  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.",
      "address_2": "",
      "product_quantity": "13,227 in total",
      "code_info": "Item Number: 00-7843-012-08  Lot Numbers:  62053685  62116471  Item Number: 00-7843-013-08  Lot Number:  62184774  Item Number: 00-7843-014-08  Lot Numbers:  62184777  62158847  62184778  Item Number: 00-7843-013-81  Lot Numbers:  61987954  61815082  62116484  61956726  Item Number: 00-7843-015-08  Lot Number:  62158851  Item Number: 00-7843-013-82  Lot Numbers:  61937673  61759258  61951276  61815083  62120674  61829373  77000819  Item Number: 00-7843-015-81  Lot Numbers:  61968101  61781981  61968103  Item Number: 00-7843-015-82  Lot Numbers:  61834826  61932605  Item Number: 00-7843-016-81  Lot Numbers:  61735809  61759260  Item Number: 00-7843-016-82  Lot Numbers:  62029071  61775246  62029071N  61981820  77000820",
      "center_classification_date": "20180314",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.",
      "state": "IN",
      "product_description": "VERSYS 8 INCH BEADED FC   12X200MM STR STD BODY/NECK  13X200MM STR STD BODY/NECK  14X200MM STR STD BODY/NECK  REV 13.5X200MM BWD LT  15X200MM STR STD BODY/NECK  REV 13.5X200MM BWD RT  REV 15.0X200MM BWD LT  REV 15.0X200MM BWD RT  REV 16.5X200MM BWD LT  REV 16.5X200MM BWD RT    Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented",
      "report_date": "20180321",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1029-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78706",
      "termination_date": "20200402",
      "more_code_info": "",
      "recall_initiation_date": "20171129",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}