{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64497",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave NE",
      "address_2": "N/A",
      "postal_code": "55432-3568",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.",
      "recall_number": "Z-1031-2013",
      "product_description": "Medtronic Prostiva RF Therapy Generator.    The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.",
      "product_quantity": "848 (696 US, 152 OUS)",
      "reason_for_recall": "Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off.  If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi",
      "recall_initiation_date": "20130227",
      "center_classification_date": "20130402",
      "termination_date": "20131121",
      "report_date": "20130410",
      "code_info": "all Model 8930 radio frequency (RF) generators are affected"
    }
  ]
}