  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1020-2014",
      "product_description": "NEXGEN OSTEOTOMY CORT BON. Item codes starting with 0052500",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60114195  60595579  61077273  60187380  60610778  60796106  61137985  61249340  60105628  60595582  60919090  61277444  60330148  60732231  60945764  61283889  60219129  60579389  60756650  60965136  61207433  61242494  60244944  60579385  60756651  60965137  61233951  61593013  61887830  60105629  60166064  60224275  60244943  60337906  60461446  60586318  60595606  60696728  60808905  60906705  61018586  61103031  61180337  61249355  61341882  61422286  61526208  61636571  61717191  61770492  61887831  61929703  62048813  60213807  60331257  60579387  60941524  61233952  61687333  60099537  60139269  60194156  60244942  60330150  60436638  60614725  60681999  60838357  60945765  61089048  61207434  61378442  61585149  61770493  60105630  60191625  60200995  60610790  60858860  61180338  61770494  60139270  60209213  60310131  60579392  60713474  60906706  61089049  61378443  61902190  60503470  60678795  60897195  61233953  61770495  60124943  60230948  60420518  60676026  61056150  61445351  60177068  60330149  60504976  60671244  60897209  61146352  61770496  60294720  60503550  61018587  61585151  60233993  60501971  60834361  61160767  60105632  60503494  60586320  60595609  61238563  60125012  60504978  60844061  61265359  60266112  60503499  60860053  61207435  60209212  60382121  60491633  60671248  60847409  61070792  61238565  61497083  61911709  60099538  60395058  60503554  60682001  61228341  61266850  60160832  60367469  60503479  60858861  61207436  60310132  60491634  60671251  60991375  60105634  60139271  60213808  60330152  60461447  60652031  60861035  61180339  61422287  60135104  60194157  60266113  60334482  60436657  60503483  60667751  60834366  61146353  61445352  62143506  60125013  60213809  60330153  60503489  60965138  61504824  60177432  60263493  60503552  60858859  61331581",
      "more_code_info": ""
    }
  ]
}