01 62052668  00-5950-017-01 62052669  00-5950-017-01 62060542  00-5950-017-01 62065390  00-5950-017-01 62066548  00-5950-017-01 62078824  00-5950-017-01 62078822  00-5950-017-01 62084439  00-5950-017-01 62092283  00-5950-017-01 62095239  00-5950-017-01 62102785  00-5950-017-01 62109668  00-5950-017-01 62112976  00-5950-017-01 62120702  00-5950-017-01 62125144  00-5950-017-01 62124820  00-5950-017-01 62134731  00-5950-017-01 62143173  00-5950-017-01 62154276  00-5950-017-01 62160283  00-5950-017-01 62170059  00-5950-017-01 62170392  00-5950-017-01 62170393  00-5950-017-05 61833598  00-5950-017-05 11400238  00-5950-017-05 61937290  00-5950-017-05 61960715  00-5950-017-05 61999572  00-5950-017-05 62036734  00-5950-017-05 62061873  00-5950-017-05 11400440  00-5950-017-05 62121230  00-5950-017-05 62126101  00-5950-017-02 61800618  00-5950-017-02 61800619  00-5950-017-02 61806282  00-5950-017-02 61815714  00-5950-017-02 61827942  00-5950-017-02 61829062  00-5950-017-02 61832308  00-5950-017-02 61835933  00-5950-017-02 61841279  00-5950-017-02 61844942  00-5950-017-02 61870726  00-5950-017-02 61880104  00-5950-017-02 61898497  00-5950-017-02 61898837  00-5950-017-02 61899628  00-5950-017-02 11400232  00-5950-017-02 61902954  00-5950-017-02 61912358  00-5950-017-02 61917758  00-5950-017-02 61920447  00-5950-017-02 61920448  00-5950-017-02 61927969  00-5950-017-02 61934965  00-5950-017-02 61934966  00-5950-017-02 61952883  00-5950-017-02 61952884  00-5950-017-02 11400312  00-5950-017-02 61956827  00-5950-017-02 61961812  00-5950-017-02 61962312  00-5950-017-02 61974125  00-5950-017-02 61974126  00-5950-017-02 61982246  00-5950-017-02 61982464  00-5950-017-02 61990918  00-5950-017-02 61990919  00-5950-017-02 62000987  00-5950-017-02 62005953  00-5950-017-02 62005954  00-5950-017-02 62005955  00-5950-017-02 62006741  00-5950-017-02 62006743  00-5950-017-02 62019996  00-5950-017-02 62026806  00-5950-017-02 62030596  00-5950-017-02 62036588  00-5950-017-02 62036609  00-5950-017-02 62051196  00-5950-017-02 62054065  00-5950-017-02 62060479  00-5950-017-02 62060480  00-5950-017-02 62067915  00-5950-017-02 62073089  00-5950-017-02 62086359  00-5950-017-02 62086360  00-5950-017-02 62091543  00-5950-017-02 62096011  00-5950-017-02 62096621  00-5950-017-02 62105718  00-5950-017-02 62110814  00-5950-017-02 62113002  00-5950-017-02 62113003  00-5950-017-02 62126732  00-5950-017-02 62126115  00-5950-017-02 62132540  00-5950-017-02 62135963  00-5950-017-02 62143620  00-5950-017-02 62149040  00-5950-017-02 62155001  00-5950-017-02 62160794  00-5950-017-02 62170055  00-5950-017-02 62170053  00-5950-017-06 61816389  00-5950-017-06 61917756  00-5950-017-06 61951701  00-5950-017-06 62000986  00-5950-017-06 62030646  00-5950-017-06 62067918  00-5950-017-06 62134748  00-5950-017-06 62160792  00-5950-017-06 62169050",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "LPS-FLEX PRECOAT FEMORAL B-R  LPS-FLEX PRECOAT FEMORAL C-L  LPS-FLEX PRECOAT FEMORAL C-R  LPS-FLEX PRECOAT FEMORAL D-L  LPS-FLEX PRECOAT FEMORAL D-R  LPS-FLEX PRECOAT FEMORAL E-L  LPS-FLEX PRECOAT FEMORAL E-R  LPS-FLEX PRECOAT FEMORAL F-L  LPS-FLEX PRECOAT FEMORAL F-R  LPS-FLEX PRECOAT FEMORAL G-L  LPS-FLEX PRECOAT FEMORAL G-R  LPS-FLEX PRECOAT FEMORAL H-L  LPS-FLEX PRECOAT FEMORAL H-R¿  LPS-FLEX POROUS FEMORAL E-R    This device is indicated for patients with severe knee pain and disability¿",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1044-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}