  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69299",
      "recalling_firm": "Animas Corporation",
      "address_1": "200 Lawrence Dr",
      "address_2": "N/A",
      "postal_code": "19380-3428",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.",
      "recall_number": "Z-1034-2015",
      "product_description": "Animas Vibe Insulin Infusion Pump.    This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.",
      "product_quantity": "1235",
      "reason_for_recall": "Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.",
      "recall_initiation_date": "20110906",
      "center_classification_date": "20150130",
      "termination_date": "20150810",
      "report_date": "20150211",
      "code_info": "Model Number(s):     100515-63  100510-63  100514-63  100512-63  100511-63  101200-03  100201-03  101202-03  101204-03  101205-03  101200-53  101202-53  101205-53  101200-02  101200-57  101201-57  101202-57  101204-57  101205-57  101200-63  101201-63  101202-63  101204-63  101205-63  101206-63"
    }
  ]
}