{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76037",
      "recalling_firm": "Fresenius Medical Care Renal Therapies Group, LLC",
      "address_1": "920 Winter St",
      "address_2": "",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign: Canada,  Mexico, Jamaica, Germany, Guam",
      "recall_number": "Z-1028-2017",
      "product_description": "Fresenius 2008K2 Series : Hemodialysis System",
      "product_quantity": "60,151 units (US)  1773 (OUS)",
      "reason_for_recall": "When the UF Rate, Goal or Time is adjusted using the \u001cup\u001d and \u001cdown\u001d arrow keys, and the change is cancelled by using the \u001cesc\u001d key, the \u001ccancelled\u001d UF Rate is actually being executed rather than rate displayed on the machine",
      "recall_initiation_date": "20161221",
      "center_classification_date": "20170113",
      "termination_date": "20230317",
      "report_date": "20170125",
      "code_info": "All serial numbers",
      "more_code_info": ""
    }
  ]
}