  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76048",
      "recalling_firm": "The Binding Site Group, Ltd.",
      "address_1": "8 Calthorpe Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA.    Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.",
      "recall_number": "Z-1043-2017",
      "product_description": "Optilite Freelite Lambda Free Kit  Product Code: LK018.OPT, LK018.OPT.A    Protein testing kit.",
      "product_quantity": "809 kits",
      "reason_for_recall": "Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.",
      "recall_initiation_date": "20160705",
      "center_classification_date": "20170118",
      "termination_date": "20170703",
      "report_date": "20170125",
      "code_info": "Lot #405566, 405567",
      "more_code_info": ""
    }
  ]
}