  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73083",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution including AZ, CA, GA, IL, MO, MS, MT, NC, NH, NY, OH, OK, and PA..",
      "recall_number": "Z-1028-2016",
      "product_description": "Kopans Breast Lesion Localization Needle  The Kopans Breast Lesion Localization Needle consists of a bevel-tip needle with centimeter markings and a hookwire. The hookwire features a burnish mark that allows the operator to ensure that the hookwire is within the needle tip during needle manipulation. The Kopans Breast Lesion Localization Needle is intended for preoperative marking of nonpalpable breast lesions.",
      "product_quantity": "134",
      "reason_for_recall": "Product potentially exposed to body fluids during processing.  Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of blood borne pathogens is thought to be minimal to non-existent since these devices were sterilized after the potential exposure to body fluids.",
      "recall_initiation_date": "20160108",
      "center_classification_date": "20160229",
      "termination_date": "20170323",
      "report_date": "20160309",
      "code_info": "Catalog # DKBL-20-7.0-A  Lot # 6215428  date of manufacture 09/17/2015"
    }
  ]
}