  "results": [
    {
      "status": "Terminated",
      "city": "Plymouth",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70048",
      "recalling_firm": "Covidien",
      "address_1": "4600 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-2890",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries Australia, Canada, Finland, France, Germany, Ireland, Italy, Spain, Sweden, Switzerland, Turkey, and United Kingdom.",
      "recall_number": "Z-1017-2015",
      "product_description": "Covidien Trellis-6 Peripheral Infusion System.  Models BVT608010V01, BVT608030V01, BVT612010V01, and BVT612030V01.  Sterile EO.    Product Usage: Intended for controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.",
      "product_quantity": "216 devices (207 US, 9 OUS)",
      "reason_for_recall": "A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units.  Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.",
      "recall_initiation_date": "20141210",
      "center_classification_date": "20150202",
      "termination_date": "20160331",
      "report_date": "20150211",
      "code_info": "Model BVT608010, Lot numbers:  9853174, 9864497, 9890772, 9890931, 9925676, 9932672, 9937308, 9937315, 9940474, 9940614, 9976340, 9978233, A001589, A001740.      Model BVT608030, Lot Numbers:  9853331, 9887597, 9887695, 9887757, 9932042.      Model BVT612010, Lot numbers:  9854121, 9868214, 9868311, 9875472, 9875632, 9876174, 9883295, 9884788, 9884951, 9886196, 9886312, 9932096, 9968511, 9974109, 9974126, 9976472, 9977091, 9978862, A002299, A002423.      Model BVT312030, Lot numbers:  9854124, 9892081, 9941213, 9971096, 9971139, 9976506, 9977140, 9978964, 9979548, A003021, A003121, A003651, A007900, A008492, A008603, A008687."
    }
  ]
}