  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "00-5996-013-01 61782577  00-5996-013-01 61863954  00-5996-013-01 61924457  00-5996-013-01 61977562  00-5996-013-01 61977561  00-5996-013-01 62003777  00-5996-013-01 62003778  00-5996-013-01 62008974  00-5996-013-01 62008972  00-5996-013-01 62051204  00-5996-013-01 62091551  00-5996-013-01 62100471  00-5996-013-01 62107986  00-5996-013-01 62107987  00-5996-013-01 62118558  00-5996-013-01 62124601  00-5996-013-01 62134289  00-5996-013-01 62151257  00-5996-013-01 62161217  00-5996-013-01 62170068  00-5996-013-01 62172687  00-5996-013-02 61782578  00-5996-013-02 61850932  00-5996-013-02 61864885  00-5996-013-02 11400172  00-5996-013-02 61922268  00-5996-013-02 61924458  00-5996-013-02 61974046  00-5996-013-02 61974131  00-5996-013-02 61974132  00-5996-013-02 61974133  00-5996-013-02 61977563  00-5996-013-02 61978530  00-5996-013-02 61989687  00-5996-013-02 61990926  00-5996-013-02 62003779  00-5996-013-02 62004852  00-5996-013-02 62009310  00-5996-013-02 62009311  00-5996-013-02 62080341  00-5996-013-02 62091041  00-5996-013-02 62100473  00-5996-013-02 62109165  00-5996-013-02 62112981  00-5996-013-02 62124859  00-5996-013-02 62134305  00-5996-013-02 62152623  00-5996-013-02 62161218  00-5996-013-02 62170070  00-5996-014-01 61782547  00-5996-014-01 61782562  00-5996-014-01 61804267  00-5996-014-01 61804268  00-5996-014-01 61808651  00-5996-014-01 61808647  00-5996-014-01 61809730  00-5996-014-01 61809731  00-5996-014-01 61820657  00-5996-014-01 61820730  00-5996-014-01 61820668  00-5996-014-01 61829136  00-5996-014-01 61835341  00-5996-014-01 61835342  00-5996-014-01 61835835  00-5996-014-01 61835836  00-5996-014-01 61843606  00-5996-014-01 61844576  00-5996-014-01 61844577  00-5996-014-01 61844578  00-5996-014-01 61844906  00-5996-014-01 61844907  00-5996-014-01 61848266  00-5996-014-01 61848267  00-5996-014-01 61848268  00-5996-014-01 61848269  00-5996-014-01 61848270  00-5996-014-01 61853388  00-5996-014-01 61853389  00-5996-014-01 61853390  00-5996-014-01 61855970  00-5996-014-01 61855971  00-5996-014-01 61868871  00-5996-014-01 61868872  00-5996-014-01 61868873  00-5996-014-01 61871529  00-5996-014-01 61871530  00-5996-014-01 61886272  00-5996-014-01 61886273  00-5996-014-01 61886274  00-5996-014-01 61886326  00-5996-014-01 61886327  00-5996-014-01 61890029  00-5996-014-01 61890030  00-5996-014-01 61890031  00-5996-014-01 61895324  00-5996-014-01 61895325  00-5996-014-01 61895327  00-5996-014-01 61896041  00-5996-014-01 61896042  00-5996-014-01 61895326  00-5996-014-01 61910215  00-5996-014-01 61910216  00-5996-014-01 61914496  00-5996-014-01 61914497  00-5996-014-01 61915316  00-5996-014-01 61914494  00-5996-014-01 61914495  00-5996-014-01 61918706  00-5996-014-01 61918707  00-5996-014-01 61918709  00-5996-014-01 61918710  00-5996-014-01 61918711  00-5996-014-01 61922271  00-5996-014-01 61922269  00-5996-014-01 61922270  00-5996-014-01 61924459  00-5996-014-01 61924460  00-5996-014-01 61924461  00-5996-014-01 61926136  00-5996-014-01 61930162  00-5996-014-01 61931676  00-5996-014-01 61935988  00-5996-014-01 61935989  00-5996-014-01 61936798  00-5996-014-01 61935991  00-5996-014-01 61935990  00-5996-014-01 61941590  00-5996-014-01 61941591  00-5996-014-01 61964442  00-5996-014-01 61964443  00-5996-014-01 61971730  00-5996-014-01 61971732  00-5996-014-01 61971731  00-5996-014-01 61972175  00-5996-014-01 61972176  00-5996-014-01 61972747  00-5996-014-01 61972748  00-5996-014-01 61972749  00-5996-014-01 61972750  00-5996-014-01 61973060  00-5996-014-01 61973063  00-5996-014-01 61973255  00-5996-014-01 61973254  00-5996-014-01 61978533  00-5996-014-01 61979136  00-5996-014-01 61978531  00-5996-014-01 61979137  00-5996-014-01 61979138  00-5996-014-01 61979139  00-5996-014-01 61990929  00-5996-014-01 61992039  00-5996-014-01 61992040  00-5996-014-01 61992041  00-5996-014-01 61997344  00-5996-014-01 61997345  00-5996-014-01 61998635  00-5996-014-01 62004855  00-5996-014-01 62004858  00-5996-014-01 62004861  00-5996-014-01 62005005  00-5996-014-01 62010337  00-5996-014-01 62010338  00-5996-014-01 62011349  00-5996-014-01 62010339  00-5996-014-01 62018147  00-5996-014-01 62029529  00-5996-014-01 62029528  00-5996-014-01 62040744  00-5996-014-01 62040764  00-5996-014-01 62040765  00-5996-014-01 62050086  00-5996-014-01 62050087  00-5996-014-01 62050088  00-5996-014-01 62050285  00-5996-014-01 62052112  00-5996-014-01 62052113  00-5996-014-01 62052114  00-5996-014-01 62058204  00-5996-014-01 62058222  00-5996-014-01 62058223  00-5996-014-01 62066791  00-5996-014-01 62066793  00-5996-014-01 62066794  00-5996-014-01 62066792  00-5996-014-01 62067351  00-5996-014-01 62074350  00-5996-014-01 62074353  00-5996-014-01 62076157  00-5996-014-01 62076322  00-5996-014-01 62076323  00-5996-014-01 62078827  00-5996-014-01 62085957  00-5996-014-01 62092269  00-5996-014-01 62092270  00-5996-014-01 62092271  00-5996-014-01 62097832  00-5996-014-01 62098067  00-5996-014-01 62098068  00-5996-014-01 62097831  00-5996-014-01 62098069  00-5996-014-01 62102068  00-5996-014-01 62102070  00-5996-014-01 62102069  00-5996-014-01 62109669  00-5996-014-01 62109670  00-5996-014-01 62110799  00-5996-014-01 62110815  00-5996-014-01 62110817  00-5996-014-01 62109168  00-5996-014-01 62111851  00-5996-014-01 62113006  00-5996-014-01 62113007  00-5996-014-01 62113008  00-5996-014-01 62114442  00-5996-014-01 62110816  00-5996-014-01 62114446  00-5996-014-01 62114449  00-5996-014-01 62118556  00-5996-014-01 62118557  00-5996-014-01 62124313  00-5996-014-01 62124315  00-5996-014-01 62124316  00-5996-014-01 62124840  00-5996-014-01 62131254  00-5996-014-01 62131255  00-5996-014-01 62134284  00-5996-014-01 62134285  00-5996-014-01 62134287  00-5996-014-01 62141727  00-5996-014-01 62141180  00-5996-014-01 62142361  00-5996-014-01 62150038  00-5996-014-01 62150039  00-5996-014-01 62152940  00-5996-014-01 62152945  00-5996-014-01 62152624  00-5996-014-01 62162590  00-5996-014-01 62162592  00-5996-014-01 62162675  00-5996-014-01 62162677  00-5996-014-01 62162680  00-5996-014-01 62169030  00-5996-014-01 62169031  00-5996-014-01 62169032  00-5996-014-01 62169052  00-5996-014-02 61782563  00-5996-014-02 61796373  00-5996-014-02 61804269  00-5996-014-02 61804270  00-5996-014-02 61804271  00-5996-014-02 61808813  00-5996-014-02 61808812  00-5996-014-02 61809806  00-5996-014-02 61810421  00-5996-014-02 61819953  00-5996-014-02 61819954  00-5996-014-02 61820652  00-5996-014-02 61821140  00-5996-014-02 61821141  00-5996-014-02 61821142  00-5996-014-02 61821143  00-5996-014-02 61835337  00-5996-014-02 61835338  00-5996-014-02 61835339  00-5996-014-02 61835340  00-5996-014-02 61843147  00-5996-014-02 61843148  00-5996-014-02 61843149  00-5996-014-02 61844237  00-5996-014-02 61844238  00-5996-014-02 61844573  00-5996-014-02 61844574  00-5996-014-02 61844575  00-5996-014-02 61848354  00-5996-014-02 61848355  00-5996-014-02 61854828  00-5996-014-02 61854829  00-5996-014-02 61855966  00-5996-014-02 61856750  00-5996-014-02 61857903  00-5996-014-02 61869855  00-5996-014-02 61869856  00-5996-014-02 61869857  00-5996-014-02 61869858  00-5996-014-02 61874468  00-5996-014-02 61874469  00-5996-014-02 61888041  00-5996-014-02 61886325  00-5996-014-02 61888042  00-5996-014-02 61888043  00-5996-014-02 61889063  00-5996-014-02 61892375  00-5996-014-02 61892376  00-5996-014-02 61892377  00-5996-014-02 61896208  00-5996-014-02 61896210  00-5996-014-02 61896212  00-5996-014-02 61896213  00-5996-014-02 61896215  00-5996-014-02 61897289  00-5996-014-02 61897290  00-5996-014-02 61904318  00-5996-014-02 61904923  00-5996-014-02 61904924  00-5996-014-02 61908306  00-5996-014-02 61908307  00-5996-014-02 61910883  00-5996-014-02 61912365  00-5996-014-02 61915317  00-5996-014-02 61915319  00-5996-014-02 61915318  00-5996-014-02 61918978  00-5996-014-02 61918980  00-5996-014-02 61918979  00-5996-014-02 61923583  00-5996-014-02 61923584  00-5996-014-02 61923585  00-5996-014-02 61926137  00-5996-014-02 61926138  00-5996-014-02 61926139  00-5996-014-02 61926140  00-5996-014-02 61932294  00-5996-014-02 61931681  00-5996-014-02 61936801  00-5996-014-02 61936804  00-5996-014-02 61937295  00-5996-014-02 61936802  00-5996-014-02 61936803  00-5996-014-02 61943194  00-5996-014-02 61943195  00-5996-014-02 61956844  00-5996-014-02 61956845  00-5996-014-02 61964959  00-5996-014-02 61964961  00-5996-014-02 61965578  00-5996-014-02 61971733  00-5996-014-02 61972173  00-5996-014-02 61972174  00-5996-014-02 61973256  00-5996-014-02 61973257  00-5996-014-02 61973259  00-5996-014-02 61974054  00-5996-014-02 61974057  00-5996-014-02 61973258  00-5996-014-02 61974135  00-5996-014-02 61974136  00-5996-014-02 61974137  00-5996-014-02 61976331  00-5996-014-02 61978527  00-5996-014-02 61978528  00-5996-014-02 61978529  00-5996-014-02 61979780  00-5996-014-02 61979781  00-5996-014-02 61979782  00-5996-014-02 61992063  00-5996-014-02 61992064  00-5996-014-02 61992065  00-5996-014-02 61998638  00-5996-014-02 61998639  00-5996-014-02 61998640  00-5996-014-02 62005006  00-5996-014-02 62005007  00-5996-014-02 62005918  00-5996-014-02 62005919  00-5996-014-02 62005959  00-5996-014-02 62007006  00-5996-014-02 62005958  00-5996-014-02 62011350  00-5996-014-02 62011351  00-5996-014-02 62025615  00-5996-014-02 62025614  00-5996-014-02 62025618  00-5996-014-02 62025619  00-5996-014-02 62026097  00-5996-014-02 62026098  00-5996-014-02 62030597  00-5996-014-02 62030412  00-5996-014-02 62030414  00-5996-014-02 62030598  00-5996-014-02 62030599  00-5996-014-02 62042326  00-5996-014-02 62042327  00-5996-014-02 62043152  00-5996-014-02 62048445  00-5996-014-02 62048446  00-5996-014-02 62049392  00-5996-014-02 62049393  00-5996-014-02 62052676  00-5996-014-02 62052678  00-5996-014-02 62052679  00-5996-014-02 62052680  00-5996-014-02 62052677  00-5996-014-02 62058789  00-5996-014-02 62064903  00-5996-014-02 62058794  00-5996-014-02 62065585  00-5996-014-02 62065586  00-5996-014-02 62065587  00-5996-014-02 62065588  00-5996-014-02 62066549  00-5996-014-02 62066550  00-5996-014-02 62074426  00-5996-014-02 62073116  00-5996-014-02 62076324  00-5996-014-02 62077296  00-5996-014-02 62077297  00-5996-014-02 62076325  00-5996-014-02 62078828  00-5996-014-02 62078829  00-5996-014-02 62079859  00-5996-014-02 62086616  00-5996-014-02 62091552  00-5996-014-02 62091553  00-5996-014-02 62098554  00-5996-014-02 62098800  00-5996-014-02 62098842  00-5996-014-02 62098843  00-5996-014-02 62105733  00-5996-014-02 62105734  00-5996-014-02 62105735  00-5996-014-02 62105736  00-5996-014-02 62105737  00-5996-014-02 62111951  00-5996-014-02 62111990  00-5996-014-02 62111991  00-5996-014-02 62111993  00-5996-014-02 62111992  00-5996-014-02 62112978  00-5996-014-02 62112979  00-5996-014-02 62115269  00-5996-014-02 62115963  00-5996-014-02 62115264  00-5996-014-02 62116019  00-5996-014-02 62117286  00-5996-014-02 62115964  00-5996-014-02 62116020  00-5996-014-02 62115965  00-5996-014-02 62118617  00-5996-014-02 62118618  00-5996-014-02 62127485  00-5996-014-02 62127483  00-5996-014-02 62128366  00-5996-014-02 11400468  00-5996-014-02 62128367  00-5996-014-02 62132105  00-5996-014-02 62132106  00-5996-014-02 62136383  00-5996-014-02 62136384  00-5996-014-02 62137088  00-5996-014-02 62137094  00-5996-014-02 62143179  00-5996-014-02 62143180  00-5996-014-02 62143623  00-5996-014-02 62148762  00-5996-014-02 62149013  00-5996-014-02 62148763  00-5996-014-02 62154047  00-5996-014-02 62154076  00-5996-014-02 62154289  00-5996-014-02 62156821  00-5996-014-02 62163081  00-5996-014-02 62164673  00-5996-014-02 62163107  00-5996-014-02 62163106  00-5996-014-02 62164674  00-5996-014-02 62170062  00-5996-014-02 62170063  00-5996-014-02 62170065  00-5996-014-02 62170064  00-5996-015-01 61796374  00-5996-015-01 61796375  00-5996-015-01 61806287  00-5996-015-01 61807833  00-5996-015-01 61807834  00-5996-015-01 61807835  00-5996-015-01 61807836  00-5996-015-01 61810425  00-5996-015-01 61810426  00-5996-015-01 61816401  00-5996-015-01 61816402  00-5996-015-01 61817757  00-5996-015-01 61817758  00-5996-015-01 61817759  00-5996-015-01 61817760  00-5996-015-01 11400109  00-5996-015-01 61817761  00-5996-015-01 61819056  00-5996-015-01 61819057  00-5996-015-01 61819058  00-5996-015-01 61821136  00-5996-015-01 61821137  00-5996-015-01 61821138  00-5996-015-01 61823893  00-5996-015-01 61826940  00-5996-015-01 61833606  00-5996-015-01 61834869  00-5996-015-01 61834870  00-5996-015-01 61834871  00-5996-015-01 61833607  00-5996-015-01 61835324  00-5996-015-01 61835390  00-5996-015-01 61835391  00-5996-015-01 61838766  00-5996-015-01 61839705  00-5996-015-01 61839706  00-5996-015-01 61844933  00-5996-015-01 61844934  00-5996-015-01 61844935  00-5996-015-01 61849038  00-5996-015-01 61849039  00-5996-015-01 61849040  00-5996-015-01 61849238  00-5996-015-01 61850208  00-5996-015-01 61850209  00-5996-015-01 61850210  00-5996-015-01 61850211  00-5996-015-01 61852850  00-5996-015-01 61852849  00-5996-015-01 61852851  00-5996-015-01 61853366  00-5996-015-01 61853367  00-5996-015-01 61853368  00-5996-015-01 61888055  00-5996-015-01 61889069  00-5996-015-01 61890032  00-5996-015-01 61892366  00-5996-015-01 61892367  00-5996-015-01 61893240  00-5996-015-01 61893241  00-5996-015-01 61897285  00-5996-015-01 61898492  00-5996-015-01 61898493  00-5996-015-01 61897284  00-5996-015-01 61897286  00-5996-015-01 61898494  00-5996-015-01 61898495  00-5996-015-01 61898496  00-5996-015-01 61899634  00-5996-015-01 61899635  00-5996-015-01 61899633  00-5996-015-01 61905195  00-5996-015-01 61905196  00-5996-015-01 61905197  00-5996-015-01 61905198  00-5996-015-01 61905199  00-5996-015-01 61905286  00-5996-015-01 61905292  00-5996-015-01 61908297  00-5996-015-01 61908298  00-5996-015-01 61908299  00-5996-015-01 61915320  00-5996-015-01 61915321  00-5996-015-01 61916156  00-5996-015-01 61916158  00-5996-015-01 61916175  00-5996-015-01 61916157  00-5996-015-01 61917761  00-5996-015-01 61918957  00-5996-015-01 61920450  00-5996-015-01 61920451  00-5996-015-01 61920452  00-5996-015-01 61920453  00-5996-015-01 61920454  00-5996-015-01 61926141  00-5996-015-01 61927972  00-5996-015-01 61927973  00-5996-015-01 61927970  00-5996-015-01 61927971  00-5996-015-01 61932298  00-5996-015-01 61933648  00-5996-015-01 61933664  00-5996-015-01 61933658  00-5996-015-01 61937762  00-5996-015-01 61937297  00-5996-015-01 61937763  00-5996-015-01 61937764  00-5996-015-01 61937804  00-5996-015-01 61937805  00-5996-015-01 61944892  00-5996-015-01 61944893  00-5996-015-01 11400304  00-5996-015-01 61958028  00-5996-015-01 61958029  00-5996-015-01 61963902  00-5996-015-01 61963903  00-5996-015-01 61963904  00-5996-015-01 61963905  00-5996-015-01 61964138  00-5996-015-01 61964145  00-5996-015-01 61964444  00-5996-015-01 61964445  00-5996-015-01 61968859  00-5996-015-01 61968868  00-5996-015-01 61968871  00-5996-015-01 61970686  00-5996-015-01 61970687  00-5996-015-01 61974134  00-5996-015-01 61976419  00-5996-015-01 61976420  00-5996-015-01 61976421  00-5996-015-01 61979784  00-5996-015-01 61979785  00-5996-015-01 61980584  00-5996-015-01 61980585  00-5996-015-01 61980586  00-5996-015-01 61980587  00-5996-015-01 61982315  00-5996-015-01 61982319  00-5996-015-01 11400335  00-5996-015-01 61982320  00-5996-015-01 61985029  00-5996-015-01 61984937  00-5996-015-01 61984938  00-5996-015-01 61992803  00-5996-015-01 61992805  00-5996-015-01 61992985  00-5996-015-01 61992804  00-5996-015-01 61992984  00-5996-015-01 61998812  00-5996-015-01 61998814  00-5996-015-01 61998817  00-5996-015-01 61999563  00-5996-015-01 61999565  00-5996-015-01 61999564  00-5996-015-01 61999999  00-5996-015-01 62011569  00-5996-015-01 62011570  00-5996-015-01 62011571  00-5996-015-01 62019496  00-5996-015-01 62019497  00-5996-015-01 62019498  00-5996-015-01 62028161  00-5996-015-01 62028163  00-5996-015-01 62028167  00-5996-015-01 62032086  00-5996-015-01 62030638  00-5996-015-01 62032087  00-5996-015-01 62032088  00-5996-015-01 62032738  00-5996-015-01 62039174  00-5996-015-01 62039175  00-5996-015-01 62039176  00-5996-015-01 62040697  00-5996-015-01 62050079  00-5996-015-01 62050080  00-5996-015-01 62050757  00-5996-015-01 62050759  00-5996-015-01 62052662  00-5996-015-01 62052661  00-5996-015-01 62052663  00-5996-015-01 62054080  00-5996-015-01 62054081  00-5996-015-01 62054837  00-5996-015-01 62054838  00-5996-015-01 62054840  00-5996-015-01 62060532  00-5996-015-01 62060533  00-5996-015-01 62060534  00-5996-015-01 62061883  00-5996-015-01 62061884  00-5996-015-01 62061885  00-5996-015-01 62061886  00-5996-015-01 62066747  00-5996-015-01 62066750  00-5996-015-01 62066753  00-5996-015-01 62067345  00-5996-015-01 62067346  00-5996-015-01 62067786  00-5996-015-01 62067787  00-5996-015-01 62072522  00-5996-015-01 11400419  00-5996-015-01 62073109  00-5996-015-01 62073110  00-5996-015-01 62080825  00-5996-015-01 62080823  00-5996-015-01 62081434  00-5996-015-01 62081438  00-5996-015-01 62086995  00-5996-015-01 62086998  00-5996-015-01 62087881  00-5996-015-01 62088449  00-5996-015-01 62088466  00-5996-015-01 62088881  00-5996-015-01 62088883  00-5996-015-01 62095263  00-5996-015-01 62095268  00-5996-015-01 62095271  00-5996-015-01 62095275  00-5996-015-01 62096012  00-5996-015-01 62096013  00-5996-015-01 62103204  00-5996-015-01 62103209  00-5996-015-01 62105726  00-5996-015-01 62105727  00-5996-015-01 62103208  00-5996-015-01 62106772  00-5996-015-01 62107115  00-5996-015-01 62107117  00-5996-015-01 62107111  00-5996-015-01 62107946  00-5996-015-01 62107947  00-5996-015-01 62111995  00-5996-015-01 62111953  00-5996-015-01 62111994  00-5996-015-01 62114099  00-5996-015-01 62114026  00-5996-015-01 62114027  00-5996-015-01 62114101  00-5996-015-01 62114451  00-5996-015-01 62114452  00-5996-015-01 62114455  00-5996-015-01 62119549  00-5996-015-01 62119550  00-5996-015-01 62124865  00-5996-015-01 62124874  00-5996-015-01 62125505  00-5996-015-01 62125506  00-5996-015-01 62125507  00-5996-015-01 62132867  00-5996-015-01 62137990  00-5996-015-01 62137991  00-5996-015-01 62138243  00-5996-015-01 62142619  00-5996-015-01 62142620  00-5996-015-01 62149475  00-5996-015-01 62149476  00-5996-015-01 62149526  00-5996-015-01 62152644  00-5996-015-01 62152645  00-5996-015-01 62152622  00-5996-015-01 62157961  00-5996-015-01 62159292  00-5996-015-01 62159293  00-5996-015-01 62164669  00-5996-015-01 62164819  00-5996-015-01 62164821  00-5996-015-01 62164863  00-5996-015-01 62164864  00-5996-015-01 62165856  00-5996-015-01 62165931  00-5996-015-01 62164858  00-5996-015-01 62165853  00-5996-015-01 62168303  00-5996-015-01 62170395  00-5996-015-01 62168304  00-5996-015-01 62170396  00-5996-015-01 62170910  00-5996-015-01 62172678  00-5996-015-01 62172681  00-5996-015-01 62172682  00-5996-015-01 62172683  00-5996-015-01 62172676  00-5996-015-02 61782568  00-5996-015-02 61797751  00-5996-015-02 61797752  00-5996-015-02 61804739  00-5996-015-02 61804740  00-5996-015-02 61804741  00-5996-015-02 61804742  00-5996-015-02 61806283  00-5996-015-02 61806284  00-5996-015-02 61813678  00-5996-015-02 61813679  00-5996-015-02 61814768  00-5996-015-02 61814769  00-5996-015-02 61814770  00-5996-015-02 61815716  00-5996-015-02 61815717  00-5996-015-02 61815718  00-5996-015-02 61815719  00-5996-015-02 11400108  00-5996-015-02 61819956  00-5996-015-02 61820646  00-5996-015-02 61820647  00-5996-015-02 61820648  00-5996-015-02 61820714  00-5996-015-02 61823898  00-5996-015-02 61824782  00-5996-015-02 61824783  00-5996-015-02 61826936  00-5996-015-02 61834872  00-5996-015-02 61834873  00-5996-015-02 61834874  00-5996-015-02 61835824  00-5996-015-02 61835826  00-5996-015-02 61835825  00-5996-015-02 61838756  00-5996-015-02 61846263  00-5996-015-02 61846264  00-5996-015-02 61846266  00-5996-015-02 61846267  00-5996-015-02 61846265  00-5996-015-02 61847677  00-5996-015-02 61847679  00-5996-015-02 61846252  00-5996-015-02 61847678  00-5996-015-02 61847709  00-5996-015-02 61846251  00-5996-015-02 61846253  00-5996-015-02 61847657  00-5996-015-02 61847658  00-5996-015-02 61847659  00-5996-015-02 61847660  00-5996-015-02 61847710  00-5996-015-02 11400143  00-5996-015-02 61850229  00-5996-015-02 61850230  00-5996-015-02 61850231  00-5996-015-02 61850232  00-5996-015-02 61850933  00-5996-015-02 61850934  00-5996-015-02 61851731  00-5996-015-02 61851732  00-5996-015-02 61851733  00-5996-015-02 61852861  00-5996-015-02 61852862  00-5996-015-02 61890015  00-5996-015-02 61890016  00-5996-015-02 61890691  00-5996-015-02 61890692  00-5996-015-02 61890703  00-5996-015-02 61890704  00-5996-015-02 61890702  00-5996-015-02 61899623  00-5996-015-02 61899622  00-5996-015-02 61899624  00-5996-015-02 61899637  00-5996-015-02 61899638  00-5996-015-02 61900454  00-5996-015-02 61900455  00-5996-015-02 61900456  00-5996-015-02 61900464  00-5996-015-02 61900465  00-5996-015-02 61899636  00-5996-015-02 61900457  00-5996-015-02 61905293  00-5996-015-02 61905294  00-5996-015-02 61906071  00-5996-015-02 61906389  00-5996-015-02 61907454  00-5996-015-02 61906390  00-5996-015-02 61909349  00-5996-015-02 61909350  00-5996-015-02 61910208  00-5996-015-02 61917762  00-5996-015-02 61917764  00-5996-015-02 61917765  00-5996-015-02 61917763  00-5996-015-02 61917766  00-5996-015-02 61918729  00-5996-015-02 61918730  00-5996-015-02 61918731  00-5996-015-02 11400220  00-5996-015-02 61920455  00-5996-015-02 61920456  00-5996-015-02 61921455  00-5996-015-02 61921456  00-5996-015-02 61921458  00-5996-015-02 61921457  00-5996-015-02 61929409  00-5996-015-02 61929411  00-5996-015-02 61929412  00-5996-015-02 61930160  00-5996-015-02 61929410  00-5996-015-02 61934968  00-5996-015-02 61934969  00-5996-015-02 61934970  00-5996-015-02 61934971  00-5996-015-02 61937807  00-5996-015-02 61937808  00-5996-015-02 61938903  00-5996-015-02 61938904  00-5996-015-02 61938905  00-5996-015-02 61939186  00-5996-015-02 61946494  00-5996-015-02 61946495  00-5996-015-02 61958544  00-5996-015-02 61958545  00-5996-015-02 61958546  00-5996-015-02 61958763  00-5996-015-02 61960722  00-5996-015-02 61958764  00-5996-015-02 61964972  00-5996-015-02 61964973  00-5996-015-02 61965580  00-5996-015-02 61965582  00-5996-015-02 61965581  00-5996-015-02 61966246  00-5996-015-02 61966247  00-5996-015-02 61971171  00-5996-015-02 61971178  00-5996-015-02 61971180  00-5996-015-02 61976422  00-5996-015-02 61977366  00-5996-015-02 61977374  00-5996-015-02 61977380  00-5996-015-02 61982471  00-5996-015-02 61982472  00-5996-015-02 61982473  00-5996-015-02 61982474  00-5996-015-02 11005547  00-5996-015-02 61982781  00-5996-015-02 61982793  00-5996-015-02 61983011  00-5996-015-02 61983009  00-5996-015-02 61983010  00-5996-015-02 61985030  00-5996-015-02 61985031  00-5996-015-02 61985032  00-5996-015-02 61986205  00-5996-015-02 61986207  00-5996-015-02 61986211  00-5996-015-02 61993954  00-5996-015-02 61993960  00-5996-015-02 61993957  00-5996-015-02 61995688  00-5996-015-02 61995689  00-5996-015-02 62000000  00-5996-015-02 62000001  00-5996-015-02 62000004  00-5996-015-02 62000977  00-5996-015-02 62000978  00-5996-015-02 62000979  00-5996-015-02 62002044  00-5996-015-02 62002045  00-5996-015-02 62007007  00-5996-015-02 62007008  00-5996-015-02 62012110  00-5996-015-02 62012115  00-5996-015-02 62012122  00-5996-015-02 62012127  00-5996-015-02 62012663  00-5996-015-02 62013563  00-5996-015-02 62012664  00-5996-015-02 62022048  00-5996-015-02 62022049  00-5996-015-02 62022328  00-5996-015-02 62023800  00-5996-015-02 62023802  00-5996-015-02 62023801  00-5996-015-02 62024962  00-5996-015-02 62024963  00-5996-015-02 62028455  00-5996-015-02 62028456  00-5996-015-02 62032745  00-5996-015-02 62033091  00-5996-015-02 62033092  00-5996-015-02 62033714  00-5996-015-02 62034612  00-5996-015-02 62033715  00-5996-015-02 62042145  00-5996-015-02 62042147  00-5996-015-02 62042151  00-5996-015-02 62042153  00-5996-015-02 62042285  00-5996-015-02 62043136  00-5996-015-02 62048259  00-5996-015-02 62048265  00-5996-015-02 62048436  00-5996-015-02 62048437  00-5996-015-02 62048438  00-5996-015-02 62050078  00-5996-015-02 62055037  00-5996-015-02 62055036  00-5996-015-02 62055038  00-5996-015-02 62056776  00-5996-015-02 62056777  00-5996-015-02 62061653  00-5996-015-02 62061654  00-5996-015-02 62061655  00-5996-015-02 62062554  00-5996-015-02 62061880  00-5996-015-02 62062556  00-5996-015-02 62064894  00-5996-015-02 62064895  00-5996-015-02 62065327  00-5996-015-02 62065340  00-5996-015-02 62065351  00-5996-015-02 62065354  00-5996-015-02 62066544  00-5996-015-02 62066543  00-5996-015-02 62073111  00-5996-015-02 62073112  00-5996-015-02 62074299  00-5996-015-02 62083135  00-5996-015-02 62081869  00-5996-015-02 62083137  00-5996-015-02 62087936  00-5996-015-02 62090402  00-5996-015-02 62091476  00-5996-015-02 62088885  00-5996-015-02 62091020  00-5996-015-02 11400433  00-5996-015-02 62088888  00-5996-015-02 62091548  00-5996-015-02 62091550  00-5996-015-02 62091485  00-5996-015-02 62096714  00-5996-015-02 62096715  00-5996-015-02 62097828  00-5996-015-02 62097830  00-5996-015-02 62098064  00-5996-015-02 62098065  00-5996-015-02 62103217  00-5996-015-02 62106545  00-5996-015-02 62105738  00-5996-015-02 62105740  00-5996-015-02 62105741  00-5996-015-02 62106543  00-5996-015-02 62106691  00-5996-015-02 62106540  00-5996-015-02 62106692  00-5996-015-02 62106693  00-5996-015-02 62106694  00-5996-015-02 62111494  00-5996-015-02 62111495  00-5996-015-02 62111496  00-5996-015-02 62111952  00-5996-015-02 62111498  00-5996-015-02 62116024  00-5996-015-02 62116026  00-5996-015-02 62116027  00-5996-015-02 62117279  00-5996-015-02 62117280  00-5996-015-02 62117281  00-5996-015-02 62117282  00-5996-015-02 62121232  00-5996-015-02 62121233  00-5996-015-02 62121777  00-5996-015-02 62121778  00-5996-015-02 62128379  00-5996-015-02 62128380  00-5996-015-02 62129225  00-5996-015-02 62129227  00-5996-015-02 62134714  00-5996-015-02 62134749  00-5996-015-02 62134750  00-5996-015-02 62134751  00-5996-015-02 62133179  00-5996-015-02 62144186  00-5996-015-02 62144187  00-5996-015-02 62149529  00-5996-015-02 62149547  00-5996-015-02 62149546  00-5996-015-02 62150036  00-5996-015-02 62152934  00-5996-015-02 62152936  00-5996-015-02 62154044  00-5996-015-02 62152935  00-5996-015-02 62160287  00-5996-015-02 62160340  00-5996-015-02 62160341  00-5996-015-02 62160801  00-5996-015-02 62160802  00-5996-015-02 62166652  00-5996-015-02 62166741  00-5996-015-02 62167495  00-5996-015-02 62166651  00-5996-015-02 62167496  00-5996-015-02 62167497  00-5996-015-02 62171593  00-5996-015-02 62166740  00-5996-015-02 62171095  00-5996-015-02 62171594  00-5996-015-02 62171595  00-5996-015-02 62171596  00-5996-015-02 62173187  00-5996-016-01 61782551  00-5996-016-01 61792785  00-5996-016-01 61794726  00-5996-016-01 61794727  00-5996-016-01 61794725  00-5996-016-01 61804272  00-5996-016-01 61804273  00-5996-016-01 61804274  00-5996-016-01 61811786  00-5996-016-01 61811787  00-5996-016-01 61817762  00-5996-016-01 61817763  00-5996-016-01 61817764  00-5996-016-01 61820742  00-5996-016-01 61821428  00-5996-016-01 61821429  00-5996-016-01 61835343  00-5996-016-01 61835344  00-5996-016-01 61835345  00-5996-016-01 61835346  00-5996-016-01 61839733  00-5996-016-01 61844593  00-5996-016-01 61844912  00-5996-016-01 61844913  00-5996-016-01 61844914  00-5996-016-01 61844952  00-5996-016-01 61848296  00-5996-016-01 61848297  00-5996-016-01 61848298  00-5996-016-01 61848299  00-5996-016-01 61848301  00-5996-016-01 61851705  00-5996-016-01 61850937  00-5996-016-01 61850938  00-5996-016-01 61851704  00-5996-016-01 61855974  00-5996-016-01 61863958  00-5996-016-01 61863959  00-5996-016-01 61894499  00-5996-016-01 61894501  00-5996-016-01 61894502  00-5996-016-01 61895328  00-5996-016-01 61895329  00-5996-016-01 61900466  00-5996-016-01 61900467  00-5996-016-01 61907456  00-5996-016-01 61907455  00-5996-016-01 61910213  00-5996-016-01 11400246  00-5996-016-01 61914456  00-5996-016-01 61914457  00-5996-016-01 61914458  00-5996-016-01 11400248  00-5996-016-01 61910874  00-5996-016-01 61914459  00-5996-016-01 11400216  00-5996-016-01 61918733  00-5996-016-01 61918732  00-5996-016-01 61918734  00-5996-016-01 61918952  00-5996-016-01 61922267  00-5996-016-01 61923578  00-5996-016-01 61923579  00-5996-016-01 61922266  00-5996-016-01 11400223  00-5996-016-01 61926142  00-5996-016-01 61926143  00-5996-016-01 61926144  00-5996-016-01 61926146  00-5996-016-01 61926145  00-5996-016-01 11400279  00-5996-016-01 61932299  00-5996-016-01 61932300  00-5996-016-01 61932301  00-5996-016-01 61936805  00-5996-016-01 61936807  00-5996-016-01 61936809  00-5996-016-01 61936806  00-5996-016-01 61936808  00-5996-016-01 61936810  00-5996-016-01 11400292  00-5996-016-01 61944902  00-5996-016-01 61979783  00-5996-016-01 61980319  00-5996-016-01 61980320  00-5996-016-01 61980580  00-5996-016-01 61980581  00-5996-016-01 61986215  00-5996-016-01 61986217  00-5996-016-01 61990279  00-5996-016-01 61990280  00-5996-016-01 61990281  00-5996-016-01 61990282  00-5996-016-01 62002071  00-5996-016-01 62002070  00-5996-016-01 62008279  00-5996-016-01 62008280  00-5996-016-01 62013564  00-5996-016-01 62014085  00-5996-016-01 62014086  00-5996-016-01 62019501  00-5996-016-01 62028457  00-5996-016-01 62029547  00-5996-016-01 62029546  00-5996-016-01 62030605  00-5996-016-01 62030606  00-5996-016-01 62036146  00-5996-016-01 62036147  00-5996-016-01 62036145  00-5996-016-01 62036148  00-5996-016-01 62047531  00-5996-016-01 62047530  00-5996-016-01 62051202  00-5996-016-01 62050778  00-5996-016-01 62051199  00-5996-016-01 62052108  00-5996-016-01 62052109  00-5996-016-01 62058220  00-5996-016-01 62058221  00-5996-016-01 62066551  00-5996-016-01 62067354  00-5996-016-01 62066552  00-5996-016-01 62067353  00-5996-016-01 62072525  00-5996-016-01 62072524  00-5996-016-01 62076326  00-5996-016-01 62076158  00-5996-016-01 62076329  00-5996-016-01 62077301  00-5996-016-01 62078810  00-5996-016-01 62083131  00-5996-016-01 62083133  00-5996-016-01 62091885  00-5996-016-01 62092096  00-5996-016-01 62092097  00-5996-016-01 62100477  00-5996-016-01 62102071  00-5996-016-01 62101634  00-5996-016-01 62107935  00-5996-016-01 62107936  00-5996-016-01 62107952  00-5996-016-01 62107991  00-5996-016-01 62117332  00-5996-016-01 62117333  00-5996-016-01 62117334  00-5996-016-01 62117335  00-5996-016-01 62121206  00-5996-016-01 62121234  00-5996-016-01 62126117  00-5996-016-01 62126118  00-5996-016-01 62126745  00-5996-016-01 62131023  00-5996-016-01 62137109  00-5996-016-01 62137520  00-5996-016-01 62144317  00-5996-016-01 62144316  00-5996-016-01 62148245  00-5996-016-01 62156811  00-5996-016-01 62156812  00-5996-016-01 62159331  00-5996-016-01 62159294  00-5996-016-01 62164359  00-5996-016-01 62164676  00-5996-016-01 62164831  00-5996-016-01 62164832  00-5996-016-01 62164871  00-5996-016-01 62164872  00-5996-016-01 62164675  00-5996-016-01 62168354  00-5996-016-01 62168355  00-5996-016-01 11400503  00-5996-016-02 61794724  00-5996-016-02 61804275  00-5996-016-02 61804276  00-5996-016-02 61804277  00-5996-016-02 61804743  00-5996-016-02 61804744  00-5996-016-02 61810418  00-5996-016-02 61810419  00-5996-016-02 61835347  00-5996-016-02 61835348  00-5996-016-02 61835841  00-5996-016-02 61835842  00-5996-016-02 61841282  00-5996-016-02 61841283  00-5996-016-02 61841284  00-5996-016-02 61844052  00-5996-016-02 61844053  00-5996-016-02 61844590  00-5996-016-02 61844591  00-5996-016-02 61844592  00-5996-016-02 61846268  00-5996-016-02 61846269  00-5996-016-02 61846270  00-5996-016-02 61849052  00-5996-016-02 61849053  00-5996-016-02 61849054  00-5996-016-02 61849055  00-5996-016-02 61849056  00-5996-016-02 61849057  00-5996-016-02 61850223  00-5996-016-02 61851706  00-5996-016-02 61851707  00-5996-016-02 61851708  00-5996-016-02 61851709  00-5996-016-02 61856761  00-5996-016-02 61856762  00-5996-016-02 61864890  00-5996-016-02 61864891  00-5996-016-02 61889065  00-5996-016-02 61889066  00-5996-016-02 61889067  00-5996-016-02 61896050  00-5996-016-02 61896052  00-5996-016-02 61896051  00-5996-016-02 61897294  00-5996-016-02 61901177  00-5996-016-02 61901178  00-5996-016-02 61907457  00-5996-016-02 61907458  00-5996-016-02 61910876  00-5996-016-02 61910877  00-5996-016-02 61914460  00-5996-016-02 61914461  00-5996-016-02 61915323  00-5996-016-02 11400217  00-5996-016-02 61915322  00-5996-016-02 61918954  00-5996-016-02 61918955  00-5996-016-02 61918953  00-5996-016-02 61918956  00-5996-016-02 11400222  00-5996-016-02 61923581  00-5996-016-02 61923582  00-5996-016-02 61924456  00-5996-016-02 61923580  00-5996-016-02 11400278",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "LPS OPTION FEMORAL C-L  LPS OPTION FEMORAL C-R  LPS OPTION FEMORAL D-L  LPS OPTION FEMORAL D-R  LPS OPTION FEMORAL E-L  LPS OPTION FEMORAL E-R  LPS OPTION FEMORAL F-L  LPS OPTION FEMORAL F-R  LPS OPTION FEMORAL G-L  LPS OPTION FEMORAL G-R  LPS OPTION FEMORAL H-L    This device is indicated for patients with severe knee pain and disability",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1035-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "00-5996-016-02 61927976  00-5996-016-02 61927975  00-5996-016-02 61927977  00-5996-016-02 61929413  00-5996-016-02 61929414  00-5996-016-02 61933842  00-5996-016-02 61933843  00-5996-016-02 61937765  00-5996-016-02 61937809  00-5996-016-02 61937810  00-5996-016-02 61937811  00-5996-016-02 61937812  00-5996-016-02 61938951  00-5996-016-02 61946501  00-5996-016-02 61946502  00-5996-016-02 11005506  00-5996-016-02 61980582  00-5996-016-02 61980583  00-5996-016-02 61982307  00-5996-016-02 61982308  00-5996-016-02 61982309  00-5996-016-02 61982467  00-5996-016-02 61986727  00-5996-016-02 61986728  00-5996-016-02 61990643  00-5996-016-02 61990646  00-5996-016-02 61990648  00-5996-016-02 62002072  00-5996-016-02 62002073  00-5996-016-02 62008281  00-5996-016-02 62008282  00-5996-016-02 62014087  00-5996-016-02 62014427  00-5996-016-02 62014432  00-5996-016-02 62015857  00-5996-016-02 62014429  00-5996-016-02 62014431  00-5996-016-02 62019503  00-5996-016-02 62019504  00-5996-016-02 62029869  00-5996-016-02 62032090  00-5996-016-02 62032869  00-5996-016-02 62032879  00-5996-016-02 62039085  00-5996-016-02 62039083  00-5996-016-02 62039084  00-5996-016-02 62046238  00-5996-016-02 62046237  00-5996-016-02 62052672  00-5996-016-02 62052673  00-5996-016-02 62052674  00-5996-016-02 62056784  00-5996-016-02 62052675  00-5996-016-02 62057272  00-5996-016-02 62059387  00-5996-016-02 62059388  00-5996-016-02 62059389  00-5996-016-02 62060481  00-5996-016-02 62060482  00-5996-016-02 62067953  00-5996-016-02 62068312  00-5996-016-02 62068313  00-5996-016-02 62068314  00-5996-016-02 62067954  00-5996-016-02 62070836  00-5996-016-02 62070837  00-5996-016-02 62078832  00-5996-016-02 62080339  00-5996-016-02 62084443  00-5996-016-02 62084444  00-5996-016-02 62068099  00-5996-016-02 62092529  00-5996-016-02 62092530  00-5996-016-02 62092531  00-5996-016-02 62099270  00-5996-016-02 62099271  00-5996-016-02 62098840  00-5996-016-02 62098841  00-5996-016-02 62099272  00-5996-016-02 62099910  00-5996-016-02 62109172  00-5996-016-02 62107992  00-5996-016-02 62109173  00-5996-016-02 62109174  00-5996-016-02 62118555  00-5996-016-02 62121802  00-5996-016-02 62121803  00-5996-016-02 62127175  00-5996-016-02 62127174  00-5996-016-02 62127176  00-5996-016-02 62131204  00-5996-016-02 62131205  00-5996-016-02 62131217  00-5996-016-02 62131218  00-5996-016-02 62131206  00-5996-016-02 62138245  00-5996-016-02 62138246  00-5996-016-02 62142599  00-5996-016-02 62142601  00-5996-016-02 62147822  00-5996-016-02 62148244  00-5996-016-02 62155685  00-5996-016-02 62155686  00-5996-016-02 62160285  00-5996-016-02 62166661  00-5996-016-02 62166747  00-5996-016-02 62167511  00-5996-016-02 62167512  00-5996-016-02 62167513  00-5996-016-02 62168311  00-5996-016-02 62168313  00-5996-016-02 62166662  00-5996-016-02 62166746  00-5996-017-01 61796376  00-5996-017-01 61796377  00-5996-017-01 61804745  00-5996-017-01 61804746  00-5996-017-01 61808890  00-5996-017-01 61808891  00-5996-017-01 61835839  00-5996-017-01 61842361  00-5996-017-01 61850225  00-5996-017-01 61850226  00-5996-017-01 61852840  00-5996-017-01 61852841  00-5996-017-01 61870738  00-5996-017-01 61889068  00-5996-017-01 61889077  00-5996-017-01 61889078  00-5996-017-01 61883275  00-5996-017-01 61883276  00-5996-017-01 61912362  00-5996-017-01 61915324  00-5996-017-01 61921459  00-5996-017-01 61930164  00-5996-017-01 61931682  00-5996-017-01 11400269  00-5996-017-01 61935993  00-5996-017-01 61935994  00-5996-017-01 61944862  00-5996-017-01 61962880  00-5996-017-01 61962881  00-5996-017-01 61962882  00-5996-017-01 61966911  00-5996-017-01 61966920  00-5996-017-01 61966928  00-5996-017-01 61977558  00-5996-017-01 61982469  00-5996-017-01 11400338  00-5996-017-01 61982468  00-5996-017-01 61997346  00-5996-017-01 62015859  00-5996-017-01 62015858  00-5996-017-01 62016808  00-5996-017-01 62026099  00-5996-017-01 62026100  00-5996-017-01 62027205  00-5996-017-01 11400386  00-5996-017-01 62034616  00-5996-017-01 62043318  00-5996-017-01 62043319  00-5996-017-01 62050083  00-5996-017-01 62050084  00-5996-017-01 62054089  00-5996-017-01 62065621  00-5996-017-01 62074427  00-5996-017-01 62079860  00-5996-017-01 62086376  00-5996-017-01 62093706  00-5996-017-01 62093707  00-5996-017-01 62099048  00-5996-017-01 62099049  00-5996-017-01 62099273  00-5996-017-01 62109747  00-5996-017-01 62117336  00-5996-017-01 62113005  00-5996-017-01 62122866  00-5996-017-01 62131270  00-5996-017-01 62145407  00-5996-017-01 62148771  00-5996-017-01 62154075  00-5996-017-01 62161225  00-5996-017-01 62165233  00-5996-017-01 62165937  00-5996-017-01 62169037  00-5996-017-02 61797756  00-5996-017-02 61798634  00-5996-017-02 61806290  00-5996-017-02 61806288  00-5996-017-02 61806289  00-5996-017-02 61835840  00-5996-017-02 61842362  00-5996-017-02 61850928  00-5996-017-02 61850930  00-5996-017-02 61850929  00-5996-017-02 61852845  00-5996-017-02 61889079  00-5996-017-02 61889080  00-5996-017-02 61884197  00-5996-017-02 61894563  00-5996-017-02 61894564  00-5996-017-02 61898502  00-5996-017-02 61913202  00-5996-017-02 61916162  00-5996-017-02 11400221  00-5996-017-02 61922264  00-5996-017-02 61930163  00-5996-017-02 61934977  00-5996-017-02 61929417  00-5996-017-02 61934978  00-5996-017-02 61939188  00-5996-017-02 61939189  00-5996-017-02 11400300  00-5996-017-02 61941599  00-5996-017-02 61962883  00-5996-017-02 61962884  00-5996-017-02 61967904  00-5996-017-02 61967905  00-5996-017-02 61968858  00-5996-017-02 61978301  00-5996-017-02 61983005  00-5996-017-02 62018152  00-5996-017-02 62025181  00-5996-017-02 62025182  00-5996-017-02 62025372  00-5996-017-02 62025376  00-5996-017-02 62043153  00-5996-017-02 62043154  00-5996-017-02 62048444  00-5996-017-02 62049391  00-5996-017-02 62055051  00-5996-017-02 62060537  00-5996-017-02 62067352  00-5996-017-02 62074436  00-5996-017-02 62074433  00-5996-017-02 62080817  00-5996-017-02 62086993  00-5996-017-02 62099934  00-5996-017-02 62099938  00-5996-017-02 62100474  00-5996-017-02 62110801  00-5996-017-02 62110818  00-5996-017-02 62110819  00-5996-017-02 62117338  00-5996-017-02 62117339  00-5996-017-02 62123889  00-5996-017-02 62132112  00-5996-017-02 62145408  00-5996-017-02 62148247  00-5996-017-02 62154285  00-5996-017-02 62161596  00-5996-017-02 62169056  00-5996-017-02 62166737  00-5996-017-02 62173190  00-5996-018-92 61836677    00-5996-012-01 11004400  00-5996-012-02 62068093  00-5996-013-51 61782571  00-5996-013-51 62118559  00-5996-013-51 62134293  00-5996-013-52 62118560  00-5996-013-52 62131269  00-5996-013-52 62170071  00-5996-013-52 62172686  00-5996-014-51 62124846  00-5996-014-51 62131280  00-5996-014-51 62134711  00-5996-014-51 62169053  00-5996-014-51 62169054  00-5996-014-52 11002937  00-5996-014-52 11003326  00-5996-014-52 11003004  00-5996-014-52 61782574  00-5996-014-52 62119552  00-5996-014-52 62137519  00-5996-014-52 62171791  00-5996-014-52 62171792  00-5996-015-51 62121199  00-5996-015-51 62120709  00-5996-015-51 62133177  00-5996-015-51 62138260  00-5996-015-51 62170913  00-5996-015-51 62170915  00-5996-015-51 62170916  00-5996-015-51 62166597  00-5996-015-51 62171094  00-5996-015-51 62173164  00-5996-015-52 62121801  00-5996-015-52 62135973  00-5996-015-52 62136386  00-5996-015-52 62160803  00-5996-015-52 62172688  00-5996-015-52 62172690  00-5996-015-52 62171830  00-5996-015-52 62171831  00-5996-015-52 62171832  00-5996-015-52 62172691  00-5996-016-51 61792784  00-5996-016-51 62121779  00-5996-016-51 62137521  00-5996-016-51 62137541  00-5996-016-52 61782556  00-5996-016-52 62122572  00-5996-016-52 62138262  00-5996-016-52 62139469  00-5996-016-52 62170933  00-5996-017-51 62122868  00-5996-017-51 62131271  00-5996-017-51 62169055  00-5996-017-51 62173166  00-5996-017-52 62123891  00-5996-017-52 62140118  00-5996-017-52 62169057  00-5996-017-52 62170066  00-5996-017-52 62173191  00-5996-018-51 61777248",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}