  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1011-2014",
      "product_description": "M/DN HUMERAL I/M NAIL. Sizes from 6 MM to 13 MM. Codes starting with 0022541, 002542 and 0022543.",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "367948  60314382  60630218  60691787  61053255  61064495  61117719  61303810  61367343  61434774  368764  369138  60219125  60391605  60468238  60645301  61040164  61040166  61092425  61271218  61298490  61332788  61415337  61565527  61711271  61748698  61565527R  368397  60136443  60269633  60393450  60894576  61031591  61079877  61191197  61242951  61294517  61490690  61578055  367232  367951  369139  60518287  60747414  61026918  61092426  61191198  61244794  61275552  61332789  61413348  61490691  00367232R  367735  367952  60205391  60439856  60917345  61049665  61117720  61191199  61260863  61328159  61394802  61428545  367953  368447  369593  60797052  60917346  61069777  61172854  61290214  61382432  61413349  367725  367954  369141  60820229  60969982  60970713  61290216  61382436  00367725R  367955  60222667  60231603  60977563  61191201  61382437  61445338  60231603R  367956  60205392  60969976  60999538  61275553  61344244  61512881  367933  369142  60165691  60269651  60583693  60622432  60973767  61092427  61266204  61298505  61446672  61453160  61631388  61667272  61747816  61747817  367957  60175558  60278617  60426189  60487390  60622433  60630460  60931489  61031586  61069779  61092428  61226794  61242943  61275555  61285979  61415338  61468148  61711274  367361  60154629  60175559  60684985  60973773  61069780  61113627  61155638  61290219  61458957  61632569  60175559R  60973773R  367934  368448  60139737  60208383  60377209  60426190  60472556  60922833  61069782  61113628  61191202  61413350  61584830  367928  369144  60913694  61159195  61313806  61328160  61415339  60913694R  61159195R  61328160R  367935  368449  369145  61053248  61097069  61191203  61415340  367959  368450  60205393  61231968  61233990  61341152  61395129  61584831  60205393R  61395129R  367960  369146  60685553  60735612  60999519  61117721  61338942  367726  367936  368192  60104823  60119139  60136445  60173954  60216381  60260479  60266612  60454980  60468259  60645310  60897830  60897855  61040167  61117722  61172853  61191230  61223093  61313802  61458958  61597136  61667273  61700135  61747820  61747821  367923  60098284  60136446  60150750  60173955  60208385  60241721  60258736  60310279  60415176  60437847  60454981  60463316  60468260  60630461  60691804  60758132  60894670  60894681  61031558  61031560  61092429  61113629  61117723  61155639  61172838  61234862  61260885  61285977  61313804  61434775  61453175  61538812  61578056  61641827  61723066  60630461R  60894671R  367961  60104825  60136447  60208387  60231264  60294034  60407616  60415177  60487391  60601779  60758133  60999573  61026915  61069788  61155640  61172834  61191231  61237554  61260860  61290222  61313803  61434776  61453166  61542204  61769268  60645318R  367937  60108198  60108199  60249228  60487393  60969979  60969980  61092430  61117724  61223094  61234857  61307943  61490694  61602077  61700153  61752962  367962  368453  61049667  61092431  61120869  61298507  61420488  61608478  61049667R  367963  60205394  61040173  61159197  61313805  61338943  61523850  61313805R  367964  369151  60973768  61290225  61382438  61523852  60187109  60388939  60468261  60684986  60897036  61064496  61281287  61490695  60468261R  367939  369152  60119140  60150751  60173957  60173958  60208389  60294037  60426192  60459737  60460805  60583694  60645302  60711442  60715136  60897037  60897038  60897039  61064497  61113630  61143371  61155641  61237555  61244806  61266200  61275557  61338944  61428546  61458972  61608479  61728582  61747825  367340  367924  60098285  60098286  60136448  60136449  60154630  60173959  60175560  60219126  60231269  60241726  60247151  60258737  60338227  60404724  60426193  60426194  60451672  60451673  60459738  60463301  60583695  60645312  60715137  60758134  60894693  60894695  60969974  60969975  60999544  61026908  61113631  61120872  61155642  61172835  61199925  61232480  61242950  61275558  61303811  61434777  61453171  61542206  61563319  61608480  367728  60098287  60098288  60119141  60136450  60136451  60202923  60260480  60387644  60407617  60447732  60451675  60583696  60684988  60820230  60820232  60820233  60925067  61031579  61113648  61120875  61172842  61199924  61232478  61242946  61275559  61307944  61475211  61523855  61597139  61618143  61632571  61700156  61769270  367966  60108205  60136452  60208391  60241728  60278619  60426197  60550489  60583697  60622435  60747416  60758135  61053250  61053252  61053253  61120879  61155643  61172837  61234859  61260858  61285978  61512882  61632572  61748716  60550489R  60622435R  60758135R  61053250R  367967  60119142  60187110  60501478  61049659  61049660  61079879  61191232  61313807  61413351  60187110R  60501478R  61313807R  367968  60187111  60973769  60973770  60973770R  367969  60208392  61072848  61159198  61382439  61445339  367733  367970  60041044  60281120  60388940  60999549  61237556  61320266  61338945  61445340  60041044R  367732  367883  60119143  60173960  60231273  60269652  60278626  60294038  60454983  60460806  60463313  60583698  60622436  60684990  60897869  60897870  61040084  61064498  61120880  61143372  61155644  61244808  61260884  61313801  61458975  61533707  61563320  61578057  61602078  61728812  61747827  60231273R  61728812R  367804  367925  60082848  60093361  60093362  60098289  60104828  60139738  60173961  60175561  60177391  60241730  60260482  60303630  60303633  60324238  60324241  60367715  60387645  60407618  60437848  60447733  60451676  60472558  60515447  60536769  60601782  60645313  60691809  60715139  60735617  60826072  60826073  60922819  60922823  60937718  60999541  61069791  61113650  61113651  61172830  61232482  61237557  61244819  61281289  61303812  61338946  61434778  61453173  61533708  61542207  61597143  61711275  61719110  61769272  61775802  62220598  367814  60098290  60098291  60136453  60173962  60175562  60208396  60231279  60260481  60363726  60380222  60404454  60407620  60426201  60447734  60472560  60536770  60583702  60583703  60601783  60645309  60735618  60826074  60925071  60925072  60925073  61069792  61113652  61117725  61120882  61172850  61232479  61242948  61271216  61281292  61293418  61608481  61700267  61748717  367972  60074311  60098292  60098293  60136454  60197223  60241731  60278630  60363729  60387646  60393451  60404455  60454984  60515449  60537993  60601784  60758136  60925074  61052171  61052172  61092432  61117726  61155646  61172841  61232484  61237558  61275560  61303813  61542208  61618146  61700279  62025381  60758136R  60176992  60358197  60747415  61053258  61103021  61275561  61290230  61468149  60747415R  367940  60175563  60806064  60950808  60969970  61748719  367806  369153  60950809  60969971  61597147  00367734R  60969971R  367203  367974  60187112  60969977  60973771  61290232  61351414  61530392  00367203R  368755  60074312  60082851  60093363  60177395  60266613  60278633  60426205  60454982  60460807  60684992  61026912  61026913  61040171  61040172  61234863  61260879  61578058  61614586  61618147  61747828  61747829  60684992R  367299  60082853  60119146  60136455  60173963  60176994  60202924  60208398  60260483  60294039  60324242  60338229  60338230  60401293  60437849  60463315  60583707  60622437  60630463  60640160  60691821  60715140  60826076  60931474  60931476  60931478  60999546  60999548  61069794  61092433  61113654  61117727  61172844  61232485  61244824  61276752  61293419  61441068  61523856  61584836  61597155  61649572  61700286  61747830  60119146R  60463314R  367977  60098294  60136456  60136457  60177396  60208401  60266614  60278635  60294040  60310281  60447742  60583719  60583720  60622438  60672021  60826077  60931463  60937719  60937720  60999574  61026917  61113655  61120884  61155647  61172845  61226791  61226795  61242949  61490698  61597159  61618149  61696585  61711350  61748722  61769277  60266614R  367930  60082854  60082855  60098295  60136458  60202925  60216383  60426206  60454985  60622439  60672022  60715141  60913699  60969981  61049672  61092434  61155648  61234861  61260880  61276753  61313797  61632573  61748723  60715141R  367929  368462  60104830  60175564  60426207  60472562  60691822  60973772  61338947  60973772R  368463  60176995  60820237  61266205  61290239  60176995R  60820237R  367979  60205395  61069789  61294521  61413352  61621129  61069789R  367205  368464  60041048  60388941  61211141  61248161  61276754  61295250  61490700  367807  60093364  60177398  60303636  60367717  60426209  60460809  60897868  61040081  61040082  61040083  61113656  61223097  61237560  61276755  61307945  61542210  61700292  61747831  60093364R  368265  60093365  60104831  60136459  60197224  60197234  60338231  60401294  60454988  60454989  60506379  60583721  60645316  60715142  60925065  60925066  61069795  61103022  61113657  61155649  61223096  61237561  61260851  61285974  61293420  61458978  61538813  61621130  61725782  61747832  367206  368266  368466  60104834  60136460  60197235  60269656  60407621  60454990  60472564  60515450  60583722  60622441  60925069  60925070  61069796  61113658  61155650  61172836  61226793  61242947  61281309  61293421  61565528  61602080  61719103  368171  60108210  60108211  60407622  60454991  60468262  60583723  60583724  60601786  61087528  61191233  61260878  61379627  61484940  61748724  60108210R  60601786R  367363  368267  60187113  60360174  60640155  61049662  61049664  61199923  61244830  61303815  61490702  60187113R  60360174R  61049664R  61244830R  367364  60139740  60205396  60678703  61040170  61079881  61271217  61441069  61608482  60139740R  60205396R  61079881R  368269  60255304  60820227  61079882  61338948  61413669  60255304R  61338948R  368270  60041055  60651190  61069797  61133047  61155651  61248160  61332790  61375272  61490703  61069797R  61248160R  61332790R  367731  368271  60074239  60165693  60367718  60404457  60506380  60999540  61211139  61226781  61234858  61285976  61341153  61747833  60074239R  368272  60074319  60074727  60367719  60732928  60950810  61069798  61113659  61155652  61179435  61275563  61453167  61538814  61641829  368470  60176996  60303638  60367720  60454992  60894701  61069799  61159200  61242945  61276756  61293422  61453162  61490705  60176996R  60342195R  60367720R  60454992R  61159200R  367795  60108214  60175567  60472567  60913697  61103023  61179437  61266201  61303816  61377323  61468150  61769282  60342196R  60472567R  368471  369129  60750987  61103024  61191234  61328161  61413353  61328161R  368275  369130  60187114  60969978  61191235  61260877  61434780  368277  368472  60797054  61020630  61313808  61395130  61608483  00368277R  60797054R  367730  368473  369534  60219127  61069801  61164611  61199922  61285975  61348461  60219127R  61069801R  368316  60082849  60583725  60601787  61040174  61226790  61276757  61313798  61584847  60583725R  367360  368317  60640158  60668407  61031578  61172832  61226788  61260856  61307947  61395131  61441070  368318  368474  60082850  60693122  60820239  60917350  61103025  61191236  61293423  61320267  61375273  61453176  60082850R  61293423R  369134  60139741  60202926  60485845  60550493  60917351  60950811  61151166  61242941  61293424  61382440  00368475R  60139741R  60485845R  60550493R  61293424R  368320  369135  60915958  61191237  61275564  61344245  61420489  368321  368476  60041066  60970714  60973766  61313809  61338949  60041066R  60973766R  369137  60217605  60977564  61395132  61584849",
      "more_code_info": ""
    }
  ]
}