{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5966-013-01 61893236  00-5966-013-01 61939720  00-5966-013-01 61976414  00-5966-013-01 61997338  00-5966-013-01 61997339  00-5966-013-01 62164866  00-5966-013-01 62173181  00-5966-013-02 61893622  00-5966-013-02 61939721  00-5966-013-02 61976415  00-5966-013-02 61997340  00-5966-013-02 61998618  00-5966-013-02 62142586  00-5966-013-02 62172419  00-5966-014-01 61782655  00-5966-014-01 61850922  00-5966-014-01 61855960  00-5966-014-01 61873439  00-5966-014-01 61883263  00-5966-014-01 61883264  00-5966-014-01 61886303  00-5966-014-01 61890020  00-5966-014-01 61890021  00-5966-014-01 61890694  00-5966-014-01 61896063  00-5966-014-01 61916149  00-5966-014-01 61918723  00-5966-014-01 61918967  00-5966-014-01 61924446  00-5966-014-01 61933809  00-5966-014-01 61943608  00-5966-014-01 61946984  00-5966-014-01 11400307  00-5966-014-01 61971724  00-5966-014-01 61971725  00-5966-014-01 61974122  00-5966-014-01 61974123  00-5966-014-01 61978521  00-5966-014-01 61978522  00-5966-014-01 61989037  00-5966-014-01 61996346  00-5966-014-01 61996347  00-5966-014-01 62002067  00-5966-014-01 62002068  00-5966-014-01 61988452  00-5966-014-01 62022322  00-5966-014-01 62033088  00-5966-014-01 62038014  00-5966-014-01 62047986  00-5966-014-01 62047985  00-5966-014-01 62067952  00-5966-014-01 62073104  00-5966-014-01 62080815  00-5966-014-01 62085948  00-5966-014-01 62092287  00-5966-014-01 62110232  00-5966-014-01 62110233  00-5966-014-01 62119544  00-5966-014-01 62123030  00-5966-014-01 62133158  00-5966-014-01 62139972  00-5966-014-01 62139995  00-5966-014-01 62142356  00-5966-014-01 62152617  00-5966-014-01 62152618  00-5966-014-01 62164318  00-5966-014-01 62173182  00-5966-014-02 61851720  00-5966-014-02 61874459  00-5966-014-02 61879226  00-5966-014-02 61883265  00-5966-014-02 61883266  00-5966-014-02 61893232  00-5966-014-02 61890696  00-5966-014-02 61893234  00-5966-014-02 61902949  00-5966-014-02 61916150  00-5966-014-02 61918968  00-5966-014-02 61918969  00-5966-014-02 61924447  00-5966-014-02 61933810  00-5966-014-02 61937797  00-5966-014-02 61947190  00-5966-014-02 61947191  00-5966-014-02 61971726  00-5966-014-02 61971727  00-5966-014-02 61971728  00-5966-014-02 61976416  00-5966-014-02 61977294  00-5966-014-02 61978523  00-5966-014-02 61979109  00-5966-014-02 61979116  00-5966-014-02 61989039  00-5966-014-02 61996348  00-5966-014-02 61996349  00-5966-014-02 62003769  00-5966-014-02 62003768  00-5966-014-02 62022323  00-5966-014-02 62028039  00-5966-014-02 62028045  00-5966-014-02 62033089  00-5966-014-02 62038015  00-5966-014-02 62047525  00-5966-014-02 62073105  00-5966-014-02 62081426  00-5966-014-02 62081425  00-5966-014-02 62085949  00-5966-014-02 62098790  00-5966-014-02 62105729  00-5966-014-02 62110794  00-5966-014-02 62120689  00-5966-014-02 62123033  00-5966-014-02 62133828  00-5966-014-02 62140115  00-5966-014-02 62141175  00-5966-014-02 62147815  00-5966-014-02 62152642  00-5966-014-02 62164670  00-5966-014-02 62164672  00-5966-014-02 62164822  00-5966-015-01 61788743  00-5966-015-01 61851725  00-5966-015-01 61851726  00-5966-015-01 61865917  00-5966-015-01 61874466  00-5966-015-01 61881044  00-5966-015-01 61881045  00-5966-015-01 61885779  00-5966-015-01 61886270  00-5966-015-01 61899631  00-5966-015-01 61899632  00-5966-015-01 61901222  00-5966-015-01 61907445  00-5966-015-01 61907446  00-5966-015-01 61907447  00-5966-015-01 61908303  00-5966-015-01 61908304  00-5966-015-01 61909335  00-5966-015-01 11400253  00-5966-015-01 61917749  00-5966-015-01 61918721  00-5966-015-01 61917750  00-5966-015-01 61918722  00-5966-015-01 61918970  00-5966-015-01 61917751  00-5966-015-01 61918972  00-5966-015-01 61918971  00-5966-015-01 61920436  00-5966-015-01 61924448  00-5966-015-01 61924449  00-5966-015-01 61937799  00-5966-015-01 11400295  00-5966-015-01 61942841  00-5966-015-01 61943188  00-5966-015-01 61946980  00-5966-015-01 61946975  00-5966-015-01 61964947  00-5966-015-01 61965576  00-5966-015-01 61964949  00-5966-015-01 61977306  00-5966-015-01 61977316  00-5966-015-01 61979126  00-5966-015-01 61979130  00-5966-015-01 61979774  00-5966-015-01 61979776  00-5966-015-01 11400336  00-5966-015-01 61987062  00-5966-015-01 61987063  00-5966-015-01 61996942  00-5966-015-01 61996943  00-5966-015-01 61998627  00-5966-015-01 61998629  00-5966-015-01 61998630  00-5966-015-01 62003109  00-5966-015-01 62003113  00-5966-015-01 62008742  00-5966-015-01 62008743  00-5966-015-01 62020808  00-5966-015-01 62033090  00-5966-015-01 62034603  00-5966-015-01 62034604  00-5966-015-01 62038575  00-5966-015-01 62038808  00-5966-015-01 62038572  00-5966-015-01 62040757  00-5966-015-01 62048441  00-5966-015-01 62048442  00-5966-015-01 62066772  00-5966-015-01 62074210  00-5966-015-01 62079856  00-5966-015-01 62080306  00-5966-015-01 62086629  00-5966-015-01 62086630  00-5966-015-01 62092284  00-5966-015-01 62092524  00-5966-015-01 62098791  00-5966-015-01 62098792  00-5966-015-01 62105730  00-5966-015-01 62110795  00-5966-015-01 62110809  00-5966-015-01 62110810  00-5966-015-01 62117283  00-5966-015-01 62123034  00-5966-015-01 62128363  00-5966-015-01 62128375  00-5966-015-01 62131027  00-5966-015-01 62137079  00-5966-015-01 62146937  00-5966-015-01 62155683  00-5966-015-01 62159281  00-5966-015-01 62156809  00-5966-015-01 62164823  00-5966-015-01 62171734  00-5966-015-01 62171789  00-5966-015-01 62174412  00-5966-015-01 62174413  00-5966-015-02 61782660  00-5966-015-02 61851727  00-5966-015-02 61851728  00-5966-015-02 61861361  00-5966-015-02 61865915  00-5966-015-02 61865916  00-5966-015-02 61870728  00-5966-015-02 61874467  00-5966-015-02 61879228  00-5966-015-02 61881671  00-5966-015-02 61881672  00-5966-015-02 61886323  00-5966-015-02 61886335  00-5966-015-02 61901223  00-5966-015-02 61901224  00-5966-015-02 61901225  00-5966-015-02 61902951  00-5966-015-02 61902952  00-5966-015-02 61902953  00-5966-015-02 61908284  00-5966-015-02 61908285  00-5966-015-02 61908286  00-5966-015-02 61909339  00-5966-015-02 61909480  00-5966-015-02 11400252  00-5966-015-02 61917752  00-5966-015-02 61917753  00-5966-015-02 61917754  00-5966-015-02 61918719  00-5966-015-02 61918720  00-5966-015-02 61920437  00-5966-015-02 61920439  00-5966-015-02 61920438  00-5966-015-02 61920440  00-5966-015-02 61926126  00-5966-015-02 61924450  00-5966-015-02 61939184  00-5966-015-02 61939185  00-5966-015-02 61939182  00-5966-015-02 61946981  00-5966-015-02 61946982  00-5966-015-02 61961808  00-5966-015-02 61961809  00-5966-015-02 61962865  00-5966-015-02 61965577  00-5966-015-02 61965934  00-5966-015-02 61965935  00-5966-015-02 61977550  00-5966-015-02 61977551  00-5966-015-02 61977552  00-5966-015-02 61982743  00-5966-015-02 61982744  00-5966-015-02 61983003  00-5966-015-02 61983004  00-5966-015-02 61988191  00-5966-015-02 61988194  00-5966-015-02 61991759  00-5966-015-02 61991760  00-5966-015-02 11400352  00-5966-015-02 61998807  00-5966-015-02 61998808  00-5966-015-02 62003609  00-5966-015-02 62003615  00-5966-015-02 62008746  00-5966-015-02 62008747  00-5966-015-02 62008744  00-5966-015-02 62020809  00-5966-015-02 62028447  00-5966-015-02 62028080  00-5966-015-02 62028084  00-5966-015-02 62038814  00-5966-015-02 62040758  00-5966-015-02 62039938  00-5966-015-02 62040759  00-5966-015-02 62049389  00-5966-015-02 62049390  00-5966-015-02 62067349  00-5966-015-02 62067903  00-5966-015-02 62067350  00-5966-015-02 62074224  00-5966-015-02 62080308  00-5966-015-02 62080813  00-5966-015-02 62080814  00-5966-015-02 62080307  00-5966-015-02 62086632  00-5966-015-02 62086631  00-5966-015-02 62092525  00-5966-015-02 62092526  00-5966-015-02 62098968  00-5966-015-02 62099266  00-5966-015-02 62099267  00-5966-015-02 62105731  00-5966-015-02 62106566  00-5966-015-02 62105732  00-5966-015-02 62106560  00-5966-015-02 62117328  00-5966-015-02 62117329  00-5966-015-02 62129349  00-5966-015-02 62128376  00-5966-015-02 62131207  00-5966-015-02 62131208  00-5966-015-02 62137083  00-5966-015-02 62142595  00-5966-015-02 62146938  00-5966-015-02 62156810  00-5966-015-02 62157840  00-5966-015-02 62157958  00-5966-015-02 62157959  00-5966-015-02 62164865  00-5966-015-02 62171845  00-5966-015-02 62171846  00-5966-016-01 61782661  00-5966-016-01 61851729  00-5966-016-01 61881674  00-5966-016-01 61881675  00-5966-016-01 61886337  00-5966-016-01 61890515  00-5966-016-01 61892006  00-5966-016-01 61892007  00-5966-016-01 61895868  00-5966-016-01 61895869  00-5966-016-01 61899630  00-5966-016-01 61912359  00-5966-016-01 61912360  00-5966-016-01 61913193  00-5966-016-01 61918724  00-5966-016-01 61917760  00-5966-016-01 61924451  00-5966-016-01 61923577  00-5966-016-01 61930158  00-5966-016-01 61935980  00-5966-016-01 61944887  00-5966-016-01 61944886  00-5966-016-01 61962873  00-5966-016-01 61962874  00-5966-016-01 61966241  00-5966-016-01 61966242  00-5966-016-01 61972741  00-5966-016-01 61977556  00-5966-016-01 61977557  00-5966-016-01 61988453  00-5966-016-01 61991778  00-5966-016-01 61991776  00-5966-016-01 61999560  00-5966-016-01 61999559  00-5966-016-01 62014391  00-5966-016-01 62014392  00-5966-016-01 62025349  00-5966-016-01 62025350  00-5966-016-01 62025351  00-5966-016-01 62028448  00-5966-016-01 62029475  00-5966-016-01 62028449  00-5966-016-01 62039939  00-5966-016-01 62042302  00-5966-016-01 62046232  00-5966-016-01 62046231  00-5966-016-01 62046233  00-5966-016-01 62046890  00-5966-016-01 62046891  00-5966-016-01 62063280  00-5966-016-01 62067905  00-5966-016-01 11400415  00-5966-016-01 62086609  00-5966-016-01 62086608  00-5966-016-01 62098794  00-5966-016-01 62098795  00-5966-016-01 62106702  00-5966-016-01 62106701  00-5966-016-01 62117331  00-5966-016-01 62121775  00-5966-016-01 62127470  00-5966-016-01 62131252  00-5966-016-01 62137515  00-5966-016-01 62140660  00-5966-016-01 62151136  00-5966-016-01 62152287  00-5966-016-01 62159285  00-5966-016-02 61782663  00-5966-016-02 61788749  00-5966-016-02 61883268  00-5966-016-02 61884184  00-5966-016-02 61888037  00-5966-016-02 61892369  00-5966-016-02 61892371  00-5966-016-02 61895871  00-5966-016-02 61895875  00-5966-016-02 61895870  00-5966-016-02 61895872  00-5966-016-02 61899629  00-5966-016-02 61913194  00-5966-016-02 61918725  00-5966-016-02 61924453  00-5966-016-02 61924454  00-5966-016-02 61926130  00-5966-016-02 61924455  00-5966-016-02 61930159  00-5966-016-02 61944891  00-5966-016-02 61944890  00-5966-016-02 11400303  00-5966-016-02 61944889  00-5966-016-02 61946488  00-5966-016-02 61957853  00-5966-016-02 61959003  00-5966-016-02 61970683  00-5966-016-02 61970684  00-5966-016-02 61970685  00-5966-016-02 61972170  00-5966-016-02 61972171  00-5966-016-02 61972172  00-5966-016-02 61972740  00-5966-016-02 61980577  00-5966-016-02 61980578  00-5966-016-02 61980579  00-5966-016-02 61982463  00-5966-016-02 61988454  00-5966-016-02 61988455  00-5966-016-02 61992056  00-5966-016-02 61999561  00-5966-016-02 61992057  00-5966-016-02 62003771  00-5966-016-02 62003772  00-5966-016-02 62014393  00-5966-016-02 62014394  00-5966-016-02 62025176  00-5966-016-02 62025177  00-5966-016-02 62025348  00-5966-016-02 62028450  00-5966-016-02 62039940  00-5966-016-02 62042313  00-5966-016-02 62042323  00-5966-016-02 62043145  00-5966-016-02 62046216  00-5966-016-02 62046230  00-5966-016-02 62046215  00-5966-016-02 62059377  00-5966-016-02 62062230  00-5966-016-02 62067908  00-5966-016-02 62074267  00-5966-016-02 62074271  00-5966-016-02 62080786  00-5966-016-02 62080787  00-5966-016-02 62086610  00-5966-016-02 62086611  00-5966-016-02 62092527  00-5966-016-02 62093279  00-5966-016-02 62098796  00-5966-016-02 62098837  00-5966-016-02 62106708  00-5966-016-02 62106709  00-5966-016-02 62118583  00-5966-016-02 62121797  00-5966-016-02 62121798  00-5966-016-02 62137537  00-5966-016-02 62137517  00-5966-016-02 62140666  00-5966-016-02 62141726  00-5966-016-02 62142358  00-5966-016-02 62151140  00-5966-016-02 62152272  00-5966-016-02 62159329  00-5966-016-02 62168049  00-5966-017-01 61782664  00-5966-017-01 61866711  00-5966-017-01 61875442  00-5966-017-01 61884187  00-5966-017-01 61890027  00-5966-017-01 61890755  00-5966-017-01 61908292  00-5966-017-01 61909340  00-5966-017-01 61913196  00-5966-017-01 61918726  00-5966-017-01 61921453  00-5966-017-01 61921454  00-5966-017-01 61922262  00-5966-017-01 61931672  00-5966-017-01 61957854  00-5966-017-01 61962316  00-5966-017-01 61962317  00-5966-017-01 61962314  00-5966-017-01 61972742  00-5966-017-01 61978274  00-5966-017-01 61989042  00-5966-017-01 61988456  00-5966-017-01 61992058  00-5966-017-01 61992059  00-5966-017-01 11400356  00-5966-017-01 62000974  00-5966-017-01 62000975  00-5966-017-01 62017250  00-5966-017-01 62017261  00-5966-017-01 62027198  00-5966-017-01 62027199  00-5966-017-01 62029521  00-5966-017-01 62047527  00-5966-017-01 62080816  00-5966-017-01 62086633  00-5966-017-01 62086634  00-5966-017-01 62107198  00-5966-017-01 62107194  00-5966-017-01 62128377  00-5966-017-01 62131268  00-5966-017-01 62140649  00-5966-017-01 62140653  00-5966-017-01 62159330  00-5966-017-01 62172420  00-5966-017-01 62174409  00-5966-017-02 61782666  00-5966-017-02 61879329  00-5966-017-02 61884189  00-5966-017-02 61885172  00-5966-017-02 61890763  00-5966-017-02 61890765  00-5966-017-02 61909343  00-5966-017-02 61909344  00-5966-017-02 61913198  00-5966-017-02 61918727  00-5966-017-02 61923574  00-5966-017-02 61923575  00-5966-017-02 61923573  00-5966-017-02 61931673  00-5966-017-02 61936774  00-5966-017-02 61957855  00-5966-017-02 61962870  00-5966-017-02 61962871  00-5966-017-02 61962872  00-5966-017-02 61964945  00-5966-017-02 61964946  00-5966-017-02 61973242  00-5966-017-02 61989045  00-5966-017-02 62000976  00-5966-017-02 62002039  00-5966-017-02 62004998  00-5966-017-02 62004999  00-5966-017-02 62017262  00-5966-017-02 62017263  00-5966-017-02 11400382  00-5966-017-02 62028075  00-5966-017-02 62027200  00-5966-017-02 62029524  00-5966-017-02 62029523  00-5966-017-02 62040756  00-5966-017-02 11006417  00-5966-017-02 62046898  00-5966-017-02 62046900  00-5966-017-02 62067912  00-5966-017-02 62086636  00-5966-017-02 62107212  00-5966-017-02 62122565  00-5966-017-02 62131282  00-5966-017-02 62131283  00-5966-017-02 62137758  00-5966-017-02 62140645  00-5966-017-02 62152311",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "CR OPTION FEM COMP SIZE C/L  CR OPTION FEM COMP SIZE C/R  CR OPTION FEM COMP SIZE D/L  CR OPTION FEM COMP SIZE D/R  CR OPTION FEM COMP SIZE E/L  CR OPTION FEM COMP SIZE E/R  CR OPTION FEM COMP SIZE F/L  CR OPTION FEM COMP SIZE F/R  CR OPTION FEM COMP SIZE G/L  CR OPTION FEM COMP SIZE G/R    This device is indicated for patients with severe knee pain and disability",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1042-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}