  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1038-2014",
      "product_description": "HERBERT BONE SCREW, 3.0MM and HERBERT CANNULATED BONE S",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60117014  60126278  60155287  60178897  60208414  60240381  60265317  60469641  60667003  60757014  60818325  60868434  60877876  60953936  61057493  61083223  61144260  61382948  61428496  61434453  60123231  60145083  60172484  60210898  60235186  60240382  60318917  60367391  60451947  60491785  60589767  60665682  60665683  60757015  60810891  60926378  60930484  60953937  60953938  61045584  61109124  61135014  61144262  61428498  61437196  60096548  60109535  60111962  60126279  60132072  60135124  60142531  60152397  60165150  60169711  60196507  60208415  60218735  60235187  60244038  60247673  60291570  60303879  60308478  60308479  60323219  60429155  60435646  60451952  60472779  60487172  60491786  60501138  60501139  60603149  60629958  60652968  60665904  60714756  60714757  60735916  60757016  60810892  60818328  60953942  61008526  61012463  61012464  61012465  61012466  61045586  61045587  61045588  61144264  61428502  61437198  61490670  61754462  60096549  60101706  60109537  60112829  60123232  60126280  60132073  60140244  60141574  60152398  60155288  60172485  60190827  60194301  60195422  60222272  60231258  60247674  60257483  60257484  60270650  60274974  60279001  60291572  60291574  60323220  60336408  60343703  60429156  60435650  60451953  60472780  60491790  60501140  60501141  60529889  60544298  60557791  60589768  60652972  60671883  60683409  60714758  60735917  60810894  60810895  60818329  60818330  60818331  60944477  60990430  61001693  61045590  61045591  61045592  61080404  61083225  61117670  61135015  61144266  61273769  61357130  61374874  61382949  61382950  61434454  61473777  61523647  369406  60096550  60101707  60109539  60112830  60117016  60126281  60130233  60132074  60140245  60141576  60148204  60152399  60155289  60162144  60169712  60172486  60178898  60189031  60190828  60194302  60195425  60228106  60235188  60244039  60247675  60253689  60257485  60270651  60274975  60279002  60291576  60291577  60303880  60308480  60323221  60336409  60343704  60372743  60380525  60425090  60429157  60435653  60451954  60472781  60478986  60501142  60501143  60529890  60544299  60544301  60557795  60652974  60667006  60675136  60677483  60677485  60683412  60757018  60810896  60810897  60810898  60818422  60818432  60818433  60818436  60973690  61001694  61012468  61012469  61012471  61045593  61060985  61071292  61071293  61112939  61112941  61117671  61120678  61135016  61372549  61374875  61382951  61434455  61465684  61473778  61490671  61523648  61729781  60096551  60101708  60105518  60112833  60123233  60126282  60130235  60132075  60135125  60139609  60141578  60142532  60148219  60152401  60162145  60167101  60172487  60178899  60185147  60189032  60190829  60194303  60195426  60218736  60228107  60244040  60247677  60253691  60257486  60265318  60270652  60274976  60279003  60291578  60291580  60308482  60308483  60308484  60323222  60336410  60338194  60343707  60372744  60380527  60425091  60435656  60451955  60472782  60478987  60514756  60514762  60529891  60537023  60544306  60557798  60589771  60652975  60652987  60671884  60677486  60683413  60735918  60757019  60810901  60810903  60818439  60818440  60818444  60818445  60962899  61001695  61001696  61008024  61008026  61008289  61008528  61008529  61083226  61111266  61112942  61129854  61144267  61156090  61166262  61245936  61270324  61374876  61380168  61382952  61396162  61473779  61490673  61523649  61707786  61827155  62319566  60096552  60101709  60105520  60117018  60123234  60126283  60130238  60135126  60140246  60141579  60145084  60152402  60154144  60162146  60167102  60167103  60172488  60178901  60190830  60194304  60208416  60208418  60210899  60215602  60215603  60244041  60247224  60247678  60257487  60257488  60270653  60274977  60279005  60291581  60291582  60323223  60336411  60338198  60343711  60343714  60380528  60406055  60425092  60429159  60435659  60435661  60435662  60451956  60455019  60472783  60491793  60491798  60514758  60514759  60529892  60537024  60544307  60557799  60652988  60652989  60675137  60677489  60677491  60683415  60757020  60810904  60810905  60810906  60818446  60818447  60818448  60818449  60818450  60818451  60818452  60818453  61019226  61060994  61060999  61083227  61109125  61111267  61137388  61166265  61185818  61273771  61374877  61451147  61513892  61523650  61675714  61729785  60096553  60105521  60111963  60117021  60126284  60130239  60130241  60132469  60135127  60139610  60140247  60141580  60152403  60155290  60165152  60172489  60178902  60178903  60189033  60215604  60228108  60235189  60244043  60247679  60257489  60265319  60270654  60274979  60279006  60291584  60291585  60323224  60336412  60343718  60343721  60380529  60425093  60435663  60451957  60472784  60487173  60491801  60514760  60544308  60561939  60652991  60665685  60667009  60675138  60735919  60757021  60810907  60810909  60810910  60818462  60818463  60868436  60877878  60914237  60914238  61019227  61071296  61080406  61112943  61117672  61120679  61144269  61166267  61290776  61513893  61523651  61675715  61729787  60094830  60097890  60101710  60105522  60123235  60130242  60132470  60135128  60141581  60142534  60152405  60165153  60167104  60172490  60181937  60185148  60201208  60210901  60222273  60228109  60253692  60265320  60291587  60303882  60308486  60308488  60318918  60323225  60380530  60425094  60451958  60472785  60478988  60528756  60561940  60652992  60665686  60667010  60675139  60677495  60757022  60757023  60818475  60818476  60926379  60962902  61008290  61012473  61045594  61080408  61109126  61111268  61137389  61224633  61380169  61428505  61490674  61513896  60099552  60105524  60117023  60123236  60126285  60130243  60132076  60140248  60148220  60152406  60167105  60181938  60185149  60194305  60208420  60215605  60222274  60228147  60265321  60279007  60308490  60323226  60338200  60363746  60367392  60367393  60372745  60380531  60406057  60425095  60435664  60451959  60472786  60478989  60514761  60528757  60550169  60585166  60603156  60629961  60677537  60735920  60810911  60810912  60818477  60818478  60877879  60885041  60914239  60914240  61008291  61019228  61112944  61117673  61135017  61144270  61238546  61357131  61380170  61428509  61473780  61490676  60109544  60127522  60175615  60235190  60343737  60435665  60630550  60646457  60901698  60980738  61117476  61144275  61567406  61596997  60285810  60356067  60757025  60987232  61144277  60097891  60123237  60142535  60159627  60166855  60247225  60279009  60318919  60336413  60378455  60435666  60487174  60538620  60585168  60714759  60757026  60812039  60845591  61017587  61075643  61179542  61356012  61425847  61451148  61556241  61589562  62304876  60265322  60514763  60630552  60757027  60938161  61117477  61166269  60092678  60094831  60107372  60152408  60166856  60218737  60247226  60265323  60318920  60336414  60367394  60378456  60435667  60472787  60487175  60528758  60605453  60640228  60686349  60714760  60752291  60757028  60859835  60938162  61020843  61025869  61141259  61166272  61266932  61283304  61473781  61556242  61567407  61685403  61733235  61799295  62304883  60285812  60686350  61087351  61144278  60092679  60102584  60145086  60152409  60166857  60222581  60247227  60247228  60285813  60298456  60303884  60367395  60378457  60406059  60435668  60472788  60487176  60538621  60566853  60630553  60646460  60686351  60714761  60757029  60812042  60845594  60845595  60868440  60914241  60944483  61040087  61117478  61136021  61166274  61238547  61245937  61266933  61283305  61287811  61451149  61465685  61542650  61705590  61707791  62040735  62310055  60179267  60367396  60399651  60422449  61075644  61120680  61156091  60088627  60102585  60115131  60132070  60141582  60155291  60166859  60172491  60208422  60231259  60240383  60285814  60298457  60351965  60378458  60425096  60435669  60472789  60491807  60538622  60566854  60585170  60605454  60686352  60757030  60812045  60845596  60859838  60914242  61020844  61024794  61120681  61141260  61156092  61179543  61185819  61185820  61206345  61266934  61287812  61523652  61538670  61564723  61584667  61655406  61729809  60154145  60218740  60367397  60396401  60576770  60757031  60885045  60944484  61075645  61120682  61283306  60107373  60115133  60132071  60152410  60159629  60175617  60179268  60218744  60265324  60298458  60367398  60385463  60429160  60455023  60491812  60605455  60752294  60812048  60845597  60859840  60938163  61117480  61117481  61117482  61144280  61238548  61538671  61620161  60150049  60298459  60396402  60502198  60690628  60812049  60885046  60930488  61025870  61117483  61144281  61556243  61584668  60088628  60107374  60126286  60154147  60165154  60178905  60231260  60265325  60285815  60318921  60367399  60406060  60502222  60538623  60605456  60757032  60812051  60914243  61087355  61117484  61166276  61206346  61529262  61584669  61596999  61612599  61655407  62271604  60088629  60177168  60279011  60367400  60502229  60573171  60812052  60901702  60944486  61033149  61120683  61425850  61556244  61675726  62223006  60094832  60115135  60152411  60166860  60218745  60265327  60285976  60303887  60380532  60435670  60502233  60538624  60585171  60757033  60812056  60868443  60980740  61117487  61141261  61206347  61655408  62293140  60265328  60367401  60451960  60502235  60640229  60752295  60845598  60930490  61020845  61025871  61149930  61156093  61206348  62223010  60115138  60165155  60178906  60231261  60279013  60376943  60451961  60502239  60566856  60646461  60757034  60885048  61075646  61117488  61156094  61287813  61589564  61640837  60179270  60502244  60576772  60757035  60885049  60901703  60980741  61017588  61087360  61238549  61584670  60082460  60115140  60159630  60178907  60247230  60298460  60451962  60502247  60566857  60630554  60812058  60868445  60980742  61087368  61144282  61529263  61589565  62293143  60115142  60145087  60178908  60279014  60285817  60325545  60336415  60378459  60502258  60646462  60757036  60812060  60845602  60885050  60930491  60944487  61117489  61156095  61685419  61774748  60179271  60265329  60380534  60566858  60605457  61117490  61144284  61224634  60094833  60152412  60235191  60298461  60380535  60422450  60422451  60502270  60606170  60930493  61040089  61080564  61256432  61819965  62258970  60263981  60393284  60502275  60585173  60605458  61033150  61117491  61141262  61439803  61584671  60109551  60303889  60378460  60502277  60576773  60606179  60812062  60845603  60901704  60990572  61075647  61144286  61414330  61774749  60285819  60318922  60566860  60646463  60901705  61080565  61144288  61224635  60285820  60325546  60380536  60566861  60640230  60845604  60914244  60987233  61075648  61117495  61144289  61602057  61733227  60233951  60393286  60514764  60589774  60757037  61033152  61117496  61144290  61395117  61597002  60094834  60152414  60218746  60298462  60367403  60514765  60544297  60714763  60845605  60938164  61033154  61075649  61144292  61556246  61758021  61835772  60231262  60318923  60528759  60646464  60845606  61117497  61224636  61589566  60265330  60325547  60514766  60555776  60752296  60885051  61017589  61075650  61141263  62223011  60279016  60393289  60561202  60589775  60845607  61040090  61141264  61224637  61395118  61513902  61556247  61584672  61616167  62258977  60135129  60265331  60380538  60451963  60514767  60555778  60566863  60611473  60812063  60914245  61025872  61087372  61141265  61438779",
      "more_code_info": ""
    }
  ]
}