      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1034-2014",
      "product_description": "6 DEGREE COCR FEM HEAD 36",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "369934  60308845  60508319  60588077  60670902  60701935  60779840  60790101  60810357  60847644  60906703  61077837  61077838  61191243  61191244  61247015  61304729  61340542  61401735  61485227  61530679  61551614  61555020  61670831  61720512  61774878  61790557  61837158  62025229  62124070  62176438  369935  60339031  60407279  60586160  60658085  60701954  60701967  60779852  60779856  60779863  60806208  60821916  60847645  60900931  61002276  61052833  61077839  61077840  61191246  61304732  61401736  61448090  61502826  61530680  61570271  61618471  61618472  61713710  61734540  61837159  61837160  61896789  62025230  62091051  62176441  62284369  60339030  60392901  60440343  60508338  60656885  60669433  60779864  60779866  60779867  60790102  60806209  60821917  60847646  61002277  61052834  61077841  61144531  61191245  61304736  61401737  61485225  61530678  61551616  61551617  61670830  61713709  61734539  61774872  61779845  61790554  61896791  61896792  62025231  62025232  62124069  62176444  62218172  60323824  60439671  60508340  60536879  60611544  60652728  60684197  60779872  60790103  60847647  60906704  61002278  61077842  61191242  61224848  61304741  61340544  61424894  61485226  61618473  61676503  61713708  61774879  61779846  61790560  61896793  62025233  62093451  62176446  62218173  60439652  60515401  60593679  60701996  60884548  60940020  61018982  61077843  61081687  61191241  61224849  61304743  61459434  61535709  61618470  61683763  61720514  61774871  61779847  61837161  61896795  62025234  62098854  62099285",
      "more_code_info": ""
    }
  ]
}