  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue related to loose screws in the \u001cHeater Head\u001d of the Infant Warmer System (IWS). Hot screws from  the \u001cHeater Head\u001d of the IWS could fall onto the bed if the \u001cHeater Head\u001d assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.",
      "address_2": "",
      "product_quantity": "32,073 (US = 24,361; OUS = 7,712)",
      "code_info": "Medical device listing number: D103053, D103054, D103056, D103057, D183529",
      "center_classification_date": "20170111",
      "distribution_pattern": "This product has been distributed domestically and internationally .",
      "state": "WI",
      "product_description": "Infant Warmer System (IWS)",
      "report_date": "20170118",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1015-2017",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "75210",
      "termination_date": "20180712",
      "more_code_info": "",
      "recall_initiation_date": "20160804",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}