{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Columbia",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82228",
      "recalling_firm": "ACell, Inc",
      "address_1": "6640 Eli Whitney Dr Ste 200",
      "address_2": "N/A",
      "postal_code": "21046-2626",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution in the state of CA.",
      "recall_number": "Z-1032-2019",
      "product_description": "Gentrix Surgical Matrix Thick, 30 x 40cm, Catalog Number PSMT3040, Lot 013948 - Product Usage: Gentrix Surgical Matrix Thick is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic and reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic &#38; reconstructive surgery includes, but is not limited to the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.",
      "product_quantity": "1",
      "reason_for_recall": "The devices were released from a lot for which one sample failed to meet the company' s tensile strength specification for lot release.",
      "recall_initiation_date": "20190214",
      "center_classification_date": "20190321",
      "termination_date": "20200507",
      "report_date": "20190327",
      "code_info": "SN KN000087"
    }
  ]
}