  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64453",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide Distribution including Puerto Rico and the states of MO, NE, CO, MT, NJ, FL, WA, TX, GA, NY, KS, NH, VA, HI, PA, MI, IA, CA, NC, AZ, UT, LA, IL, TN, ME, MN, KY, CT, OR, NV, AL and  NM.",
      "recall_number": "Z-1028-2013",
      "product_description": "Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001.  Used with the cobas¿ CT/NG 4800 System, for in vitro diagnostics.",
      "product_quantity": "189 units",
      "reason_for_recall": "Certain lots of Reagent reservoirs,  may not have a separation in the lower part of the reservoir.  As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.",
      "recall_initiation_date": "20130301",
      "center_classification_date": "20130401",
      "termination_date": "20140604",
      "report_date": "20130410",
      "code_info": "Lot / SN :  477415,477416, 477898, and 477899"
    }
  ]
}