  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "KEMPELE",
      "state": "N/A",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70068",
      "recalling_firm": "INNOKAS MEDICAL OY",
      "address_1": "TARJUSOJANTIE 12",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the US and Europe.",
      "recall_number": "Z-1022-2015",
      "product_description": "VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.",
      "product_quantity": "20 units",
      "reason_for_recall": "If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.",
      "recall_initiation_date": "20141203",
      "center_classification_date": "20150128",
      "termination_date": "20150424",
      "report_date": "20150204",
      "code_info": "SK514470022YA, SK514470023YA, SK514470024YA, SK514470020YA, SK514470021YA, SK514100003YA, SK514470019YA, SK514470001YA, SK514470017YA, SK514470018YA, SK514450021YA, SK514450022YA, SK514450019YA, SK514470004YA, SK514470005YA, SK514390001YA, SK514150002YA, SK514150004YA, SK514150005YA, SK514300002YA"
    }
  ]
}