  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1016-2014",
      "product_description": "MOORE HIP PROS LONG FENES, MOORE HIP PROS NRW STR FE, MOORE HIP PROS STR FENEST, THOMPSON HIP PROSTHESIS 3, 4, 5 and 6.",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60484301  60536380  60565715  60753818  60817438  60864746  61003323  61059514  61081364  61121949  61135827  61387138  61783644  365466  60096346  60097010  60259990  60422473  60459752  60515365  60674982  60715057  60768520  60787821  60799626  60799627  61012710  61029791  61029797  61075345  61103552  61103553  61121950  61121951  61135828  61883369  61901575  61938344  369678  369843  60242811  60260004  60317143  60370104  60411929  60411932  60441611  60501819  60501820  60501821  60540465  60553615  60715059  60759283  60759284  60759285  60787991  60799628  60979642  61003325  61012729  61059519  61121952  61121954  61222215  61339786  61398793  61528632  61541580  61565702  61589080  61613237  61921999  61938346  60085561  60096349  60097011  60124640  60134157  60148294  60154764  60154765  60169374  60169375  60187551  60187552  60242814  60260197  60274423  60301862  60313429  60335338  60395245  60395246  60422533  60442559  60484312  60501825  60511856  60553618  60574472  60600313  60609939  60753835  60768538  60768539  60768541  60768542  60787295  60799629  60799630  60990033  61003326  61003327  61012713  61029792  61042227  61059521  61059540  61059541  61121955  61121958  61286077  61293836  61336916  61356578  61365617  61427119  61506293  61550873  61567517  61685104  61708167  61744439  61803839  61846938  61886948  61987166  365462  366297  367557  370033  60091185  60110058  60110136  60110138  60124644  60124646  60134158  60134159  60141863  60148295  60166638  60169376  60169383  60204468  60242818  60242819  60242820  60260206  60274450  60274453  60274486  60293234  60293235  60293236  60293237  60313507  60326529  60395249  60395250  60411943  60411944  60422535  60441625  60459776  60484317  60501827  60523550  60552243  60552244  60565799  60574476  60574477  60574478  60609942  60609943  60646735  60700464  60700465  60743733  60743734  60768553  60787311  60787313  60842140  60900070  60919566  61003328  61029793  61029794  61033867  61033868  61075309  61103522  61103523  61103525  61121959  61166175  61183999  61197657  61218568  61218569  61356579  61365618  61369497  61435961  61468751  61484505  61490076  61541589  61550869  61565703  61579856  61651260  61658293  61736305  61806778  61812090  61832976  61843323  61868575  61899073  61911421  61922002  61943374  61999753  365434  367541  60093205  60099414  60110153  60110154  60110156  60124815  60124816  60124817  60124818  60134160  60134161  60134162  60141864  60154806  60154807  60154808  60154809  60166765  60169391  60169392  60187557  60187558  60187561  60204503  60204505  60242826  60242827  60242828  60260214  60260215  60260216  60274535  60274537  60274538  60293272  60301870  60301879  60313514  60317111  60326535  60326536  60335396  60335398  60335399  60335501  60355069  60370156  60395256  60411951  60411952  60422537  60441686  60441687  60459782  60484351  60501906  60506708  60532115  60553758  60565844  60574481  60600321  60609949  60609951  60609952  60700473  60720006  60727912  60727913  60743737  60757942  60768559  60768560  60768562  60768563  60768564  60787902  60787903  60799631  60799632  60808890  60817439  60829549  60842142  60919567  60979643  61023526  61027849  61029796  61045164  61045165  61059522  61075346  61109415  61121960  61121961  61121962  61121963  61121964  61135833  61205903  61207319  61266251  61282906  61293827  61293830  61318404  61329738  61336918  61339791  61365609  61365610  61381009  61398794  61401529  61427120  61453527  61468750  61472280  61472281  61509857  61522982  61534386  61541602  61562536  61599943  61601327  61613239  61631083  61688425  61708172  61727052  61750634  61779334  61806760  61830025  61866243  61907705  61911422  61938348  61952543  61957373  365440  367558  60089773  60089774  60110160  60110161  60110162  60124825  60134163  60134164  60148325  60154858  60154871  60169410  60187563  60204691  60204694  60204710  60204711  60204712  60204713  60204714  60204715  60242834  60242835  60242836  60260223  60274553  60288258  60293290  60293295  60301883  60317120  60335402  60335504  60370198  60383043  60411960  60441732  60442565  60450000  60480771  60480772  60506709  60506710  60506720  60511581  60522920  60540641  60540642  60562992  60565845  60574493  60574494  60581906  60590161  60600288  60617061  60664710  60678497  60700476  60700477  60727917  60727918  60768574  60768575  60768576  60787912  60817440  60829551  60842143  60864748  60887285  60945804  60945805  61003329  61003331  61012715  61012750  61027850  61027858  61029798  61042229  61042230  61059523  61059544  61059545  61102978  61118575  61118576  61135834  61153016  61153017  61166173  61178685  61178686  61207327  61231912  61250449  61255859  61255860  61307978  61312310  61318407  61365619  61395458  61427121  61468756  61481943  61516817  61522981  61548471  61550871  61562537  61589077  61640526  61688423  61708159  61727053  61736302  61748927  61750629  61779329  61798095  61832981  61885875  61917990  61925006  61928447  61983703  61996726  62018413  365436  365897  366217  366989  367559  60099412  60099413  60110231  60110232  60110277  60120596  60120597  60124829  60139196  60139197  60141865  60141866  60141867  60141868  60148326  60148516  60154899  60154901  60169414  60169415  60169416  60169439  60187564  60187565  60187566  60187567  60187571  60187572  60187573  60187574  60187575  60187576  60204797  60204798  60204801  60204802  60242839  60242841  60274560  60301887  60301888  60313519  60313521  60313522  60326541  60335404  60355182  60355183  60355185  60370205  60395269  60411980  60411981  60411982  60411983  60423090  60423091  60441896  60441897  60459857  60459860  60484395  60484396  60484399  60506723  60506724  60523089  60532201  60532202  60540748  60540749  60574543  60574546  60574548  60581314  60610010  60610011  60646764  60646765  60664753  60664756  60678499  60678500  60727937  60727938  60727939  60743744  60768585  60768586  60768588  60787992  60808892  60817441  60829552  60829553  60834997  60864749  60900074  60919569  60945806  60945807  61012716  61020393  61029799  61042231  61042232  61042233  61046011  61046012  61075348  61081369  61103531  61121965  61135836  61135838  61166171  61166172  61179461  61205891  61218561  61243386  61266252  61278831  61282992  61282993  61293835  61307975  61318402  61356524  61356581  61365620  61369501  61389637  61426170  61426171  61435958  61435959  61435960  61495879  61506286  61506287  61528631  61541581  61541582  61555618  61565704  61585030  61589045  61596758  61620011  61708176  61711959  61736283  61783646  61815822  61831251  61899077  61905768  61968014  61983704  365433  366216  367561  367568  369721  60110311  60110312  60110313  60124916  60124918  60134243  60134244  60134245  60148329  60154904  60154905  60154909  60154910  60169471  60187595  60187596  60204820  60242891  60242892  60242893  60260247  60274752  60301889  60313523  60326545  60335412  60370218  60383045  60395276  60395277  60411999  60436451  60442573  60459905  60484697  60484698  60506735  60506736  60506737  60532242  60540825  60574556  60574557  60574558  60581993  60610176  60664771  60678509  60727964  60727965  60743752  60768595  60768597  60768598  60787301  60817443  60823736  60877919  60887288  60919570  60945808  60979644  61003332  61003334  61012751  61027851  61029801  61059525  61059546  61103548  61103549  61103551  61122092  61122093  61199982  61205901  61207314  61243390  61304660  61318403  61329735  61336920  61356582  61380156  61389636  61495911  61522990  61534383  61550868  61555619  61667276  61683767  61727055  61736303  61767766  61794979  61817937  61837087  61899079  61905767  61976165  61996727  62022711  365463  365575  369722  60089775  60110322  60110323  60120599  60124926  60124927  60124928  60134239  60134240  60148333  60148334  60148335  60148336  60169475  60169476  60169477  60187597  60187635  60204823  60204824  60231605  60242897  60242898  60242902  60274753  60293323  60301898  60317113  60335416  60335417  60335418  60395278  60395279  60395280  60412004  60423126  60423128  60442056  60460001  60476458  60484705  60484706  60484707  60506738  60506739  60506740  60532260  60553889  60553890  60582001  60610183  60646767  60664774  60664775  60664776  60678536  60678537  60718282  60727989  60727990  60743774  60743775  60768605  60768608  60787305  60787307  60817444  60829554  60829555  60842144  60887289  60900075  61003335  61003336  61012752  61029802  61029803  61045169  61059527  61059528  61059547  61075349  61103537  61122095  61122096  61122097  61135840  61136034  61207331  61290075  61290076  61290077  61312306  61318401  61336922  61339815  61365621  61381010  61427122  61483139  61506292  61541590  61550870  61589079  61601330  61613240  61651263  61658298  61685099  61688420  61731618  61779326  61822290  61846940  61866242  61895475  61907706  61917991  61928451  61987168  61992514  367543  60110324  60110325  60124929  60124930  60134260  60187654  60187656  60231606  60242903  60242904  60274756  60293338  60313534  60313535  60326562  60335423  60370243  60370245  60442144  60450167  60460037  60484716  60523720  60532263  60540842  60565898  60574592  60610198  60646768  60678539  60728009  60728010  60728011  60768614  60787914  60808893  60817445  60829556  60842145  60864750  60900076  60979645  61023520  61029804  61059529  61059548  61109417  61135841  61171527  61243389  61312312  61356584  61370548  61483140  61509858  61522978  61538257  61550867  61579857  61613241  61651265  61708168  61758281  61783658  61790446  61831248  61895480  61925014  61951216  61976167  61993152  365465  60124932  60124933  60134276  60148377  60154911  60154912  60169562  60242906  60326596  60335425  60370246  60370247  60395286  60395287  60404750  60423133  60442180  60460056  60484718  60484719  60540846  60540847  60574608  60610217  60610219  60646773  60665147  60678558  60716279  60728038  60743782  60768620  60768621  60808894  60817446  60829557  60829558  60845450  60887290  60900077  61003338  61012719  61029805  61029806  61047073  61047074  61122099  61122101  61262140  61282990  61304650  61336923  61387144  61398795  61490084  61504106  61538258  61555621  61607583  61613242  61667275  61685097  61750633  61840592  61938350  61964724  365432  367571  369724  370107  60110326  60124935  60134282  60148378  60154913  60169589  60187858  60187859  60242908  60260570  60274761  60301905  60326603  60405962  60405963  60423136  60442240  60460072  60484798  60506781  60506782  60532469  60565934  60610269  60610270  60678572  60716282  60728039  60743783  60800009  60834999  60887291  60900078  60940763  60979561  61003339  61029807  61059549  61118577  61199983  61207325  61273019  61356577  61483136  61522983  61538256  61550872  61613243  61640525  61694182  61727051  61783660  61822291  61868576  61911423  61928448  365431  366293  367550  367608  60096351  60097012  60154914  60187861  60231607  60242911  60274762  60326622  60326623  60335426  60335427  60412039  60442287  60450256  60460077  60484801  60506784  60506786  60506790  60523043  60553911  60610273  60610275  60610276  60665151  60728049  60743786  60887292  60900079  60919573  61012730  61042185  61042235  61122103  61135854  61166165  61222212  61290078  61293841  61332290  61336924  61341664  61468753  61472279  61533992  61613244  61631084  61691778  61731619  61832975  61983705  366429  367553  60096353  60097013  60187862  60187863  60242912  60242913  60293366  60326624  60335430  60395294  60412040  60423145  60460086  60484822  60484825  60523677  60540999  60566196  60610292  60700478  60728053  60743790  60829559  60887293  60900080  60945810  61042236  61102979  61118578  61122104  61183995  61222213  61237798  61282907  61336926  61370534  61520889  61534374  61589067  61762045  61790445  61901581  61906270  365435  366290  367555  369772  60089808  60096357  60097014  60124961  60154915  60169594  60187864  60187865  60204951  60242914  60293371  60313606  60335438  60412108  60412109  60442299  60484835  60506792  60553913  60610319  60728057  60743791  60835001  60887294  60900081  61019422  61045170  61118579  61135855  61250448  61298520  61332294  61516816  61555622  61640523  61711958  61855209  61917992  365430  365518  367552  367611  369845  60096363  60097015  60124962  60134299  60154917  60242915  60395296  60442324  60484841  60506802  60553914  60574625  60728060  60768642  60900082  61012754  61045171  61207323  61329751  61341662  61555623  61601347  61651792  61758274  61895482  61951217  365437  367554  369848  370109  60096364  60097016  60260650  60370263  60383088  60523724  60716274  60768645  60835002  61059550  61273009  61293837  61332284  61565706  61640524  61817992  61917995  60200150  60412116  60506824  60835003  61332285  60096366  60097043  61059530  61976168  60162047  60242918  60395242  60523668  60565709  60808895  61027852  61075342  61651269  367711  60085695  60093206  60162048  60169599  60231610  60231611  60260192  60274402  60326523  60335337  60411933  60411934  60501822  60501823  60501824  60540468  60664649  60768528  60823738  60829560  61184006  61197658  61286070  61356528  61468752  61555624  61685101  61736287  61783671  61832979  61914804  61987169  60134367  60411940  61167111  61447724  61809376  60134372  60169381  60187878  60204455  60204456  60274461  60288267  60326530  60395251  60411945  60484325  60501831  60646751  60646753  60664660  60715064  60787316  60799636  60808897  61042238  61072123  61243387  61328305  61369504  61401533  61495909  61667277  61688422  61708169  61718336  61779333  61806775  61817935  61840591  61895483  61968013  61993154  60110339  60110343  60124819  60134378  60162050  60169394  60169395  60169398  60187559  60187560  60204508  60204537  60242829  60242830  60274541  60274546  60293275  60293276  60313513  60317099  60317101  60345744  60370157  60395261  60411953  60441673  60459784  60501921  60501924  60506707  60553756  60565843  60574483  60574484  60624836  60646760  60664692  60678492  60700474  60727915  60739553  60768567  60787896  60787898  60823740  60829564  60945811  61029811  61029812  61045175  61081370  61118580  61122108  61171526  61175114  61179462  61222217  61255824  61262143  61282909  61339902  61356530  61356588  61370527  61380157  61420397  61447723  61453528  61460712  61472271  61501407  61538260  61548469  61601329  61744443  61783675  61803841  61864099  61886960  61943376  61952546  61968012  61983702  60274554  60423086  60553760  61189782  61472283  60110268  60124838  60141894  60141895  60141896  60148517  60169451  60187593  60204819  60242842  60242843  60260234  60313516  60335503  60345745  60345746  60370207  60383170  60383171  60395272  60411987  60411989  60441945  60484405  60484406  60523094  60532207  60540801  60581930  60600340  60624787  60624788  60646766  60727942  60727943  60743745  60768590  60829565  60835004  60842150  61012731  61059553  61109418  61109420  61153019  61164599  61178687  61260561  61286072  61365629  61460750  61481944  61490079  61596749  61647178  61696346  61758270  61914805  61932405  61957377  61999754  62019780  60484699  60787302  61420424  61691779  61841591  61917993  61942541  365527  60093208  60097244  60148519  60169478  60242899  60301901  60326556  60335505  60540830  60787303  60933808  61023530  61029817  61072128  61103544  61159369  61159372  61218565  61266249  61427124  61534373  61555627  61679038  61691777  61762042  61914802  61942544  60134261  60158526  60175946  60293348  60313547  60326563  60553894  60565923  60610190  60768617  60787913  60817449  60864755  61012757  61171524  61205897  61207313  61685105  61999755  61166168  60093209  60134401  60169592  60187860  60242909  60260575  60460074  60565926  60624795  60646785  60678575  60768625  61197659  61205894  61356534  61370535  61460713  61868578",
      "more_code_info": ""
    }
  ]
}