      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5964-014-51 62171767  00-5964-014-52 61804723  00-5964-014-52 61804724  00-5964-014-52 61804726  00-5964-014-52 61804725  00-5964-014-52 61811767  00-5964-014-52 61811768  00-5964-014-52 61812820  00-5964-014-52 61814746  00-5964-014-52 61814749  00-5964-014-52 61814750  00-5964-014-52 61814747  00-5964-014-52 61814751  00-5964-014-52 61815695  00-5964-014-52 61829046  00-5964-014-52 61832272  00-5964-014-52 61832273  00-5964-014-52 61832274  00-5964-014-52 61839689  00-5964-014-52 61839691  00-5964-014-52 61847653  00-5964-014-52 61847654  00-5964-014-52 61847655  00-5964-014-52 61860204  00-5964-014-52 61860203  00-5964-014-52 61860205  00-5964-014-52 61860206  00-5964-014-52 61866693  00-5964-014-52 61869843  00-5964-014-52 61870709  00-5964-014-52 61874450  00-5964-014-52 61878263  00-5964-014-52 61878262  00-5964-014-52 61883260  00-5964-014-52 61884178  00-5964-014-52 61886312  00-5964-014-52 61890007  00-5964-014-52 61890008  00-5964-014-52 61895306  00-5964-014-52 61894531  00-5964-014-52 61895308  00-5964-014-52 61896038  00-5964-014-52 61901210  00-5964-014-52 61901211  00-5964-014-52 61902939  00-5964-014-52 61901209  00-5964-014-52 61906064  00-5964-014-52 61906381  00-5964-014-52 61915295  00-5964-014-52 61915294  00-5964-014-52 61920422  00-5964-014-52 61931652  00-5964-014-52 61931653  00-5964-014-52 61939172  00-5964-014-52 61939175  00-5964-014-52 61949694  00-5964-014-52 61936489  00-5964-014-52 61951104  00-5964-014-52 61954538  00-5964-014-52 61954539  00-5964-014-52 61954540  00-5964-014-52 61954537  00-5964-014-52 61965930  00-5964-014-52 61965929  00-5964-014-52 61973058  00-5964-014-52 61973228  00-5964-014-52 61973229  00-5964-014-52 61979767  00-5964-014-52 61979768  00-5964-014-52 61980293  00-5964-014-52 61980296  00-5964-014-52 61979766  00-5964-014-52 61988446  00-5964-014-52 61989660  00-5964-014-52 61989661  00-5964-014-52 61993903  00-5964-014-52 61993907  00-5964-014-52 61993910  00-5964-014-52 61993913  00-5964-014-52 62000966  00-5964-014-52 61989662  00-5964-014-52 62010323  00-5964-014-52 62010324  00-5964-014-52 62017251  00-5964-014-52 62017253  00-5964-014-52 62018047  00-5964-014-52 62022925  00-5964-014-52 62022927  00-5964-014-52 62027189  00-5964-014-52 62028442  00-5964-014-52 62028443  00-5964-014-52 62035834  00-5964-014-52 62036141  00-5964-014-52 62041496  00-5964-014-52 62048233  00-5964-014-52 62048236  00-5964-014-52 62052657  00-5964-014-52 62052658  00-5964-014-52 62052659  00-5964-014-52 62076318  00-5964-014-52 62077284  00-5964-014-52 62077839  00-5964-014-52 62079815  00-5964-014-52 62084433  00-5964-014-52 62085524  00-5964-014-52 62085525  00-5964-014-52 62085941  00-5964-014-52 62098052  00-5964-014-52 62096703  00-5964-014-52 62098053  00-5964-014-52 62099804  00-5964-014-52 62099848  00-5964-014-52 62099860  00-5964-014-52 62100456  00-5964-014-52 62105722  00-5964-014-52 62105723  00-5964-014-52 62103206  00-5964-014-52 62105721  00-5964-014-52 62112971  00-5964-014-52 62112988  00-5964-014-52 62112989  00-5964-014-52 62112994  00-5964-014-52 62114093  00-5964-014-52 62114094  00-5964-014-52 62119153  00-5964-014-52 62114367  00-5964-014-52 62124787  00-5964-014-52 62125121  00-5964-014-52 62126096  00-5964-014-52 62130977  00-5964-014-52 62131185  00-5964-014-52 62131201  00-5964-014-52 62134704  00-5964-014-52 62134745  00-5964-014-52 11007251  00-5964-014-52 62143136  00-5964-014-52 62143139  00-5964-014-52 62148749  00-5964-014-52 62154066  00-5964-014-52 62154067  00-5964-014-52 62154036  00-5964-014-52 62160333  00-5964-014-52 62161196  00-5964-014-52 62163070  00-5964-014-52 62164813  00-5964-014-52 62163097  00-5964-014-52 62173159  00-5964-014-52 62173160  00-5964-014-52 62173162  00-5964-014-52 62173175  00-5964-014-52 62173176  00-5964-014-52 62173178  00-5964-014-52 62173179  00-5964-014-52 62173180  00-5964-014-52 62163071  00-5964-015-51 61820639  00-5964-015-51 61820640  00-5964-015-51 61820641  00-5964-015-51 61821139  00-5964-015-51 61829067  00-5964-015-51 61829068  00-5964-015-51 61829069  00-5964-015-51 61829071  00-5964-015-51 61829070  00-5964-015-51 61829846  00-5964-015-51 61835354  00-5964-015-51 61835356  00-5964-015-51 61835357  00-5964-015-51 61840459  00-5964-015-51 61840460  00-5964-015-51 61844225  00-5964-015-51 61844227  00-5964-015-51 61844228  00-5964-015-51 61844885  00-5964-015-51 11400142  00-5964-015-51 61850913  00-5964-015-51 61850914  00-5964-015-51 61856736  00-5964-015-51 61857882  00-5964-015-51 61856735  00-5964-015-51 61857881  00-5964-015-51 61858464  00-5964-015-51 61861342  00-5964-015-51 61861343  00-5964-015-51 61861344  00-5964-015-51 61864866  00-5964-015-51 61867593  00-5964-015-51 61870710  00-5964-015-51 61870711  00-5964-015-51 61870713  00-5964-015-51 61879221  00-5964-015-51 61881047  00-5964-015-51 61881048  00-5964-015-51 61881676  00-5964-015-51 61881678  00-5964-015-51 61886313  00-5964-015-51 61886314  00-5964-015-51 61886315  00-5964-015-51 61888026  00-5964-015-51 61889084  00-5964-015-51 61889085  00-5964-015-51 61889086  00-5964-015-51 61889083  00-5964-015-51 61896039  00-5964-015-51 61897278  00-5964-015-51 61897279  00-5964-015-51 61902941  00-5964-015-51 61902942  00-5964-015-51 61904182  00-5964-015-51 61904183  00-5964-015-51 61904184  00-5964-015-51 61904916  00-5964-015-51 61904915  00-5964-015-51 61904917  00-5964-015-51 11400242  00-5964-015-51 61906382  00-5964-015-51 61907433  00-5964-015-51 61907434  00-5964-015-51 61918714  00-5964-015-51 61918715  00-5964-015-51 61918716  00-5964-015-51 61918717  00-5964-015-51 61918958  00-5964-015-51 61918959  00-5964-015-51 61922244  00-5964-015-51 61922245  00-5964-015-51 61923559  00-5964-015-51 61923560  00-5964-015-51 61926121  00-5964-015-51 61927955  00-5964-015-51 61933492  00-5964-015-51 61935973  00-5964-015-51 61936376  00-5964-015-51 61947006  00-5964-015-51 61943490  00-5964-015-51 61947181  00-5964-015-51 61947183  00-5964-015-51 61947020  00-5964-015-51 61947182  00-5964-015-51 61943489  00-5964-015-51 61952862  00-5964-015-51 61952863  00-5964-015-51 61952864  00-5964-015-51 61943463  00-5964-015-51 61960816  00-5964-015-51 61962852  00-5964-015-51 61972318  00-5964-015-51 61974115  00-5964-015-51 61974117  00-5964-015-51 61976406  00-5964-015-51 61976407  00-5964-015-51 61976408  00-5964-015-51 61976409  00-5964-015-51 61977235  00-5964-015-51 61982271  00-5964-015-51 61982451  00-5964-015-51 61982452  00-5964-015-51 61983002  00-5964-015-51 61982453  00-5964-015-51 61982282  00-5964-015-51 61987073  00-5964-015-51 61988468  00-5964-015-51 61988467  00-5964-015-51 61987053  00-5964-015-51 61995684  00-5964-015-51 61995685  00-5964-015-51 61995686  00-5964-015-51 61996350  00-5964-015-51 61996948  00-5964-015-51 61996950  00-5964-015-51 61996951  00-5964-015-51 61996952  00-5964-015-51 61999573  00-5964-015-51 62000988  00-5964-015-51 62000989  00-5964-015-51 62002051  00-5964-015-51 62003650  00-5964-015-51 62003776  00-5964-015-51 62009308  00-5964-015-51 62015855  00-5964-015-51 62015856  00-5964-015-51 62017264  00-5964-015-51 62016807  00-5964-015-51 62017265  00-5964-015-51 62017266  00-5964-015-51 62017597  00-5964-015-51 62017599  00-5964-015-51 62017600  00-5964-015-51 62029266  00-5964-015-51 62027453  00-5964-015-51 62028235  00-5964-015-51 62028237  00-5964-015-51 62029263  00-5964-015-51 62029856  00-5964-015-51 62027465  00-5964-015-51 62028236  00-5964-015-51 62029877  00-5964-015-51 62030480  00-5964-015-51 62046037  00-5964-015-51 62046036  00-5964-015-51 62046235  00-5964-015-51 62046236  00-5964-015-51 62052711  00-5964-015-51 62054075  00-5964-015-51 62054076  00-5964-015-51 62059369  00-5964-015-51 62059370  00-5964-015-51 62063717  00-5964-015-51 62064893  00-5964-015-51 62065289  00-5964-015-51 62069327  00-5964-015-51 62070227  00-5964-015-51 62070226  00-5964-015-51 62070820  00-5964-015-51 62078048  00-5964-015-51 62079812  00-5964-015-51 62081419  00-5964-015-51 62081865  00-5964-015-51 62086334  00-5964-015-51 62086604  00-5964-015-51 62086605  00-5964-015-51 62086626  00-5964-015-51 62096035  00-5964-015-51 62096721  00-5964-015-51 62096033  00-5964-015-51 62096720  00-5964-015-51 62096722  00-5964-015-51 62098857  00-5964-015-51 62098858  00-5964-015-51 62099077  00-5964-015-51 62099080  00-5964-015-51 62098859  00-5964-015-51 62099260  00-5964-015-51 62102872  00-5964-015-51 62105756  00-5964-015-51 62105758  00-5964-015-51 62105759  00-5964-015-51 62105760  00-5964-015-51 62113022  00-5964-015-51 62113023  00-5964-015-51 62114104  00-5964-015-51 62114105  00-5964-015-51 62114505  00-5964-015-51 62113024  00-5964-015-51 62114499  00-5964-015-51 62114503  00-5964-015-51 62119528  00-5964-015-51 62119530  00-5964-015-51 62119529  00-5964-015-51 62119532  00-5964-015-51 62124886  00-5964-015-51 62124888  00-5964-015-51 62124608  00-5964-015-51 62125119  00-5964-015-51 62131029  00-5964-015-51 62131192  00-5964-015-51 62131211  00-5964-015-51 62131212  00-5964-015-51 62134264  00-5964-015-51 62134265  00-5964-015-51 62142580  00-5964-015-51 62143148  00-5964-015-51 62143154  00-5964-015-51 62143616  00-5964-015-51 62143617  00-5964-015-51 62150842  00-5964-015-51 62150849  00-5964-015-51 62151225  00-5964-015-51 62151230  00-5964-015-51 62152196  00-5964-015-51 62152615  00-5964-015-51 62152215  00-5964-015-51 62160278  00-5964-015-51 62160332  00-5964-015-51 62160279  00-5964-015-51 62161211  00-5964-015-51 62166751  00-5964-015-51 62167520  00-5964-015-51 62168057  00-5964-015-51 62169045  00-5964-015-51 62168350  00-5964-015-51 62168352  00-5964-015-51 62169036  00-5964-015-51 62169044  00-5964-015-51 62169046  00-5964-015-52 61793540  00-5964-015-52 11400106  00-5964-015-52 61814744  00-5964-015-52 61815692  00-5964-015-52 61819044  00-5964-015-52 61819045  00-5964-015-52 61819046  00-5964-015-52 61819921  00-5964-015-52 61819923  00-5964-015-52 61819924  00-5964-015-52 61819922  00-5964-015-52 61829051  00-5964-015-52 61829052  00-5964-015-52 61829053  00-5964-015-52 61829054  00-5964-015-52 61835358  00-5964-015-52 61835818  00-5964-015-52 61835819  00-5964-015-52 61841273  00-5964-015-52 61841271  00-5964-015-52 61844887  00-5964-015-52 61844888  00-5964-015-52 61844889  00-5964-015-52 61844929  00-5964-015-52 61851713  00-5964-015-52 61853363  00-5964-015-52 61853364  00-5964-015-52 61853365  00-5964-015-52 61854839  00-5964-015-52 61855955  00-5964-015-52 61856738  00-5964-015-52 61856739  00-5964-015-52 61862123  00-5964-015-52 61864868  00-5964-015-52 61872323  00-5964-015-52 61873425  00-5964-015-52 61873426  00-5964-015-52 61880089  00-5964-015-52 61880090  00-5964-015-52 61881030  00-5964-015-52 61881679  00-5964-015-52 61881680  00-5964-015-52 61881681  00-5964-015-52 61883270  00-5964-015-52 61883269  00-5964-015-52 61888027  00-5964-015-52 61888028  00-5964-015-52 61888029  00-5964-015-52 61888030  00-5964-015-52 61897280  00-5964-015-52 61897281  00-5964-015-52 61898483  00-5964-015-52 11400237  00-5964-015-52 61904185  00-5964-015-52 61904187  00-5964-015-52 61904186  00-5964-015-52 61905145  00-5964-015-52 61905146  00-5964-015-52 61905147  00-5964-015-52 61907436  00-5964-015-52 61907437  00-5964-015-52 61908278  00-5964-015-52 61907438  00-5964-015-52 61917741  00-5964-015-52 61917743  00-5964-015-52 61918712  00-5964-015-52 61917778  00-5964-015-52 61918713  00-5964-015-52 61917742  00-5964-015-52 61920424  00-5964-015-52 61920425  00-5964-015-52 61920426  00-5964-015-52 61920427  00-5964-015-52 61920428  00-5964-015-52 61921432  00-5964-015-52 61921433  00-5964-015-52 61921434  00-5964-015-52 61921435  00-5964-015-52 61924437  00-5964-015-52 61926116  00-5964-015-52 61926117  00-5964-015-52 61926120  00-5964-015-52 61926119  00-5964-015-52 61933827  00-5964-015-52 61933829  00-5964-015-52 61934948  00-5964-015-52 61934950  00-5964-015-52 11400298  00-5964-015-52 61939791  00-5964-015-52 61940348  00-5964-015-52 61946455  00-5964-015-52 61946456  00-5964-015-52 61949691  00-5964-015-52 61949692  00-5964-015-52 61948134  00-5964-015-52 61953034  00-5964-015-52 61943491  00-5964-015-52 61953037  00-5964-015-52 61953036  00-5964-015-52 61970671  00-5964-015-52 61970672  00-5964-015-52 61971150  00-5964-015-52 61970673  00-5964-015-52 61971151  00-5964-015-52 61972744  00-5964-015-52 61972746  00-5964-015-52 61972319  00-5964-015-52 61973249  00-5964-015-52 61973250  00-5964-015-52 61973251  00-5964-015-52 61973252  00-5964-015-52 61973253  00-5964-015-52 61974043  00-5964-015-52 61980344  00-5964-015-52 61982822  00-5964-015-52 61984203  00-5964-015-52 61982821  00-5964-015-52 61985043  00-5964-015-52 61984205  00-5964-015-52 61986758  00-5964-015-52 61987052  00-5964-015-52 61990662  00-5964-015-52 61990664  00-5964-015-52 61991999  00-5964-015-52 61992001  00-5964-015-52 61990663  00-5964-015-52 61992780  00-5964-015-52 61992992  00-5964-015-52 61992070  00-5964-015-52 61992781  00-5964-015-52 61993977  00-5964-015-52 61993979  00-5964-015-52 62002059  00-5964-015-52 62002060  00-5964-015-52 62005003  00-5964-015-52 62005896  00-5964-015-52 62003762  00-5964-015-52 62005004  00-5964-015-52 62012313  00-5964-015-52 62012315  00-5964-015-52 62012660  00-5964-015-52 62012662  00-5964-015-52 62011568  00-5964-015-52 62015868  00-5964-015-52 62015869  00-5964-015-52 62017274  00-5964-015-52 62017275  00-5964-015-52 62017907  00-5964-015-52 62016825  00-5964-015-52 62022913  00-5964-015-52 62022917  00-5964-015-52 62023339  00-5964-015-52 62023337  00-5964-015-52 62024590  00-5964-015-52 62029557  00-5964-015-52 62029556  00-5964-015-52 62029558  00-5964-015-52 62029559  00-5964-015-52 62024589  00-5964-015-52 62030482  00-5964-015-52 62030620  00-5964-015-52 62030622  00-5964-015-52 62030623  00-5964-015-52 62043314  00-5964-015-52 62046906  00-5964-015-52 62047528  00-5964-015-52 62048276  00-5964-015-52 62048277  00-5964-015-52 62054077  00-5964-015-52 62054650  00-5964-015-52 62054651  00-5964-015-52 62061868  00-5964-015-52 62061869  00-5964-015-52 62061870  00-5964-015-52 62062236  00-5964-015-52 62063719  00-5964-015-52 62064892  00-5964-015-52 62072145  00-5964-015-52 62072149  00-5964-015-52 62072202  00-5964-015-52 62072507  00-5964-015-52 62072203  00-5964-015-52 62087943  00-5964-015-52 62087012  00-5964-015-52 62087944  00-5964-015-52 62087947  00-5964-015-52 62087946  00-5964-015-52 62088863  00-5964-015-52 62091538  00-5964-015-52 62091539  00-5964-015-52 62098077  00-5964-015-52 62098078  00-5964-015-52 62098571  00-5964-015-52 62098572  00-5964-015-52 62098805  00-5964-015-52 62098804  00-5964-015-52 62100195  00-5964-015-52 62100213  00-5964-015-52 62100484  00-5964-015-52 62100485  00-5964-015-52 62107942  00-5964-015-52 62107944  00-5964-015-52 62107960  00-5964-015-52 62109186  00-5964-015-52 62109751  00-5964-015-52 62109754  00-5964-015-52 62109184  00-5964-015-52 62110246  00-5964-015-52 62110796  00-5964-015-52 62116035  00-5964-015-52 62116036  00-5964-015-52 62116038  00-5964-015-52 62116037  00-5964-015-52 62117292  00-5964-015-52 62117294  00-5964-015-52 62118407  00-5964-015-52 62121213  00-5964-015-52 62121217  00-5964-015-52 62121240  00-5964-015-52 62121214  00-5964-015-52 62121241  00-5964-015-52 62125513  00-5964-015-52 62125515  00-5964-015-52 62126107  00-5964-015-52 62132072  00-5964-015-52 62132076  00-5964-015-52 62132094  00-5964-015-52 62132740  00-5964-015-52 62132872  00-5964-015-52 62135958  00-5964-015-52 62136379  00-5964-015-52 62136380  00-5964-015-52 62137052  00-5964-015-52 62144123  00-5964-015-52 62144288  00-5964-015-52 62144312  00-5964-015-52 62144929  00-5964-015-52 62149361  00-5964-015-52 62149362  00-5964-015-52 62149363  00-5964-015-52 62149518  00-5964-015-52 62149533  00-5964-015-52 62149551  00-5964-015-52 62152911  00-5964-015-52 62152640  00-5964-015-52 62152916  00-5964-015-52 62161204  00-5964-015-52 62161222  00-5964-015-52 62162207  00-5964-015-52 62162229  00-5964-015-52 62164877  00-5964-015-52 62164878  00-5964-015-52 62165873  00-5964-015-52 62165945  00-5964-015-52 62166665  00-5964-015-52 62170035  00-5964-015-52 62170390  00-5964-015-52 62170036  00-5964-015-52 62170389  00-5964-015-52 62170391  00-5964-015-52 62170033  00-5964-015-52 62170037  00-5964-015-52 62170388  00-5964-016-51 61783051  00-5964-016-51 61784877  00-5964-016-51 61784878  00-5964-016-51 61783049  00-5964-016-51 61783050  00-5964-016-51 61815795  00-5964-016-51 61817744  00-5964-016-51 61817742  00-5964-016-51 61817746  00-5964-016-51 61821407  00-5964-016-51 61821409  00-5964-016-51 61822964  00-5964-016-51 61821408  00-5964-016-51 61822939  00-5964-016-51 61822963  00-5964-016-51 61822965  00-5964-016-51 61822966  00-5964-016-51 61832304  00-5964-016-51 61833584  00-5964-016-51 61833585  00-5964-016-51 61833588  00-5964-016-51 61835820  00-5964-016-51 61841269  00-5964-016-51 61842336  00-5964-016-51 61844230  00-5964-016-51 61844231  00-5964-016-51 61848276  00-5964-016-51 61859217  00-5964-016-51 61859218  00-5964-016-51 61862128  00-5964-016-51 61862126  00-5964-016-51 61872326  00-5964-016-51 61876276  00-5964-016-51 61876907  00-5964-016-51 61881037  00-5964-016-51 61881038  00-5964-016-51 61885165  00-5964-016-51 61885166  00-5964-016-51 61885167  00-5964-016-51 61885168  00-5964-016-51 61890036  00-5964-016-51 61890759  00-5964-016-51 61890760  00-5964-016-51 61892380  00-5964-016-51 61892382  00-5964-016-51 61899641  00-5964-016-51 61899646  00-5964-016-51 61899645  00-5964-016-51 61900470  00-5964-016-51 61901204  00-5964-016-51 61901207  00-5964-016-51 61901206  00-5964-016-51 61902961  00-5964-016-51 61901412  00-5964-016-51 61909333  00-5964-016-51 61909467  00-5964-016-51 61909471  00-5964-016-51 61916144  00-5964-016-51 61917744  00-5964-016-51 61917746  00-5964-016-51 61917748  00-5964-016-51 61918739  00-5964-016-51 61917747  00-5964-016-51 61920464  00-5964-016-51 61923587  00-5964-016-51 61926125  00-5964-016-51 61926122  00-5964-016-51 61931663  00-5964-016-51 61931662  00-5964-016-51 61936787  00-5964-016-51 61946954  00-5964-016-51 61952869  00-5964-016-51 61943495  00-5964-016-51 61943496  00-5964-016-51 61951655  00-5964-016-51 61959249  00-5964-016-51 61959250  00-5964-016-51 61959252  00-5964-016-51 61960706  00-5964-016-51 11400319  00-5964-016-51 61960707  00-5964-016-51 61970688  00-5964-016-51 61970690  00-5964-016-51 61971186  00-5964-016-51 61971187  00-5964-016-51 61971739  00-5964-016-51 61971740  00-5964-016-51 61971741  00-5964-016-51 61971742  00-5964-016-51 61978516  00-5964-016-51 61976432  00-5964-016-51 61986240  00-5964-016-51 61986243  00-5964-016-51 61985037  00-5964-016-51 61986245  00-5964-016-51 61986733  00-5964-016-51 61987070  00-5964-016-51 61988463  00-5964-016-51 61990657  00-5964-016-51 61990661  00-5964-016-51 61994834  00-5964-016-51 61994838  00-5964-016-51 61994843  00-5964-016-51 61996971  00-5964-016-51 62000984  00-5964-016-51 62002023  00-5964-016-51 62002063  00-5964-016-51 62000983  00-5964-016-51 62002077  00-5964-016-51 62002079  00-5964-016-51 62008258  00-5964-016-51 62010325  00-5964-016-51 62010327  00-5964-016-51 11005907  00-5964-016-51 62011224  00-5964-016-51 62018137  00-5964-016-51 62020797  00-5964-016-51 62020798  00-5964-016-51 62024953  00-5964-016-51 62024954  00-5964-016-51 62026092  00-5964-016-51 62027191  00-5964-016-51 62030583  00-5964-016-51 62030585  00-5964-016-51 62033833  00-5964-016-51 62036524  00-5964-016-51 62041511  00-5964-016-51 62042334  00-5964-016-51 62042335  00-5964-016-51 62041520  00-5964-016-51 62048451  00-5964-016-51 62049418  00-5964-016-51 62049420  00-5964-016-51 62050095  00-5964-016-51 62050096  00-5964-016-51 62061688  00-5964-016-51 62062240  00-5964-016-51 62062242  00-5964-016-51 62070236  00-5964-016-51 62070240  00-5964-016-51 62072204  00-5964-016-51 62077290  00-5964-016-51 62077843  00-5964-016-51 62077844  00-5964-016-51 62082375  00-5964-016-51 62083081  00-5964-016-51 62086987  00-5964-016-51 62087934  00-5964-016-51 62092003  00-5964-016-51 62092257  00-5964-016-51 62092259  00-5964-016-51 62092260  00-5964-016-51 62099275  00-5964-016-51 62099276  00-5964-016-51 62099941  00-5964-016-51 62100478  00-5964-016-51 62100481  00-5964-016-51 62101679  00-5964-016-51 62101704  00-5964-016-51 62102075  00-5964-016-51 62100480  00-5964-016-51 62112032  00-5964-016-51 62113013  00-5964-016-51 62112983  00-5964-016-51 62112982  00-5964-016-51 62115992  00-5964-016-51 62117311  00-5964-016-51 62117313  00-5964-016-51 62117314  00-5964-016-51 62117315  00-5964-016-51 62120595  00-5964-016-51 62124289  00-5964-016-51 62124759  00-5964-016-51 62125260  00-5964-016-51 62131291  00-5964-016-51 62132080  00-5964-016-51 62132098  00-5964-016-51 62132101  00-5964-016-51 62137510  00-5964-016-51 62143614  00-5964-016-51 62143654  00-5964-016-51 62143734  00-5964-016-51 62150066  00-5964-016-51 62150068  00-5964-016-51 62150087  00-5964-016-51 62154069  00-5964-016-51 62154068  00-5964-016-51 62159325  00-5964-016-51 62154039  00-5964-016-51 62165929  00-5964-016-51 62165930  00-5964-016-51 62166649  00-5964-016-51 62166647  00-5964-016-51 62170884  00-5964-016-51 62171085  00-5964-016-51 62171086  00-5964-016-51 62171730  00-5964-016-51 62170893  00-5964-016-51 62171726  00-5964-016-51 62170890  00-5964-016-52 61816395  00-5964-016-52 61816396  00-5964-016-52 61820650  00-5964-016-52 61820651  00-5964-016-52 61821125  00-5964-016-52 61823875  00-5964-016-52 61823877  00-5964-016-52 61830375  00-5964-016-52 61834855  00-5964-016-52 61834856  00-5964-016-52 61837315  00-5964-016-52 61843117  00-5964-016-52 61843119  00-5964-016-52 61844583  00-5964-016-52 61844584  00-5964-016-52 61855951  00-5964-016-52 61855952  00-5964-016-52 61857886  00-5964-016-52 61857887  00-5964-016-52 61857888  00-5964-016-52 61869844  00-5964-016-52 61869846  00-5964-016-52 61876909  00-5964-016-52 61876908  00-5964-016-52 61878268  00-5964-016-52 61883272  00-5964-016-52 61883271  00-5964-016-52 61884190  00-5964-016-52 61884192  00-5964-016-52 61885768  00-5964-016-52 61888047  00-5964-016-52 61888048  00-5964-016-52 61895303  00-5964-016-52 61895304  00-5964-016-52 61895305  00-5964-016-52 61901215  00-5964-016-52 61902943  00-5964-016-52 61902944  00-5964-016-52 61901214  00-5964-016-52 61902945  00-5964-016-52 61904191  00-5964-016-52 61902947  00-5964-016-52 61902946  00-5964-016-52 61904188  00-5964-016-52 61904189  00-5964-016-52 61904190  00-5964-016-52 61915298  00-5964-016-52 61915301  00-5964-016-52 61915302  00-5964-016-52 61922249  00-5964-016-52 61922250  00-5964-016-52 61923561  00-5964-016-52 61927959  00-5964-016-52 61932281  00-5964-016-52 61937792  00-5964-016-52 61940950  00-5964-016-52 61943181  00-5964-016-52 61943182  00-5964-016-52 61943183  00-5964-016-52 61953086  00-5964-016-52 61952877  00-5964-016-52 61953089  00-5964-016-52 61953093  00-5964-016-52 61953096  00-5964-016-52 61953095  00-5964-016-52 61960708  00-5964-016-52 61962879  00-5964-016-52 61962876  00-5964-016-52 61965932  00-5964-016-52 61965933  00-5964-016-52 61971720  00-5964-016-52 61971721  00-5964-016-52 61971723  00-5964-016-52 61980301  00-5964-016-52 61980302  00-5964-016-52 61980569  00-5964-016-52 61980570  00-5964-016-52 61980571  00-5964-016-52 61980572  00-5964-016-52 61982291  00-5964-016-52 61989035  00-5964-016-52 61989665  00-5964-016-52 61996344  00-5964-016-52 61996345  00-5964-016-52 61997334  00-5964-016-52 61997336  00-5964-016-52 61997337  00-5964-016-52 61998614  00-5964-016-52 61998615  00-5964-016-52 62006438  00-5964-016-52 62006449  00-5964-016-52 62006465  00-5964-016-52 62014083  00-5964-016-52 62014084  00-5964-016-52 62014396  00-5964-016-52 62014397  00-5964-016-52 62018587  00-5964-016-52 62023296  00-5964-016-52 62023798  00-5964-016-52 62023799  00-5964-016-52 62023797  00-5964-016-52 62024561  00-5964-016-52 62024570  00-5964-016-52 62024960  00-5964-016-52 62024961  00-5964-016-52 62033086  00-5964-016-52 62034601  00-5964-016-52 62039091  00-5964-016-52 62044629  00-5964-016-52 62044630  00-5964-016-52 62046964  00-5964-016-52 62048290  00-5964-016-52 62048291  00-5964-016-52 11006368  00-5964-016-52 62056792  00-5964-016-52 62057299  00-5964-016-52 62059398  00-5964-016-52 62059372  00-5964-016-52 62057306  00-5964-016-52 62058232  00-5964-016-52 62067920  00-5964-016-52 62072206  00-5964-016-52 62072511  00-5964-016-52 62077288  00-5964-016-52 62078816  00-5964-016-52 62078817  00-5964-016-52 62083121  00-5964-016-52 62083123  00-5964-016-52 62083124  00-5964-016-52 62083125  00-5964-016-52 62098667  00-5964-016-52 62098670  00-5964-016-52 62098280  00-5964-016-52 62098786  00-5964-016-52 62098787  00-5964-016-52 62098788  00-5964-016-52 62098789  00-5964-016-52 62102055  00-5964-016-52 62102056  00-5964-016-52 62102057  00-5964-016-52 62102712  00-5964-016-52 62102714  00-5964-016-52 62114023  00-5964-016-52 62114334  00-5964-016-52 62114346  00-5964-016-52 62115228  00-5964-016-52 62119534  00-5964-016-52 62119151  00-5964-016-52 62119152  00-5964-016-52 62121772  00-5964-016-52 62121794  00-5964-016-52 62126683  00-5964-016-52 62127080  00-5964-016-52 62132450  00-5964-016-52 62132454  00-5964-016-52 62132862  00-5964-016-52 62132876  00-5964-016-52 62138238  00-5964-016-52 62143738  00-5964-016-52 62144287  00-5964-016-52 62145398  00-5964-016-52 62146935  00-5964-016-52 62146956  00-5964-016-52 62148238  00-5964-016-52 62148258  00-5964-016-52 62154272  00-5964-016-52 62154957  00-5964-016-52 62154958  00-5964-016-52 62156819  00-5964-016-52 62157853  00-5964-016-52 62166739  00-5964-016-52 62167493  00-5964-016-52 62171785  00-5964-016-52 62171837  00-5964-016-52 62171838  00-5964-016-52 62171839  00-5964-016-52 62171840  00-5964-016-52 62172413  00-5964-017-51 61823882  00-5964-017-51 61823883  00-5964-017-51 61829831  00-5964-017-51 61842343  00-5964-017-51 61844892  00-5964-017-51 61861346  00-5964-017-51 61865905  00-5964-017-51 61874455  00-5964-017-51 61874456  00-5964-017-51 61881661  00-5964-017-51 61881663  00-5964-017-51 61886260  00-5964-017-51 61886261  00-5964-017-51 61886262  00-5964-017-51 61892362  00-5964-017-51 61893239  00-5964-017-51 61898487  00-5964-017-51 61898485  00-5964-017-51 61905152  00-5964-017-51 61905155  00-5964-017-51 61905169  00-5964-017-51 61909456  00-5964-017-51 61910862  00-5964-017-51 61912346  00-5964-017-51 61912347  00-5964-017-51 61918718  00-5964-017-51 61923565  00-5964-017-51 61938744  00-5964-017-51 61939177  00-5964-017-51 61939178  00-5964-017-51 61949697  00-5964-017-51 61943517  00-5964-017-51 61966236  00-5964-017-51 61966237  00-5964-017-51 61970679  00-5964-017-51 61971153  00-5964-017-51 61976411  00-5964-017-51 61977546  00-5964-017-51 61982454  00-5964-017-51 61988126  00-5964-017-51 61990911  00-5964-017-51 61990912  00-5964-017-51 61992052  00-5964-017-51 61990913  00-5964-017-51 62000967  00-5964-017-51 62000968  00-5964-017-51 62006726  00-5964-017-51 62012394  00-5964-017-51 62012654  00-5964-017-51 62017256  00-5964-017-51 62018139  00-5964-017-51 62018140  00-5964-017-51 62028013  00-5964-017-51 62029429  00-5964-017-51 62035837  00-5964-017-51 62042284  00-5964-017-51 62043132  00-5964-017-51 62044610  00-5964-017-51 62054831  00-5964-017-51 62058757  00-5964-017-51 62059371  00-5964-017-51 62068311  00-5964-017-51 62073097  00-5964-017-51 62074152  00-5964-017-51 62085944  00-5964-017-51 62085945  00-5964-017-51 62086606  00-5964-017-51 62096003  00-5964-017-51 62096697  00-5964-017-51 62092518  00-5964-017-51 62093708  00-5964-017-51 62107981  00-5964-017-51 62115243  00-5964-017-51 62115255  00-5964-017-51 62115256  00-5964-017-51 62117299  00-5964-017-51 62117301  00-5964-017-51 62122655  00-5964-017-51 62126686  00-5964-017-51 62131222  00-5964-017-51 62131246  00-5964-017-51 62132055  00-5964-017-51 62141170  00-5964-017-51 62142287  00-5964-017-51 62148755  00-5964-017-51 62154269  00-5964-017-51 62154994  00-5964-017-51 62159252  00-5964-017-51 62168117  00-5964-017-51 62169033  00-5964-017-51 62169034  00-5964-017-51 62170490  00-5964-017-51 62170491  00-5964-017-51 62170492  00-5964-017-52 61807817  00-5964-017-52 61820636  00-5964-017-52 61823880  00-5964-017-52 61822968  00-5964-017-52 61830363  00-5964-017-52 61843123  00-5964-017-52 61844893  00-5964-017-52 61844931  00-5964-017-52 61860211  00-5964-017-52 61866703  00-5964-017-52 61875434  00-5964-017-52 61875435  00-5964-017-52 61883261  00-5964-017-52 61883262  00-5964-017-52 61886317  00-5964-017-52 61886318  00-5964-017-52 61892359  00-5964-017-52 61892360  00-5964-017-52 61899617  00-5964-017-52 61901208  00-5964-017-52 61899618  00-5964-017-52 61899616  00-5964-017-52 61905263  00-5964-017-52 61910186  00-5964-017-52 61909460  00-5964-017-52 61910185  00-5964-017-52 61910189  00-5964-017-52 61910193  00-5964-017-52 61924441  00-5964-017-52 61935976  00-5964-017-52 61935978  00-5964-017-52 61939795  00-5964-017-52 61939181  00-5964-017-52 61946486  00-5964-017-52 61949695  00-5964-017-52 11400311  00-5964-017-52 61954543  00-5964-017-52 61955558  00-5964-017-52 61966807  00-5964-017-52 61966806  00-5964-017-52 61966808  00-5964-017-52 61978256  00-5964-017-52 61978264  00-5964-017-52 61982458  00-5964-017-52 61986695  00-5964-017-52 61986416  00-5964-017-52 11400341  00-5964-017-52 61992981  00-5964-017-52 61993919  00-5964-017-52 61993921  00-5964-017-52 61998799  00-5964-017-52 61998800  00-5964-017-52 62003765  00-5964-017-52 62004792  00-5964-017-52 62004794  00-5964-017-52 62004993  00-5964-017-52 62014073  00-5964-017-52 62014074  00-5964-017-52 62014387  00-5964-017-52 62019486  00-5964-017-52 62019487  00-5964-017-52 62019489  00-5964-017-52 62029441  00-5964-017-52 62029446  00-5964-017-52 62038010  00-5964-017-52 62043281  00-5964-017-52 62043283  00-5964-017-52 62046217  00-5964-017-52 62055031  00-5964-017-52 62058211  00-5964-017-52 62055523  00-5964-017-52 62068306  00-5964-017-52 62076154  00-5964-017-52 62074167  00-5964-017-52 62086981  00-5964-017-52 62086983  00-5964-017-52 62098049  00-5964-017-52 62098051  00-5964-017-52 62098278  00-5964-017-52 62106733  00-5964-017-52 62106732  00-5964-017-52 62115851  00-5964-017-52 62115852  00-5964-017-52 62117296  00-5964-017-52 62117297  00-5964-017-52 62122860  00-5964-017-52 62127394  00-5964-017-52 62132108  00-5964-017-52 62132078  00-5964-017-52 62138255  00-5964-017-52 62140623  00-5964-017-52 62140628  00-5964-017-52 62149037  00-5964-017-52 62149365  00-5964-017-52 62155191  00-5964-017-52 62156806  00-5964-017-52 62159249  00-5964-017-52 62168118  00-5964-017-52 62169043  00-5964-017-52 62169042  00-5964-017-52 62170028  00-5964-017-52 62170029  00-5972-015-01 61777593  00-5972-018-02 4002192  00-5970-012-02 61878272  00-5970-013-02 61777576  00-5970-013-02 62135909  00-5970-014-02 11400192  00-5970-014-02 62122787  00-5970-014-02 62135920  00-5970-014-02 62132399  00-5970-014-02 62132399R  00-5970-014-02 62135921  00-5970-015-02 11002229  00-5970-015-02 61777581  00-5970-015-02 62135936  00-5970-016-02 61777583  00-5970-016-02 62128113  00-5970-017-01 11003049  00-5970-018-01 61758971  00-5970-018-01 11003323  00-5970-018-02 11003330  00-5970-018-02 11003322  00-5956-013-01 61782485  00-5956-013-01 62075014  00-5956-013-01 62056264  00-5956-013-01 62117143  00-5956-013-02 62086765  00-5956-013-02 62117144  00-5956-013-06 62124799  00-5956-014-01 61844569  00-5956-014-01 61850921  00-5956-014-01 61855959  00-5956-014-01 61881040  00-5956-014-01 61972166  00-5956-014-01 61980310  00-5956-014-01 61985025  00-5956-014-01 62034470  00-5956-014-01 62094382  00-5956-014-01 62107136  00-5956-014-01 62118576  00-5956-014-01 62125123  00-5956-014-01 62160791  00-5956-014-05 62166743  00-5956-014-02 61835926  00-5956-014-02 61839709  00-5956-014-02 61869873  00-5956-014-02 61881043  00-5956-014-02 61906386  00-5956-014-02 62023791  00-5956-014-02 62067898  00-5956-014-02 62094383  00-5956-014-02 62122981  00-5956-014-02 62154041  00-5956-014-02 62161213  00-5956-014-06 62126727  00-5956-014-06 62137069  00-5956-015-01 61803565  00-5956-015-01 61821415  00-5956-015-01 61821416  00-5956-015-01 61826928  00-5956-015-01 11400132  00-5956-015-01 61835330  00-5956-015-01 61854818  00-5956-015-01 61858473  00-5956-015-01 61867606  00-5956-015-01 61869874  00-5956-015-01 61876279  00-5956-015-01 61895313  00-5956-015-01 61915309  00-5956-015-01 61926128  00-5956-015-01 61934960  00-5956-015-01 11005209  00-5956-015-01 61958022  00-5956-015-01 61962868  00-5956-015-01 61972737  00-5956-015-01 61982306  00-5956-015-01 61986702  00-5956-015-01 61986420  00-5956-015-01 62024955  00-5956-015-01 62046227  00-5956-015-01 62076994  00-5956-015-01 62076999  00-5956-015-01 62084437",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "LPS-FLEX OPTION FEMORAL C-L  LPS-FLEX OPTION FEMORAL C-L  LPS-FLEX OPTION FEMORAL C-R  LPS-FLEX OPTION FEMORAL D-L  LPS-FLEX OPTION FEMORAL D-R  LPS-FLEX OPTION FEMORAL E-L  LPS-FLEX OPTION FEMORAL E-R  LPS-FLEX OPTION FEMORAL F-L  LPS-FLEX OPTION FEMORAL F-R  LPS-FLEX OPTION FEMORAL G-L  LPS-FLEX OPTION FEMORAL G-R    This device is indicated for patients with severe knee pain and disability",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1043-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "00-5956-015-01 62109654  00-5956-015-01 62119541  00-5956-015-01 62127087  00-5956-015-01 62110605  00-5956-015-01 62138241  00-5956-015-01 62144132  00-5956-015-01 62152926  00-5956-015-01 62162199  00-5956-015-01 62172418  00-5956-015-05 62091540  00-5956-015-05 62119542  00-5956-015-05 62122982  00-5956-015-05 62127088  00-5956-015-05 62138258  00-5956-015-02 61803567  00-5956-015-02 61822947  00-5956-015-02 61822948  00-5956-015-02 11400116  00-5956-015-02 61835393  00-5956-015-02 61842348  00-5956-015-02 61849045  00-5956-015-02 61859222  00-5956-015-02 61864878  00-5956-015-02 61864879  00-5956-015-02 61869876  00-5956-015-02 61876938  00-5956-015-02 61897287  00-5956-015-02 61901218  00-5956-015-02 61916147  00-5956-015-02 61927965  00-5956-015-02 61929400  00-5956-015-02 61963686  00-5956-015-02 61973239  00-5956-015-02 61973241  00-5956-015-02 61982462  00-5956-015-02 61982742  00-5956-015-02 61988148  00-5956-015-02 62002066  00-5956-015-02 62012658  00-5956-015-02 62022038  00-5956-015-02 62028028  00-5956-015-02 62047517  00-5956-015-02 62072210  00-5956-015-02 62078811  00-5956-015-02 62085937  00-5956-015-02 62098284  00-5956-015-02 62110230  00-5956-015-02 62127119  00-5956-015-02 62127468  00-5956-015-02 62144290  00-5956-015-02 62154043  00-5956-015-02 62162764  00-5956-015-02 62169024  00-5956-015-02 62171732  00-5956-015-06 62085936  00-5956-015-06 62127118  00-5956-016-01 61810703  00-5956-016-01 61826933  00-5956-016-01 61833599  00-5956-016-01 61864881  00-5956-016-01 61895864  00-5956-016-01 61907448  00-5956-016-01 61915313  00-5956-016-01 61932289  00-5956-016-01 61932291  00-5956-016-01 61971163  00-5956-016-01 61978525  00-5956-016-01 61986178  00-5956-016-01 61989682  00-5956-016-01 62030345  00-5956-016-01 62030593  00-5956-016-01 62054911  00-5956-016-01 62083126  00-5956-016-01 62083127  00-5956-016-01 62109161  00-5956-016-01 62114098  00-5956-016-01 62126728  00-5956-016-01 62117146  00-5956-016-01 62154274  00-5956-016-01 62163103  00-5956-016-05 62121616  00-5956-016-05 62127085  00-5956-016-05 62138681  00-5956-016-05 62162202  00-5956-016-05 62170903  00-5956-016-02 61811794  00-5956-016-02 61817755  00-5956-016-02 61827940  00-5956-016-02 61865912  00-5956-016-02 61873444  00-5956-016-02 61880101  00-5956-016-02 61894398  00-5956-016-02 61910871  00-5956-016-02 61921452  00-5956-016-02 61964438  00-5956-016-02 61978526  00-5956-016-02 61986709  00-5956-016-02 61986713  00-5956-016-02 62026095  00-5956-016-02 62043310  00-5956-016-02 62077294  00-5956-016-02 62083130  00-5956-016-02 62100468  00-5956-016-02 62107984  00-5956-016-02 62115961  00-5956-016-02 62126729  00-5956-016-02 62117147  00-5956-016-02 62145821  00-5956-016-02 62156807  00-5956-016-02 62162770  00-5956-016-06 62075017  00-5956-016-06 62107985  00-5956-016-06 62127086  00-5956-016-06 62169047  00-5956-017-01 61797740  00-5956-017-01 61813670  00-5956-017-01 61814765  00-5956-017-01 61837325  00-5956-017-01 61876918  00-5956-017-01 61903674  00-5956-017-01 61918975  00-5956-017-01 61946490  00-5956-017-01 61971164  00-5956-017-01 61973245  00-5956-017-01 61996944  00-5956-017-01 62018673  00-5956-017-01 62052670  00-5956-017-01 62078825  00-5956-017-01 62084440  00-5956-017-01 62086771  00-5956-017-01 62112977  00-5956-017-01 62125785  00-5956-017-01 62143177  00-5956-017-01 62160337  00-5956-017-05 62126102  00-5956-017-05 62134747  00-5956-017-02 61803569  00-5956-017-02 61815712  00-5956-017-02 61855965  00-5956-017-02 61885778  00-5956-017-02 61901226  00-5956-017-02 61929404  00-5956-017-02 61971165  00-5956-017-02 61991756  00-5956-017-02 62007004  00-5956-017-02 62030636  00-5956-017-02 62051197  00-5956-017-02 62073095  00-5956-017-02 62096010  00-5956-017-02 62114025  00-5956-017-02 62126736  00-5956-017-02 62117149  00-5956-017-02 62143640  00-5956-017-02 62161216  00-5956-017-02 62170057  00-5956-017-06 62079857  00-5956-017-06 62160793  00-5956-017-06 62169051  00-5990-014-02 61776963  00-5990-015-02 11003324  00-5990-015-02 61776965  00-5990-017-91 61771841  00-5964-013-01 77001358  00-5964-013-51 61808871  00-5964-013-51 61876900  00-5964-013-51 61890747  00-5964-013-51 61890748  00-5964-013-51 61990909  00-5964-013-51 62009298  00-5964-013-51 62018048  00-5964-013-51 62080785  00-5964-013-51 62120585  00-5964-013-51 62160331  00-5964-013-51 62170030  00-5964-013-52 61858467  00-5964-013-52 61876901  00-5964-013-52 61876902  00-5964-013-52 61992032  00-5964-013-52 61999924  00-5964-013-52 62010044  00-5964-013-52 11400373  00-5964-013-52 62083118  00-5964-013-52 62107109  00-5964-013-52 62131121  00-5964-013-52 62170031  00-5964-014-51 61796364  00-5964-014-51 61794708  00-5964-014-51 61804727  00-5964-014-51 61809705  00-5964-014-51 61809745  00-5964-014-51 61810398  00-5964-014-51 61809707  00-5964-014-51 61813688  00-5964-014-51 61813689  00-5964-014-51 61826922  00-5964-014-51 61826923  00-5964-014-51 61827924  00-5964-014-51 61827925  00-5964-014-51 61827926  00-5964-014-51 61827927  00-5964-014-51 61829824  00-5964-014-51 61839693  00-5964-014-51 61839694  00-5964-014-51 61840456  00-5964-014-51 61846248  00-5964-014-51 61846249  00-5964-014-51 61847652  00-5964-014-51 61859212  00-5964-014-51 61859214  00-5964-014-51 61859215  00-5964-014-51 61866692  00-5964-014-51 61873428  00-5964-014-51 61876904  00-5964-014-51 61876905  00-5964-014-51 61884182  00-5964-014-51 61885169  00-5964-014-51 61886259  00-5964-014-51 61886258  00-5964-014-51 61890011  00-5964-014-51 61890010  00-5964-014-51 61890690  00-5964-014-51 61905149  00-5964-014-51 61905256  00-5964-014-51 61914474  00-5964-014-51 61914475  00-5964-014-51 61915292  00-5964-014-51 61914478  00-5964-014-51 61914479  00-5964-014-51 61915293  00-5964-014-51 61918961  00-5964-014-51 61930141  00-5964-014-51 61931654  00-5964-014-51 61939789  00-5964-014-51 61951646  00-5964-014-51 61951648  00-5964-014-51 61951105  00-5964-014-51 61951647  00-5964-014-51 61964434  00-5964-014-51 61965568  00-5964-014-51 61965569  00-5964-014-51 61965928  00-5964-014-51 61965927  00-5964-014-51 61968882  00-5964-014-51 61972732  00-5964-014-51 61978248  00-5964-014-51 61986694  00-5964-014-51 61987055  00-5964-014-51 61987054  00-5964-014-51 61988120  00-5964-014-51 61988444  00-5964-014-51 61992034  00-5964-014-51 61992051  00-5964-014-51 61992698  00-5964-014-51 61992699  00-5964-014-51 61992701  00-5964-014-51 61999966  00-5964-014-51 61999969  00-5964-014-51 62000965  00-5964-014-51 62010054  00-5964-014-51 62010322  00-5964-014-51 62016674  00-5964-014-51 62016669  00-5964-014-51 62023783  00-5964-014-51 62023784  00-5964-014-51 62024952  00-5964-014-51 62026045  00-5964-014-51 62026044  00-5964-014-51 62026089  00-5964-014-51 62034568  00-5964-014-51 62047521  00-5964-014-51 62051185  00-5964-014-51 62051186  00-5964-014-51 62051187  00-5964-014-51 62074155  00-5964-014-51 62077286  00-5964-014-51 62077837  00-5964-014-51 62079851  00-5964-014-51 62084425  00-5964-014-51 62088867  00-5964-014-51 62089663  00-5964-014-51 62096004  00-5964-014-51 62096006  00-5964-014-51 62096005  00-5964-014-51 62100461  00-5964-014-51 62100460  00-5964-014-51 62102053  00-5964-014-51 62102695  00-5964-014-51 62109151  00-5964-014-51 62109649  00-5964-014-51 62109650  00-5964-014-51 62112541  00-5964-014-51 62112972  00-5964-014-51 62112685  00-5964-014-51 62118561  00-5964-014-51 62118562  00-5964-014-51 62124294  00-5964-014-51 62130970  00-5964-014-51 62130973  00-5964-014-51 62130974  00-5964-014-51 62134259  00-5964-014-51 62134260  00-5964-014-51 62141164  00-5964-014-51 62141723  00-5964-014-51 62142299  00-5964-014-51 62146620  00-5964-014-51 62151772  00-5964-014-51 62152904  00-5964-014-51 62152897  00-5964-014-51 62152639  00-5964-014-51 62161209  00-5964-014-51 62163072  00-5964-014-51 62163099  00-5964-014-51 62171782  00-5964-014-51 62171783  00-5964-014-51 62171833  00-5964-014-51 62171834  00-5964-014-51 62170387  00-5964-014-51 62171766  00-5964-014-51 62171767",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}