{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81783",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution to states to: CA, ME. MI, and WV.",
      "recall_number": "Z-1016-2019",
      "product_description": "Fructosamine Calibrator, FR2993    This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.",
      "product_quantity": "85 kits",
      "reason_for_recall": "The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.",
      "recall_initiation_date": "20181203",
      "center_classification_date": "20190315",
      "report_date": "20190327",
      "code_info": "498FR"
    }
  ]
}