      "status": "Ongoing",
      "city": "Norcross",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82197",
      "recalling_firm": "CP Medical Inc",
      "address_1": "1775 Corporate Dr Ste 150",
      "address_2": "N/A",
      "postal_code": "30093-2960",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide to IA, IL, NY, SC.      Distributed internationally to Canada.",
      "recall_number": "Z-1027-2019",
      "product_description": "Plain and Chromic Gut Absorbable Surgical Sutures, sterile. 12 sutures per box.    Product Usage:  Plain Gut Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular and neural tissues.",
      "product_quantity": "1,905 boxes (22,860 sutures total)",
      "reason_for_recall": "Potential compromise of product sterility after routine monitoring for Endotoxins showed an out of limit test result.",
      "recall_initiation_date": "20190204",
      "center_classification_date": "20190319",
      "report_date": "20190327",
      "code_info": "Product Code (lot code): 822PG-HUF (170927-52, 180611-59); 822PG-DAR (171205-52); 810PG-1 (180924-51); 822PG-1 (180924-53)"
    }
  ]
}