  "results": [
    {
      "status": "Ongoing",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81783",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution to states to: CA, ME. MI, and WV.",
      "recall_number": "Z-1017-2019",
      "product_description": "Fructosamine Control 1, FR2994     This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.",
      "product_quantity": "141 kits",
      "reason_for_recall": "The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.",
      "recall_initiation_date": "20181203",
      "center_classification_date": "20190315",
      "report_date": "20190327",
      "code_info": "499FR"
    }
  ]
}