{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1039-2014",
      "product_description": "COMPRESSION SCREW (1/2, 1-1/2, 1-1/4, 1-3/4, 1. 2. 3/4 IN)",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60102586  60132471  60142537  60145090  60208438  60210902  60218747  60265332  60316667  60380272  60387522  60425078  60451935  60491775  60504962  60538423  60569271  60629869  60629870  60671868  60690620  60735901  60790633  60828403  60859841  60885052  60910023  60926381  60938165  60997341  61008512  61045578  61049489  61063879  61085581  61087377  61120002  61160484  61160485  61179531  61276833  61441956  61457387  61513903  61626104  61660625  61664773  61677697  61677702  61705589  61707801  61729145  61729146  61729745  61750828  62109253  60049193  60096554  60101711  60102587  60102588  60123238  60130244  60139604  60140249  60140250  60141583  60142539  60145093  60165156  60165157  60169713  60169714  60175619  60177171  60193845  60195824  60195825  60218748  60222629  60222631  60231265  60231266  60240376  60249923  60249924  60257490  60257491  60265333  60275014  60275015  60297515  60307915  60307916  60316668  60316669  60322678  60330555  60332252  60338176  60370097  60380274  60391503  60391504  60393274  60393275  60401261  60405226  60471147  60491776  60491777  60514747  60536910  60536915  60538424  60550151  60557766  60563823  60585131  60591894  60610147  60629871  60629873  60629876  60639748  60639754  60646706  60652934  60652937  60671869  60671870  60698228  60714739  60735902  60735903  60756001  60790637  60794404  60843741  60868448  60885053  60885055  60926383  60930495  60932378  60938166  60938167  60938168  60944479  60997342  60997343  61008514  61012474  61045579  61063880  61063881  61063882  61085583  61120003  61120004  61140029  61140030  61140031  61160486  61160487  61160488  61179533  61206340  61233884  61245938  61262515  61276834  61276835  61296488  61305259  61313619  61316547  61354692  61354693  61354694  61359714  61359715  61420257  61425119  61425120  61457388  61461974  61468652  61523661  61584920  61620163  61634702  61657584  61660646  61707798  61707799  61713678  61713679  61729746  61729747  61740684  61750832  62118597  62304910  60044879  60086616  60086618  60096556  60097892  60097893  60097894  60101712  60101713  60101714  60102590  60102594  60102595  60105526  60105527  60108360  60108362  60109554  60109556  60109558  60111964  60111965  60112836  60114740  60123239  60123240  60123241  60123242  60123243  60124842  60124844  60126287  60130245  60132472  60132473  60132474  60132475  60135130  60135134  60139605  60139606  60140251  60141584  60141585  60142541  60142545  60142547  60142552  60145094  60145095  60145096  60145097  60149084  60149085  60154148  60154149  60154152  60155292  60165158  60165160  60165161  60165162  60165164  60165165  60165166  60165167  60169715  60169716  60169717  60172492  60172493  60175620  60175621  60179272  60179274  60186808  60190832  60190833  60193846  60195826  60195827  60195828  60195829  60195830  60208456  60208458  60208459  60208460  60208461  60208462  60210903  60210904  60210905  60210906  60218749  60218751  60218752  60218753  60222622  60222623  60222624  60222625  60222626  60222627  60231267  60231268  60239229  60239230  60240371  60240372  60240373  60240374  60240375  60244016  60244019  60244020  60244021  60244022  60244023  60244025  60245719  60249925  60249926  60249927  60253664  60253665  60253666  60253667  60253668  60257492  60257493  60257494  60265334  60265335  60265336  60265337  60265338  60265339  60265340  60265341  60275016  60275017  60275018  60275019  60297520  60297521  60297522  60297523  60297524  60297525  60307919  60307922  60307924  60307925  60307928  60307930  60307932  60307934  60307937  60316670  60316671  60318910  60318911  60318912  60322679  60322680  60322681  60322682  60322683  60322684  60330551  60330552  60330553  60332253  60332254  60332255  60338181  60338183  60338184  60338186  60349484  60349485  60364852  60370098  60370099  60380275  60380276  60380277  60380279  60391505  60391506  60391507  60391508  60393277  60393278  60396391  60396392  60396393  60401262  60401263  60401264  60405227  60405230  60405231  60406061  60406062  60425084  60425085  60425087  60425088  60431826  60431828  60431830  60435603  60439909  60439910  60439911  60439912  60439913  60439914  60439915  60439916  60439917  60448275  60448276  60448277  60448278  60451936  60451937  60451938  60451939  60460759  60460760  60460761  60460762  60460764  60465835  60465836  60465838  60465839  60471148  60471150  60471151  60471152  60478872  60478873  60478874  60478875  60478876  60504963  60504964  60504965  60504966  60525777  60525786  60525790  60555770  60557767  60557768  60557769  60557770  60563824  60563825  60566839  60566841  60566842  60585136  60585138  60585139  60585141  60585145  60591895  60591896  60591897  60610148  60610149  60610150  60610151  60610152  60610153  60610154  60610155  60639755  60639758  60639759  60639767  60639768  60639818  60646708  60646710  60646712  60646713  60646714  60646715  60652936  60652938  60652939  60652940  60652941  60652942  60652943  60671871  60671872  60671873  60671874  60671876  60671877  60671878  60698238  60698240  60698241  60698242  60698243  60714740  60714741  60714743  60724464  60724465  60724466  60724467  60735904  60735905  60756002  60756003  60756004  60756005  60756006  60756007  60790659  60790660  60790661  60794406  60794407  60794408  60794409  60797063  60797064  60797065  60797066  60797067  60828404  60828405  60836006  60836007  60836008  60885056  60885057  60885058  60910024  60910025  60910026  60910027  60910028  60910029  60910030  60910031  60926384  60926385  60926386  60926387  60926388  60930496  60930497  60938169  60938170  60944480  60944481  61008515  61008516  61008519  61012475  61012476  61012477  61012478  61012479  61012480  61012481  61012482  61012483  61019229  61019230  61019231  61019232  61019233  61020846  61045581  61045582  61045583  61063883  61085584  61085585  61085586  61085587  61085589  61087381  61087384  61087385  61087386  61120005  61120007  61120008  61120009  61120011  61129856  61129857  61140032  61140033  61140034  61140035  61140036  61160489  61160493  61160494  61160495  61160496  61166277  61179534  61184526  61184527  61184528  61200594  61200595  61200596  61206341  61206342  61206343  61224638  61224639  61224640  61224641  61233885  61245939  61245940  61256434  61262516  61262517  61262518  61269374  61274525  61282034  61282035  61282036  61282037  61305260  61313620  61316548  61354695  61354696  61354697  61354698  61354699  61354701  61354702  61354703  61359717  61359719  61359720  61359721  61359722  61359723  61378426  61378427  61380171  61382953  61382954  61395119  61396163  61420258  61420259  61425121  61439804  61439805  61441960  61461975  61461976  61465686  61468653  61468656  61468657  61513904  61513906  61513911  61516709  61523662  61523663  61523664  61538674  61580813  61580814  61580815  61584921  61612605  61612607  61612608  61612645  61620165  61620166  61620167  61620168  61620169  61620170  61640839  61640840  61640841  61649746  61649747  61650515  61657598  61664775  61713680  61713681  61729748  61729754  61750837  61750840  61750842  61750845  62036738  60096557  60108363  60112838  60114741  60132476  60141586  60142554  60145098  60149086  60155293  60165168  60165169  60175622  60179275  60193847  60195831  60195832  60208454  60210907  60210908  60210909  60222628  60231270  60231271  60253669  60253670  60257495  60265342  60265344  60275021  60297526  60297527  60297528  60297529  60307939  60307940  60316672  60316673  60318913  60318914  60322685  60322686  60330554  60332256  60338188  60370102  60405233  60451940  60465840  60471153  60514748  60525791  60550152  60557771  60563826  60566844  60591898  60610156  60629874  60639769  60646716  60652949  60671879  60671880  60698244  60714746  60752297  60756008  60756010  60756011  60828406  60836009  60843744  60885060  60910032  60926391  60930498  60938171  60944482  61063886  61063888  61067749  61085590  61087388  61120012  61140037  61140039  61160497  61184529  61206344  61224642  61233886  61256435  61282038  61296489  61359724  61382955  61441961  61457389  61484347  61516710  61523665  61584922  61612646  61634706  61640843  61653981  61852995  60102596  60109561  60130246  60132477  60135135  60142555  60149087  60165171  60172494  60195833  60208455  60218750  60222632  60231274  60231276  60265345  60297530  60297532  60316674  60380280  60393279  60405235  60550154  60557772  60566845  60591899  60652950  60698245  60724468  60768296  60790666  60836010  60910034  60926392  60932379  60938172  60997344  61008520  61120013  61120014  61160498  61160499  61185829  61269375  61420260  61484348  61490677  61523666  61612647  61634707  61653982  60115144  60201209  60210910  60218754  60265346  60307941  60504967  60836011  60938173  61120015  61156303  61184530  61185830  60139607  60210911  60367364  60504968  60714747  60790667  60828407  60890911  61012484  61120016  61140042  61184531  61185831",
      "more_code_info": ""
    }
  ]
}