{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Issaquah",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89856",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "22010 SE 51st St",
      "address_2": "",
      "postal_code": "98029-7298",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of Washington, Alaska, Ohio, South Carolina, Pennsylvania, Arizona, New York, Texas and the countries of Australia, Czech Republic, France, United Kingdom, Indonesia, Mexico, New Zealand.",
      "recall_number": "Z-1010-2022",
      "product_description": "Acuson 9vE4 Transducer, Model (240)11289564 CE0129, Rx Only ,",
      "product_quantity": "31 units",
      "reason_for_recall": "Transducers (component of ultrasound system) may exceed recommended thermal index limits",
      "recall_initiation_date": "20220307",
      "center_classification_date": "20220427",
      "report_date": "20220504",
      "code_info": "UDI/DI -  (01)4056869243566; Affected Serial Numbers: BT10672 BT10582 BT10617 BT10643 BT10669 BT10580 BT10619 BT10659   BT10611 BT10612 BT10614 BT10673   BT10581 BT10494 BT10610 BT10641   BT10497 BT10615 BT10661 BT10667   BT00248 BT10613 BT10657 BT10640   BT10584 BT10488 BT10481 BT10644",
      "more_code_info": ""
    }
  ]
}