  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1035-2014",
      "product_description": "MODULAR AUSTIN MOORE SOLI and MOORE HIP PROS REG SOLID",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60169597  365445  60096116  60096117  60134409  60845244  61059560  61613250  366221  60096118  60553617  60743729  60990038  61298521  365622  60096119  60411949  60501830  60540531  60609946  60700469  61657255  365458  60110443  60154855  60187556  60301871  60484369  60590160  60716272  60768572  60887301  61118584  61211680  61304669  61589078  61619986  366386  60096121  60124827  60169413  60242838  60301886  60411970  60484390  60540735  60565849  60610009  61122126  61401537  61694181  61878452  61934378  61992515  365461  366387  367572  60109037  60110290  60124913  60169470  60204787  60242845  60326542  60395275  60423101  60474918  60484409  60768593  60845456  61199171  61298517  61613251  61817985  61992516  365499  367548  369841  60096125  60110448  60148883  60242895  60293315  60484704  60540826  60581312  60737211  61103540  61324218  61619209  61659604  61758278  61817987  366388  369882  60109039  60124924  60134410  60169479  60187599  60242901  60293321  60335415  60442118  60506747  60993378  60993379  61118585  61276626  61398796  61651794  61756168  61758280  60096126  60187653  60412010  60537009  60574597  60787916  61118586  61932403  366389  369842  60124934  60187682  60274760  60423134  60980807  61231910  61651793  61837093  365464  365604  60096127  60355236  60540889  60565935  61662668  61976171  365444  60085816  60148884  60484813  61042186  61290074  61528630  61744433  60301911  60541000  61059561  61555631  61658313  366382  60154916  60395295  60900095  61324217  60096128  60541001  61370524  365448  60096186  60553915  61496597  60096187  60096188  366307  60096189  60096190  60166763  367542  60110458  60301876  60906529  60169412  60326978  60484392  61895487  60099418  60242846  60506729  60096192  60096193  365496  60523403  60331362  365481  366300  61222214  365500  60906530  61886563  366301  60096195  60484817  61886564  365449  366304  60423158  60906531  60096202  60331367  365616  366578  367657  369317  60101998  60101999  60113196  60113197  60145797  60156493  60178358  60187897  60213589  60274027  60284125  60284126  60294382  60313728  60382721  60411695  60467652  60474624  60614956  60708479  60793758  60822078  60822089  60876798  60899020  61092517  61104296  61172285  61199985  61243392  61282926  61293847  61336987  61435528  61555658  61664257  61696348  369400  60102000  60102001  60113198  60135105  60135108  60156495  60156496  60178359  60187902  60213591  60222247  60234255  60273998  60275695  60294383  60313730  60313732  60324334  60345750  60382722  60411698  60457745  60474626  60535059  60551516  60589262  60708480  60793770  60822079  60822090  60876799  60899021  61092518  61172276  61172277  61199987  61263822  61282922  61293846  61337003  61435527  61460812  61522994  61550878  61555659  61696349  365615  365617  369401  60145802  60187905  60324336  60429016  60474628  60474629  60803409  61115345  61351331",
      "more_code_info": ""
    }
  ]
}