  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66500",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand,  and United Kingdom.",
      "recall_number": "Z-1010-2014",
      "product_description": "INTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG    Some items starting with codes 0022530",
      "product_quantity": "7,044,680 total",
      "reason_for_recall": "Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.",
      "recall_initiation_date": "20131011",
      "center_classification_date": "20140220",
      "termination_date": "20170414",
      "report_date": "20140226",
      "code_info": "60625043  60678699  60690742  60728338  60747676  60757045  60813551  60864020  60864021  60915935  60915936  61003973  60806395  60829584  60864023  60867711  60867767  60550320  60569951  60597681  60653347  60690746  60735651  60757047  60757048  60808005  60817585  60829586  60829587  60864037  60864038  60864040  60867771  61161394  61190864  61194691  61215177  61319429  61379510  60518291  60606752  60620346  60711145  60728343  60757050  60806397  60806398  60829592  60829594  60867782  60867785  61071422  61150075  61173568  61319435  61319436  61379514  60748952  60762088  60806399  60876899  60876901  60876902  61248174  61325065  60123303  60130220  60135214  60142918  60150073  60218846  60243102  60279663  60309141  60323311  60338439  60349093  60387528  60393392  60407659  60422612  60422615  60422617  60422618  60505967  60505968  60569960  60597683  60597684  60602462  60620369  60620371  60690750  60690751  60714161  60714162  60735662  60735663  60757051  60757052  60808010  60808011  60814567  60814568  60829596  60829597  60867787  60867788  60867789  60867790  60867792  60956540  60956547  61003994  61004003  61004006  61004010  61035979  61035980  61064854  61066610  61066611  61071425  61071426  61084123  61100815  61158640  61211625  61211626  61221436  61249859  61271267  61286461  61421225  61496669  61509881  61535667  61564743  61664934  61729550  61729552  61746802  61746803  62262065  60123304  60132558  60142920  60152971  60253769  60253770  60280195  60288490  60310563  60310564  60310565  60323343  60338521  60367544  60393393  60393395  60407661  60410614  60438323  60451198  60472688  60487155  60487156  60491717  60491718  60505969  60602464  60602465  60602466  60602467  60620387  60653365  60690752  60690753  60714165  60735665  60735666  60757057  60757058  60757059  60817613  60829604  60867807  60867809  60867811  60867813  60956553  60979622  60985458  60985459  60985460  61024754  61024757  61031015  61031016  61035968  61035972  61041707  61041709  61041713  61064855  61066614  61066615  61067796  61073739  61091734  61096524  61104693  61133860  61133862  61150093  61153543  61199894  61210272  61224276  61269415  61295067  61362462  61382019  61489978  61504080  61504081  61564747  61650565  61657650  61664935  60099580  60103529  60109578  60109579  60123308  60123309  60127535  60132562  60132563  60135218  60140394  60142924  60142925  60150077  60154195  60154196  60154197  60155336  60167149  60169750  60186875  60218855  60229996  60242017  60243123  60243124  60253778  60253779  60253780  60265723  60265724  60265725  60288499  60288501  60288502  60288503  60288504  60288505  60310578  60310582  60323351  60323352  60338534  60338537  60338539  60338540  60358127  60367556  60367560  60380468  60385469  60393402  60393403  60407668  60407669  60410616  60422634  60425160  60425161  60444819  60444820  60452048  60455028  60472693  60472694  60479134  60479135  60487157  60491724  60491725  60505973  60505974  60505975  60505976  60505977  60569969  60569970  60575726  60602498  60602499  60602501  60602502  60602504  60602505  60602506  60620394  60620395  60620396  60630485  60653377  60653381  60674212  60690760  60690761  60714182  60714183  60714184  60714186  60735674  60735676  60757069  60757070  60757072  60757073  60757074  60817603  60817604  60817605  60817607  60817608  60817609  60876980  60876984  60876985  60876987  60926474  60926475  60926476  60926478  60926479  60926480  60926481  60926482  60966586  60966587  60966588  60966590  60966591  60979628  60985467  61009884  61024666  61030980  61030983  61035951  61041617  61041618  61041698  61064857  61066619  61066621  61067798  61071429  61071440  61075712  61075713  61084125  61084126  61096534  61096536  61133873  61133877  61150110  61150112  61150114  61150116  61161408  61199901  61202928  61202931  61228843  61269436  61269439  61307114  61329372  61343035  61343037  61356918  61356922  61356926  61362470  61373379  61379529  61383095  61387094  61464949  61464950  61489983  61489984  61489985  61496688  61496690  61504085  61504087  61504088  61516681  61523814  61538690  61538691  61542364  61559084  61564751  61564752  61564753  61564754  61564755  61613178  61613180  61613181  61621149  61621155  61652156  61652452  61652453  61652455  61652456  61664943  61746771  61746808  62093620  62093621  62298581  60293437  60338548  60367572  60380479  60393407  60399622  60407675  60429165  60431516  60452052  60487160  60491729  60513247  60537348  60569990  60602514  60602515  60602517  60653400  60674217  60690765  60714198  60735679  60817615  60877003  60877005  60877006  60926499  60926501  60926504  60985470  61003947  61003948  61003950  61031017  61035934  61064859  61066623  61071442  61073741  61075716  61084128  61100822  61258938  61383096  61421234  61484479  61516685  61564759  61641812  61657687  61664954  61746813  61746814  62262076  62296452  62296454  60638362  60757081  60789887  60859226  60958482  60958483  61003985  61112660  61125221  61186054  61221446  61256013  61267717  61319683  61464952  61578715  62319585  60099582  60099583  60099584  60101724  60109580  60135224  60145211  60154215  60165654  60172518  60186882  60196128  60208206  60208640  60242022  60242197  60253790  60265735  60288512  60288513  60310607  60310608  60338564  60338565  60338566  60338567  60367578  60367580  60393422  60407681  60407682  60410620  60425164  60438340  60444823  60451199  60455029  60472701  60487161  60487162  60505985  60505986  60505987  60550367  60550370  60602562  60653415  60653416  60691442  60691443  60714214  60714215  60735691  60735692  60757093  60757094  60757095  60829609  60829610  60829611  60829613  60829615  60877034  60877035  60877037  60877043  60877044  60926505  60926506  60926507  60926508  60926509  60926511  61009674  61133881  61150126  61162603  61231299  61249941  61288956  61319478  61325080  61343060  61356955  61367130  61373394  61373396  61379536  61379537  61383101  61388293  61395923  61729580  61729581  61746774  61746776  61746778  61746819  61746820  62296604  62296606  60452059  60513278  60538413  60570004  60597686  60620421  60653424  60692745  60711169  60714222  60757101  60806461  60877319  60877320  60979625  61003951  61003952  61035978  61071455  61073743  61084131  61096561  61256014  61267732  61535675  62262077  62298592  60797060  60813553  60832655  60867821  60979624  61120293  61158661  61199890  61249868  61277978  61597022  60099586  60099587  60109581  60154222  60169752  60196129  60208644  60210696  60242205  60265742  60304030  60304031  60304032  60323376  60338586  60358147  60358149  60385496  60393430  60407689  60438345  60444826  60468386  60479143  60491737  60506002  60506003  60506004  60506005  60506006  60506007  60653436  60653437  60691463  60714244  60714245  60735708  60757106  60757107  60757108  60829616  60829617  60829619  60829620  60877343  60877347  60877348  60877349  61009684  61009685  61009687  61009688  61009689  61059957  61104707  61133890  61133891  61199903  61269475  61288982  61295128  61329390  61329391  61336006  61349206  61382038  61387097  61395925  61652160  61652476  60385771  60422652  60452064  60468391  60513281  60538414  60597279  60597688  60728340  60735712  60817591  60829621  60832658  60979618  61041736  61066626  61125230  61229535  61231199  60748953  60792402  60806469  60838122  60867932  60867934  60979619  61120298  61158662  61199891  61226101  61249874  61267737  60099588  60103531  60109582  60165666  60218875  60288593  60323385  60338598  60367602  60393436  60407695  60422655  60451204  60452067  60472705  60506026  60513283  60570008  60597690  60620447  60630494  60654022  60714261  60735718  60757115  60814580  60829662  60829664  60831082  60831083  60831084  60831085  60831086  60877687  60877688  61009706  61009707  61178969  61202972  61224299  61238883  61244476  61295173  61307171  61319508  61336012  61336017  61383107  61388294  61395928  61395929  61759093  60383971  60422663  60538417  60606753  60747678  60908919  61031037  61086382  61249878  62298614  60748595  60813554  60832662  60857986  60877699  61003986  61109140  61158665  61211654  61242347  61256017  61277981  61597039  60087147  60123322  60129685  60323391  60363834  60399640  60517101  60532036  60532037  60561122  60570011  60597694  60714269  60735724  60757123  60757124  60817617  60831087  60831090  60877707  60877709  60877711  61009726  61009728  61194738  61215262  61238886  61249977  61307182  61329420  61395930  61395931  62306324  60123323  60155344  60253810  61091750  61226107  62306325  60748599  60813555  60853678  60915947  61023560  61158667  61186067  61258953  61293781  60103532  60132575  60165671  60253813  60253814  60573904  60575747  60654048  60691548  60714275  60735790  60757138  60757139  60814488  60831092  60831093  60877724  60877725  61009856  61009860  61202977  61224304  61231379  61288989  61307192  61319519  60598248  60657681  60789868  60831094  60857987  61066632  61104713  61226109  61564768  60535279  60538418  60573905  60597695  60620466  60654110  60691555  60711168  60714287  60757145  60814574  60858302  60858303  60858305  60877737  60950627  60950634  61173764  61173765  61224306  61295186  61307196  61307197  60085611  60253818  60817589  60853665  61066636  61535677  60099589  60573906  60620230  60620555  60714295  60735802  60757152  60757153  60813558  60831095  60831096  60877749  60877751  61035976  60757157  60910708  61091762  61578722  61746793  60693121  60714308  60732959  60735806  60757162  60813559  60847860  60847861  60877766  61004017  61004018  61267751  60678702  60754402  60789877  60883669  61067806  61091765  61535679  61746794  60691641  60721969  60757169  60807457  60831097  60857989  60869321  60877781  60950655  61031033  61031034  60757172  60807461  60831098  60873781  60877795  61191333  60099590  60103533  60670024  60686898  60711170  60757174  60789819  60814353  60831099  60853681  60877798  60915944  60915945  60757178  60831101  60873774  60686905  60711167  60714326  60757180  60789884  60853685  60869322  60869442  60927671",
      "more_code_info": ""
    }
  ]
}