{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irwindale",
      "address_1": "15715 Arrow Hwy",
      "reason_for_recall": "Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated.  When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.",
      "address_2": "",
      "product_quantity": "54 units",
      "code_info": "Impacted serial Numbers: 120, 132, 134,136, 151, 342, 350, 354, 356, 363, 368, 373, 379, 380, 384, 385, 402, 405, 424, 919, 953, 954, 960, 963, 968, 970, 977, 980, 993, 998, 1001, 1123, 1148, 1378, 1596, 1729, 1745, 1747, 1749, 1750, 1751, 1753, 1754, 1758, 1759, 1760, 1761, 1797, 1799, 1802, 1803, 1804, 1805.    Lot Numbers:  P1210, P1220, P1221, P1231.",
      "center_classification_date": "20130328",
      "distribution_pattern": "Worldwide Distribution - USA (Nationwide) including the states of: DE, NL, CH, FR, GB, DK.",
      "state": "CA",
      "product_description": "20-Pole Eco Cable, Catalog #EM-5050-060    Product Usage:  The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System.",
      "report_date": "20130403",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biosense Webster, Inc.",
      "recall_number": "Z-1023-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64579",
      "termination_date": "20131017",
      "more_code_info": "",
      "recall_initiation_date": "20130305",
      "postal_code": "91706-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}