  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91320",
      "recalling_firm": "Align Technology Inc",
      "address_1": "2820 Orchard Pkwy",
      "address_2": "N/A",
      "postal_code": "95134-2019",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "U.S. Nationwide distribution in the states of  AZ, CA, CO, FL, KY, MA, NJ, NY, OH, OK, and WI.",
      "recall_number": "Z-1017-2023",
      "product_description": "Invisalign Express 10 REF 8512    Invisalign System   Comprehensive REF 9000    Invisalign System   Moderate REF 9001",
      "product_quantity": "14 cases (637 aligner and temples)",
      "reason_for_recall": "3D orthodontic planning software has a defect that leads to an issue where incorrect number of aligners may be produced (less or greater than the doctor requested) and incorrect packaging is provided.",
      "recall_initiation_date": "20221205",
      "center_classification_date": "20230126",
      "report_date": "20230201",
      "code_info": "Invisalign Express 10 Product Code: 8512 SO/PID: 215086492 / 18752002  Invisalign System - Comprehensive Product Code: 9000 SO / PID: 215072477  / 18747673 213766469 / 17291917 215237342 / 16525818 215211007 / 18799320 215234241 / 15985771 215275953 / 18824845 215035759 / 16193974 215038858 / 18715205 215349325 / 18799786 215125907 / 18746963 215541695 / 16576036 215451877 / 14994856  Invisalign System - Moderate Product Code: 9001 SO/PID: 214866511 / 18672529",
      "more_code_info": ""
    }
  ]
}